Libraries and Learning Services

University of Auckland Research Repository, ResearchSpace

Version

This is the Accepted Manuscript version. This version is defined in the NISO recommended practice RP-8-2008 http://www.niso.org/publications/rp/

Suggested Reference

Schell, A., Budgett, D., Nielsen, P., Smalldridge, J., Hayward, L., Dumoulin, C., &

Kruger, J. (2016). Design and development of a novel intra-vaginal pressure sensor array. In Neurourology and Urodynamics Vol. 35 (pp. S355-S356).

Tokyo, Japan: Wiley. doi: 10.1002/nau.23074

Copyright

Items in ResearchSpace are protected by copyright, with all rights reserved, unless otherwise indicated. Previously published items are made available in accordance with the copyright policy of the publisher.

This is the peer reviewed version of the article above which has been published in final form at 10.1002/nau.23074

This article may be used for non-commercial purposes in accordance With Wiley Terms and Conditions for self-archiving.

For more information, see General copyright, Publisher copyright,

SHERPA/RoMEO.

(2)

414

Schell A1, Budgett D1, Nielsen P1, Smalldridge J1, Hayward L2, Dumoulin C3, Kruger J1 1. University of Auckland, 2. Counties Manukau District Health, 3. University of Montreal

DESIGN AND DEVELOPMENT OF A NOVEL INTRA-VAGINAL PRESSURE SENSOR ARRAY

Hypothesis / aims of study

The measurement of the vaginal pressure profile is likely to be a desirable metric for both the monitoring and assessment of effective pelvic floor muscle training (PFMT), and the assessment of change in the vaginal pressure profile pre- and post-surgical correction of pelvic organ prolapse (POP) [1].

There are several commercially available ‘pelvic floor trainers’ but none have the ability to simultaneously measure abdominal pressure and PFM pressure, and is feasible to be worn during everyday activities [2, 3]. The aim of this project was to design and develop a wireless intra-vaginal device which would conform to the anatomical shape of the vagina; could be used during everyday activities; measure changes in the pressures exerted along the length of the vagina without distorting the vaginal walls; remain in position; display the pressures measured on a user friendly interface and have a recording frequency fast enough to capture rapid changes in pressure. The pressure sensor array had to also show good repeatability and reproducibility of the pressure profiles.

Study design, materials and methods

The intra-vaginal pressure sensor device (IVPSD) contains an array of eight pressure sensors (MS5803-02BA, Measurement Specialties, United States) which are mounted onto a flexible printed circuit (FPC) board to allow the device to conform to the anatomy of the vagina. The device went through several iterative shape designs to determine the most suitable for comfort and retention. The final design has a total length of 80 mm and a maximum width of 20 mm (Figure 1). The contoured edges cover a distance of 55 mm and are designed to sit within the rugae of the vaginal wall to reduce device movement. The cover is made out of a soft, biocompatible silicone (MED-4901, NuSil, United States), and cast using a silicone transfer press. The electronics module uses a nRF52832 (Nodic Semiconductor) low energy Bluetooth radio system. Data is transmitted to an Android tablet for data logging and realtime display and user feedback. Each pressure sensor sampled at a rate of 140 Hz.

Bench testing was performed to verify the performance of the device. It included the assessment of: pressure sensor drift, response to rapid pressure changes, bending and flexing of the device, hysteresis, hydration of the silicone encapsulation an d temperature sensitivity. In order to validate the repeatability of the device, pressure profile measurements were obtained from four subjects (no symptoms of POP) across four different tasks: maximum PFM contractions (3x 5s), rapid PFM contractions (15s), Valsalva (3x 5s) and coughing (5x). For each test the device was self-inserted with the sensors facing towards the pubic bone (anterior). The order of tasks was random. After each set of tasks the device was removed and re-inserted to evaluate the repeatability of the pressure profiles. Data was analysed using MATLAB.

Results

Bench testing identified that bending and the effect of body temperature dominated the accuracy of the pressure measurements and their combined effect could introduce an error of approximately 5 mmHg.

All subjects for intra-vaginal testing were parous (2 vaginal, 2 caesarean). Mean age 49.2 years SD± 8.5. Due to the small sample size, no statistical analysis could be performed. The vaginal pressure profiles were distinctive for each task and subject. Figure 2 shows the profiles in blue for the initial insertion and in red for the re-inserted device. The overall reproducibility of the profiles was good. However, for some subjects the profiles of the second insertion were shifted. This was likely due the difference in device positioning for the re-inserted device. Retrospective data processing could compensate for the shift.

Silicone cover

FPC

1 2 3 4 5 6 7 8

Pressure sensor

Figure 1: A) Prototypes of different shape designs for trialling. B)The IVPSD. The head of the device is sitting deep within the vagina. The top sensors measure abdominal pressure. The lower end of the device is located at the introitus. The sensors measure near atmospheric pressure.

A B

(3)

Figure 2: The vaginal pressure profiles of different tasks and subjects. Atmospheric pressure has been subtracted for the baseline pressure. The baseline pressures have been subtracted from all other profiles. Blue graphs show the profiles of insertion one, red graphs of insertion two.

Interpretation of results

The reproducibility of the vaginal pressure profiles appears to be good. When comparing data from different sessions, it seems the position of the device can vary, particular during tasks such as cough. This data provides confidence that the magnitude of the pressure differences generated when performing PFMT tasks is much larger than the magnitude of the error.

Concluding message

Bench-top and in-vivo testing verified that the IVPS device is reliable and repeatable when measuring the pressure profile along the length of the vagina at rest and during exercise.

References

1. Neurourol. Urodyn., vol. 24, no. 3, pp. 243–247, 2005.

2. “PeriCoach.” Online. Available: http://www.pericoach.com/. Accessed: 24-Feb-2016.

3. “M!NNA kGoal.” Online. Available: http://www.minnalife.com/products/kgoal. Accessed: 24-Feb-2016.

Disclosures

Funding: J Kruger - Aotearoa Postdoctoral Fellowship Clinical Trial: No Subjects: HUMAN Ethics Committee: University of Auckland Human Participants Ethics Committee REf no 010994 Helsinki: Yes Informed Consent: Yes

-5 15 35 55 75 95 115

1 2 3 4 5 6 7 8

ΔPressure (mmHg)

Sensor Number

Subject #1 Max PFM Contractions

-55 15 25 35 45 55 65 75 85

1 2 3 4 5 6 7 8

ΔPressure (mmHg)

Sensor Number

Subject #2 Max PFM Contractions

-55 1525 3545 5565 7585

1 2 3 4 5 6 7 8

ΔPressure (mmHg)

Sensor Number

Subject #3 Max PFM Contractions

-55 1525 3545 5565 7585

1 2 3 4 5 6 7 8

ΔPressure (mmHg)

Sensor Number

Subject #4 Max PFM Contractions

-5 5 15 25 35

1 2 3 4 5 6 7 8

Pressure (mmHg)

Sensor Number Subject #3 Baseline

-5 15 35 55 75 95

1 2 3 4 5 6 7 8

Pressure (mmHg)

Sensor Number Subject #3 Max P Coughing

Cough1 Test1 Cough2 Test1 Cough3 Test1

Cough4 Test1 Cough5 Test1 Cough1 Test2

Cough2 Test2 Cough3 Test2 Cough4 Test2

Cough5 Test2

Introitus Abdomen

Referensi

Dokumen terkait

Sex, ethnicity, work status, smoking status and clinical characteristics of disease duration, current pharmacological management, the lo- cation of foot pain, and extra articular foot

oleraceum seed to retain viability despite extended immersion in salt water, coupled with an inability to germinate while in seawater and the long attachment time of seed mucilage to

The efficacy of the proposed approach was evaluated on ten distinct, single and multiple simultaneous disturbances which were simulated following the definitions of IEEE 1159-2009

But this attribution immediately poses the question: “Does the absence of macroscopic force at this short muscle length belie the continued cycling of crossbridges, thereby inflating

3  Has the proper credentials as defined by the supervisor’s discipline or profession  Has the appropriate knowledge of and skills for clinical supervision and an awareness of his

Baseado no exposto, o presente estudo teve como objetivos: 1 avaliar a cober- tura do SISVAN no Estado e nas Coordenadorias Regionais de Saúde do Rio Grande do Sul; 2 des- crever o

Intercalibration of CERES, MODIS, and MISR reflected solar radiation and its application to albedo trends Yizhe Zhan1and Roger Davies1 1Department of Physics, University of Auckland,