NO
JML
SATUAN
2
UNIT
*. Max 250 g
*. Readabilty ± 0.1 mg
*. 0,2 mg
*. Max 0.2 mg
*. 0,6 mg/piece
*. WDH : 210 x 330 x 335 mm
*. 80 mm
*. Instalation, Training, Guarantee min. a year, Original operation Manual, Dust Cover.
1
UNIT
*. Min 1800 x 875 x 1500 mm
*. Internal dims Min 1720 X 680 X 1200 mm
*. ± 270 mm ( 10.6")
*. 580 mm, vertical rising sash
*. Max : 1795 cmh, Min : 900 cmh
*. 780 lux
*. Phenolic resin Laminate or equivalent
*. 220 - 240 V, 50-60 hz
*. Base Storage Cabinet with Standard Fittings for 6ft/; Exhaust Fan/Blower Polypropylene (diameter : 250mm)
*. Instalation, Guarantee min. a year, Original operation Manual
2
UNIT
*.
2 set
*. Parallel-type double plunger *. 0.0001mL/min. to 10.0000mL/min.
*. No more than 1% or 0.5uL/min whichever is greater (0.01 to 2m L/min).
*. No more than 0.06% or 0.02 min SD. whichever is greater
10 uL. 40MPa.
0.03 MPa (for water at 1.0 mL/min and 7 MPa). Supported.
*. Manual rinsing or automatic rinsing using optional product.
*. Liquid-leakage sensor, high pressure/low-pressure limits.
*.
2 set
*. 4°C to 85°C or wider
Min piece wight at piece count Overall dimensions(include draft shield)
Weighing plate
NAMA ALAT DAN SPESIFIKASI
Face velocity 0,5 m.s
Accesories
Included of Light intensity
Construction Main Body
1 Analytical balance*)
Specification : Capacity
Reproducibility Linearity
Includes
ADDENDUM SPESIFIKASI TEKNIS
PENGADAAN ALAT LABORATORIUM BPOM DI PALANGKARAYA
TAHUN ANGGARAN 2012
2 Fumehood *)
Specification :
External dims ( LxDxH in mm )
Exhaust outlet Max sash opening
Power Supply
Typical Pulsation
Constant-pressure solvent delivery
3 HPLC with UV-Vis detector and autosampler*)
Specification : PUMPS
Solvent delivery method Flow-rate setting range Flow rate accuracy
Flow rate precission
Plunger capacity
Maximum discharge pressure
Plunger rinsing mechanism
Safety measures
*.
4 set
*. Total-volume sample injection, variable injectionvolume
*. 0.1µL to 100µL
*. Max 1 %
*. RSD 0,3 % max
*. 175 (1ml vials)
*. 260 x 415 x 500 mm (w x h x d), 27 kg *. 100 VAC, 100 VA, 50/60 Hz
*.
2 set
*.
2 set
*. D2 lamp
*. 190 nm to 700 nm or wider
*.
*. Intel Core i3-2100 Processor, 2GB 1x2G
CMel/CTino D3-1333, Windows 7 Professional 32-bit, HDD 500GB 7200RPMSATA, Intel Graphics Media Accelerator (GMA) X4500HD, HP USB Optical Mouse , USB Standard Value Keyboard, HP 15-in-1 Media Card Reader, HP SuperMulti (JB) 1st ALL, Monitor HP LCD 18,50", Laser Jet HP 1102, speed up 17 ppm, 1200 dpi effective output, USB 2 port
*. Data analysis & reporting for food and drug analysis *.
*.
2 set
*.
2 set
*.
2 set
*.
4 set
*. ANALYTICAL COLUMN C18, 25CM, 5 um
4 set
*. ANALYTICAL COLUMN C8, 15CM, 5 um
4 set
*.
2 set
*.
2 set
*.
2 set
*.
2 set
*.
2 set
*. *.
2
UNIT
*. -2 to 19.999 or wider
*. ± 0.002 or less
*. -1999 to +1999 or wider
*. ± 0.1 or less
*. -30.0 dC to +130.0 dC or wider
*. Max three pH points
*. 2 unit buffer pH 4.01 250 ml : 2 unit buffer pH 7.00 250 ml; 2 unit buffer pH 9.21 250 ml; 2 unit InLab® Expert Pro "3-In-1" pH electrode
*. Installation, training min a year, original operation manual
Wavelength range Lamp
Injection volume accuracy Injection volume precision Number of processed samples Dimensions wight
Power requirements
WORKSTATION : AUTO SAMPLER
Injection method
Injection-volume setting range
DEGASSER
UV-VIS DETECTOR
Reservoir Tray HUB W/ Cable Hardware
Software ACCESSORIS
Maintenance Kit Auto rinsing kit for Mixer
Sample vial, 1.5 ml (100pc/ pk)
MICRO SYRINGE 100 uL
4 pH meter*)
Specification : pH Range
pH Relative accuracy mV Range
mV Relative Accuracy Range temperature Calibration Acessories
Dust Cover & Tools Kit Stabilizer 3 KVA INCLUDE
Instalasi, Training, Guarantee min 1year, Manual operating book, Factory certificate
1
UNIT
*. 340-900 nm or wider
*. ± 1.5 nm
*. 0.1 nm or less
*. 1 nm or less
*. 5 nm or less
*. Internal, automatic at power on, visual feedback
*. 10 to 40oC
*. Tungsten
*. Line : 100 to 240 V, 47/63 Hz; automatic changeover, Battery : Lithium-Ion 11V/4400mAh
*. 216 x 132 x 330 mm
*. Included : 1-in.square matched glass sample cells; Cell adapters for 1-inch round/AccuVac cells, 1x1 cells and multi-path 1-inch/1-cm cells; Universal power supply, 100 to 240V, 47/63Hz, with plug adapters for EU,GB,US,China; Protective cover for storing adapters; Dust Cover
*. Instalation, Training, Guarantee min. a year, Original operation Manual
4
UNIT
*. Touch control keypad
*. 2 l
*. 60W (at 43 kHz ± 2 kHz)
*. 153 x 138 x 100 (l x w x d) mm
*. 180 x 219 x 179 (l x w x h) mm
*. 2, 4 appr kg
*. Original operation manual, guarantee min a year
1
UNIT
*. 0.1 uS/cm or less
*. < 6 ppb
*. < 1/ml
*. < 1 CFU/ml
*. 0,6 l/min
*.
*. Min 3 L/ h (output per 8 hours : 24 L) at 15 °C
*. 1 6 bar
*.
The System includes a pressure pump for supplying equipment connected downstream :
Type I Ultrapure water for Molecular biology, microbiology, PCR, inorganic and organic trace analysis, HPLC, ICP, TOC-analysis
Type II High purity water for Rinsing laboratory glassware, Preparing and diluting buffers, reagents, tissue culture media and dyes, Sample preparation for analytical methods such as flame AAS
Bath Size
5 Spektrofotometer portable untuk Laboratorium Keliling *)
Specification : Wavelength range Wavelength accuracy
Operating Temperature Source lamp
Wavelength Calibration
Power Requirements
Dimensions
Wavelength reproducibility Wavelength rosolution Spectral Bandwidth
6 Ultrasonic cleaner *)
Specification
Pengoperasian Penggunaan Volume
Acsesories
Included of
Pulse swept power
Include
Footprint Net Weight
7 Water Deionisation App / Aqua Demineralisator/ Water Puripication *)
Specification : With UV-Lamp Conductivity
TOC Particles Bacteria
Included of UV/UF direct system, 3 Pre-treatment with integrated reverse osmosis ultra pure water, Sterilizing filter, Ultra filter, UV oxidation 185/254 nm,Tank 6 litres with pressure, Recirculation pump Instalation, Training, Guarantee min. a year, Original operation Manual Flow rate
1
UNIT
*. Min 8 L/h
*. Stainless steel or equivalent
*. 2.3 µs/cm or less
*. Conform
*. 16 Liter (built in)
*. 220 V, 50/60 Hz, 1 phase, 3.0 KW
*. Pre-filter with filling; *Dechlorite Filter (Cat. # 2904); *Phospate Cartridge (Cat. # 2906); *Pre-Filter (Cat. # 2912)
*. Instalation, Guarantee min. a year, Original operation Manual
1
UNIT
Specification :
*. Min 45 Liters
*. Stainless steel or equivalent
*. l x w h : 818 x 516 x 296 mm *. l x w h : 590 x 350 x220 mm
*. 10 - 95⁰C or wider
*. LED digital display with PID microprocessor *. 1 minutes to 99,59 hours or better
*. 220-230 V;50/60 Hz
*. Flat Stainless steel cover with opening concentric ring sets (8 openings); Guarantee min. a year, Original operation Manual
Digestion Apparatus Kjeldahl *)
1
UNIT
*. Max 600o
C
*. 6
*. LED digital display with PID microprocessor
*. 600x250x300 (wxdxh) mm
*. 250 mL
*. 220-230 V;50/60 Hz
*. Water jet pump WSP, 10-0088 1 unit; lsoversinic tubing 1,2/17 lSO, 10-009L 1 unit
*. Guarantee min. a year, Original operation Manual
Interior Heating Concept Stainless Steel Casing
Electrical Accessories
Casing, container and cover made of
Temperature range and control
8
Water Distillation App. *)
Specification :
Destilation Capacity Heating material Conductivity
Energy saving through distillation of the heated cooling water
Storage tank contents
Included of
9
Waterbath *)
Capacity
10
Specification : Temperature Heating places
Temperature range and control
Included of Dimensions Size of flasks Electrical supply Accessories Control panel Timer