• Tidak ada hasil yang ditemukan

Bacterial transfer, biofilm formation and effect of device design in four peripheral IV blood-control catheters

N/A
N/A
Protected

Academic year: 2024

Membagikan "Bacterial transfer, biofilm formation and effect of device design in four peripheral IV blood-control catheters"

Copied!
5
0
0

Teks penuh

(1)

INTRODUCTION

Marcia  Ryder

1

,  Elinor  deLancey  Pulcini

2

,  Albert  Parker

2

,  and  Garth  James

2  

(1)  

Ryder  Science,  Nashville,  TN,  

(2)  

Center  for  Biofilm  Engineering,  Montana  State  University-­‐Bozeman,  MT  

           …..medical biofilm research

Four  PIVCs  were  tested:  ViaValve®  Safety  I.V.  Catheter  (Smiths),  the   Autoguard™  BC  Catheter  (BD  Medical),  the  Introcan  Safety®  3  Catheter   (B.  Braun),  and  the  Surflash®  Plus  Safety  I.V.  Catheter  (Terumo/  

CareFusion).    

      The  inserJon  of  peripheral  intravenous  catheters  (PIVC)  is  the  most  

common  invasive  procedure  performed  in  hospitals.  PrevenJng   needlesJck  injury  and  bloodborne  pathogen  exposure  to  healthcare   workers  during  PIVC  inserJon  has  been  an  ongoing  challenge  and   places  healthcare  personnel  at  risk  for  HBV,  HCV,  or  HIV  infecJon.  

The  vascular  catheter  industry  has  responded  with  a  new  generaJon   of  advanced  technologies  that  eliminates  blood  exposure  during  the   inserJon  procedure.    

 

Valved  blood  control  PIVC  (BC-­‐PIVC)  technology  requires  the  addiJon   of  internal  components  within  the  hub  to  prevent  blood  exposure  to   clinicians.    The  added  components  increase  the  internal  surface  area,   dead  space  and  volume  within  the  catheter  hub  that  is  thought  to   increase  biofilm  formaJon  and  subsequent  transfer  of  bacteria  into   the  bloodstream  increasing  the  risk  of  catheter-­‐related  bloodstream   infecJon  (CRBSI).    

 

The  2011  Infusion  Nursing  Standards  of  PracJce  indicates  that  the   nurse  should  consider  replacement  of  the  short  peripheral  catheter   when  clinically  indicated  rather  than  the  previous  72-­‐96  hour  resite   standard.    The  increased  potenJal  for  biofilm  formaJon  in  the  valved   blood  control  catheters  may  increase  the  risk  of  PIV-­‐CRBSI  when  le\  

in  place  for  extended  periods  of  Jme.  

Nine  experiments  were  run  with  three  Jme  points  measured  within   each  run:  0,  72  and  96  hours.  Catheters  were  precondiJoned  with   Bovine  Serum  Albumin  (BSA)  using  a  pressurized  serum  bo^le  to  mimic   venous  inserJon  of  the  catheters.  

                 

A\er  precondiJoning,  a    needleless  connector  (SmartSite®)  was   a^ached  to  each  catheter,  inoculated  by  flushing  with  0.5  ml  of  a  104   colony  forming  units  per  ml  (CFU/ml)  of  Staphylococcus  aureus,  and   incubated  at  room  temperature  for  2  hours.    

               

The  connectors  were  then  replaced  with  new  sterile  connectors  and   una^ached  bacteria  were  rinsed  from  the  fluid  path  using  sterile   Phosphate  Buffered  Saline  (PBS).    

 

Catheters  were  then  either  sampled  or  subjected  to  simulated  clinical   use  by  flushing  17  Jmes  daily  with  0.5  ml  sterile  nutrient  and  1  flush  at   the  end  of  the  day  with  normal  saline  for  72  and  96  hours.  The   catheters  were  then  either  sampled  or  subjected  to  simulated  clinical   use  by  flushing  17  Jmes  daily  with  0.5  ml  sterile  nutrient  and  1  flush  at   the  end  of  the  day  with  normal  saline  for  72  and  96  hours.    

                        PURPOSE

The  purpose  of  the  this  study  was  1)  to  compare  the  bacterial   transfer  rate  and  intraluminal  biofilm  formaJon  between   three  valved  blood  control  PIVCs  in  a  clinically  simulated  in   vitro  model  and  2)  to  determine  if  the  added  components   within  the  catheter  hub  with  larger  surface  area  and  volume   would  also  have  larger  bacterial  transfer  when  flushed.    

METHODS METHODS

ViaValve™  Safety  I.V.  Catheter  

BD  Insyte™  Autoguard™  BC  Shielded  I.V.  Catheter  

B.Braun  Introcan  Safety®  3  catheter  

 Surflash®  Plus  Safety  I.V.  catheter    

(2)

RESULTS

Bacterial transfer, biofilm formation and effect of device

design in four peripheral IV blood-control catheters

Marcia  Ryder1,  Elinor  deLancey  Pulcini2,  Albert  Parker2,  and  Garth  James2  

There  were  staJsJcally  significantly  smaller  bacterial  mean  log  densiJes  (LD)    in  Flush  1  and  Flush  2  for  the  ViaValve™  compared  to  either  the  Autoguard™  BC  (p  =  0.003  and  0.001   respecJvely),  Introcan®  Safety  3  (p  =  0.014  and  0.010  respecJvely)  or  Surflash®Plus  Safety  IV  catheters  (p  =  0.014  and  0.005  respecJvely).    There  were  no  differences  in  the  bacterial   mean  log  densiJes  between  the  Autoguard™  BC,  Introcan®  Safety  3  or  Surflash®  Plus  Safety  I.V.  catheters  in  either  Flush  1  or  Flush  2  (p-­‐value  =  0.273  and  0.687  respecJvely).  

 

For  the  ViaValve™,  the  rate  of  increase  of  the  bacterial  mean  LD  over  Jme  was  small  but  not  staJsJcally  significant  in  either  Flush  1  (0.16/day)  or  Flush  2  (0.23/day).    For  the   Autoguard™  BC,  there  was  a  staJsJcally  significant  rate  of  decrease  over  Jme  in  the  Flush  1  (-­‐0.21/day)  but  not  Flush  2  (-­‐0.05/day).    For  the  Introcan®  Safety  3,  a  staJsJcally   significant  increase  was  found  in  both  Flush  1  (0.71/day)  and  Flush  2  (0.91/day).    For  the  Surflash®,  a  staJsJcally  significant  increase  was  found  in  both  Flush  1  (0.43/day)  and  Flush  2   (0.61/day).  The  rates  of  change  in  bacterial  mean  LD  over  Jme  amongst  the  three  catheters  were  not  significantly  different  in  either  Flush  1  (p-­‐value  ≥  0.121)  or  Flush  2  (p-­‐value  ≥   0.063).  

         

Table  1.    Yellow  indicates  that  the  mean  LD  of  bacteria  in  Flush  1  through  the   ViaValve™  catheter  was  staWsWcally  significantly  smaller  than  either  the  mean  LD  for   the  Autoguard™  BC  (p-­‐value  =  0.0030)  or  the  mean  LD  for  the  Introcan®  Safety  3  (p-­‐

value  =  0.014)  and  Surflash®  Plus  Safety  I.V.  catheter  (p-­‐value  =  0.014).      Green   indicates  that  the  mean  LD  the  Autoguard™  BC,  Introcan®  Safety  3  and  Surflash®  Plus   Safety  I.V.  catheter  were  not  staWsWcally  significantly  different  (p-­‐value  =  0.197).  

Table  2.  Yellow  indicates  that  the  mean  LD  of  bacteria  in  Flush  2  through   the  ViaValve  Safety  catheter  was  staWsWcally  significantly  smaller  than   either  the  mean  LD  for  the  Autoguard  BC  (p-­‐value  <  0.0005),  the  mean  LD   for  the  Introcan®  Safety  3  (p-­‐value  =  0.010)  or  the  Surflash®  Plus  Safety  I.V.  

catheter  (p-­‐value  =  0.005).  Green  indicates  that  the  mean  LD  the   Autoguard™  BC,  Introcan®  Safety  3  and  Surflash®  Plus  Safety  I.V.  catheter   were  not  staWsWcally  significantly  different  (p-­‐value  =  0.687).  

Catheters  were  sampled  with  a  two-­‐step  procedure.  First,   each  catheter  was  flushed  to  recover  planktonic  bacteria  and   plated  to  determine  CFU/ml  (Flush  1).    

                   

The  connector  surface  was  disinfected,  sonicated  in  PBS  to   remove  firmly  a^ached  (biofilm)  bacteria,  and  flushed  a   second  Jme  and  plated  (Flush  2).    

METHODS

In  over  nine  experimental  runs,  the  bacterial  log  densiJes   (LD)  in  Flush  1  and  Flush  2  were  compared  among  four   catheters  types:  the  ViaValve™,  Autoguard™  BC,  Introcan®   Safety  3,  and  Surflash®.  Three  experiments  were  performed   with  the  Autoguard™  BC  and  the  ViaValve™.    Three  other   experiments  with  the  Introcan®  Safety  3  and  the  ViaValve™.    

The  last  three  experiments  with  the  Surflash®  and  the   ViaValve™.    In  each  experiment,  the  bacterial  log  densiJes  in   Flush  1  and  Flush  2  measured  at  three  different  Jme  points:  

0,  72  and  96  hours.  

     

RESULTS

Figure  1   Figure  2  

Figure  1  and  3:    Each  point  in  the  graphs  in  Figure  1  and  2  is  a  LD  for  a  single  catheter  in  a  single  experiment.  The  lines  indicate  how  the  mean  LDs  for  each   catheter  type  change  over  Wme  in  each  experiment.  

Table  2   Table  1  

(3)

RESULTS

Introcan 3 ViaValve

BD  BC  

B.Braun  Introcan  Safety®  3  Catheter   BD  Insyte™  Autoguard™  BC  Shielded  I.V.  Catheter  

A A

The  log  sum  CFU  for  biofilm  formed  on  all  internal  fluid  pathway  surfaces  within  Smiths  Medical  ViaValve™  Safety  I.V.  catheter  was  3.84,  BD  Insyte™  Autoguard™  BC  Shielded  I.V.  Catheter  was  4.02,    B.Braun  Introcan  Safety®  3  catheter  was  4.90  and  the  Surflash®Plus  Safety  IV  catheter  was  5.43.    Biofilm  was  observed  on  SEM  on  internal  component  surfaces  with  higher  CFU  counts  

ViaValve™  Safety  I.V.  Catheter  

               Figure  3.  Cross-­‐secWonal  schemaWc  of  the  catheter  hub    accessed  with  syringe  

E                                        C              B                                                    A  

D  

E D C B A  Surflash®  Plus  Safety  I.V.  Catheter    

               Figure  4.  DestrucWve  sampling  of  catheter  at  96  hours.  Each  part  was  placed  in  a    

H                G            F          E      D  C              B                    A  

               Table  3.  DestrucWve  sampling  CFU  counts  for  each  design  component.    The  log  sum  CFU  increased  as  the  internal  surface  area  increased.      

B.Braun  Introcan  Safety®  3  Catheter   BD  Insyte™  Autoguard™  BC  Shielded  I.V.  Catheter  

ViaValve™  Safety  I.V.  Catheter  

 Surflash®  Plus  Safety  I.V.  Catheter    

(4)

METHODS RESULTS

The  hypothesis  was  that  catheters  with  larger  SAs  and  volumes  would   also  have  larger  bacterial  densiJes  in  Flush  1  and  Flush  2.  On  the   average,  the  staJsJcs  support  this  hypothesis.    

 

At  each  Jme  point  separately,  there  were  significant  increases  in  the   Flush  1  and  Flush  2  LDs  as  either  SA  or  Volume  increased  (p  ≤  0.0001).  

Across  the  Jme  points,  the  effect  of  SA  differed  significantly  (p  ≤   0.0150);  and  the  effect  of  volume  also  differed  significantly  (p  ≤   0.0048).  Based  on  R2  and  informaJon  criteria,  neither  SA  nor  Volume   emerged  as  being  be^er  than  the  other  when  predicJng  Flush  1  and   Flush  2  results.  In  other  words,  both  SA  and  volume  are  important.  

InteresJngly,  for  both  Flush  1  and  Flush  2,  the  effect  of  SA  was   significantly  different  for  different  volumes  (or,  equivalently,  the   effect  of  volume  was  significantly  different  for  different  SAs  (p  ≤   0.040).    

 

The  predicJve  power  of  the  models  fit  to  the  Flush  1  and  Flush  2  LDs   as  a  funcJon  of  both  SA  and  volume  and  Jme  can  be  roughly   assessed  by  observing  that  the  R2  ranged  from  60  -­‐  63%  for  the  Flush   1  and  Flush  2  data  respecJvely.  In  other  words,  60  -­‐  63%  of  the   variability  in  the  Flush  1  and  Flush  2  LDs  are  explained  by  the  model,   with  37-­‐40%  of  the  variability  is  due  to  unknown  sources.(p-­‐

value=0.014  and  0.010  respecJvely).    

Figure  5.  Comparison  for  surface  area  (in2)  and  biofilm  count  (CFU)  

Bacterial transfer, biofilm formation and effect of device design in four peripheral IV blood-control catheters

Marcia  Ryder1,  Elinor  deLancey  Pulcini2,  Albert  Parker2,  and  Garth  James2  

in plunger

in catheter in hub

in seal in sleeve

between seal and hub in back of seal in actuator

BD  BC   in  &  around  

plunger   in  seal  

in  hub   in  eyelet  

in  tube  (enJre  length   not  shown)   ViaValve

 Figure  6.  Internal  volume  and  surface  area  in  contact  with  blood  (in  red).  Blood  locaWons  used  to  calculate  surface  area  and  volume    in  Table  4.      The  cross-­‐secWons,  surface  area  (SA)  and  volume  for  each  catheter  was  provided  by  Smith  Medical    engineers.  .      

 

Three  different  linear  mixed  effect  models  were  fit  to  the  Flush  1  LDs,  and   separately  to  the  Flush  2  LDs.    All  of  these  models  included  a  random  effect  for   experiment,  and  a  fixed  effect  for  Jme  (with  3  levels:  0,  72  and  96  hours).    The   first  model  included  a  covariate  for  SA  and  the  2-­‐way  interacJon  between  SA   and  Jme.    The  second  model  included  a  covariate  for  volume  and  the  2-­‐way   interacJon  between  volume  and  Jme.    The  third  model  included  covariates   for  both  SA  and  volume,  their  2-­‐way  interacJon,  and  the  3-­‐way  interacJon   between  SA,  volume  and  Jme.    Random  slopes  for  SA  and  Volume  separately   were  also  considered,  but  since  it  was  not  possible  to  fit  random  slopes  to  

“model  3”  described  above  (that  contains  both  SA  and  Volume),  then  this   report  does  not  consider  any  model  with  random  slopes.  

   

Residual  and  normal  probability  plots  were  used  to  check  for  outliers,  and  to   check  the  normality  and  homogeneity  of  variance  assumpJons  of  the  models   used.  

    RESULTS

Given  the  observaJon  of  increased  biofilm  formaJon  in  relaJon   to  surface  area  and  volume,  an  analysis  was  done  to    determine    if  the  added  components  within  the  catheter  hub  with  larger     surface  area  and  volume  would  also  have  larger  bacterial  transfer     when  flushed.    

Table  4.    Comparison  of  the  internal  volume  and  surface  area  of   the  Autoguard  BC,  Introcan  Safety®3,  and  Surflash    Safety   catheter  hubs  to  the  ViaValve  

           Internal  Volume  and  Surface  Area  in  Contact  with  Blood  

Internal  surface  area  (in2)  0.34                          0.96                          2.8  x  more  SA                                                                                                                                                                                              

ViaValve™          Introcan®3                Difference                                    

Internal  volume  (in3)              0.0038                      0.017                      4.5  x  more  vol   Internal  surface  area  (in2)  0.34                          0.73                          2.1  x  more  SA     Internal  volume  (in3)              0.0038                      0.0082                  2.2  x  more  vol   ViaValve™  Autoguard  BC™        Difference    

ViaValve™    Sureflash  Plus®      Difference      

Internal  volume  (in3)              0.0038                      0.0063                  1.7  x  more  vol     Internal  surface  area  (in2)  0.34                          0.66                          1.9  x  more  SA                            

                                                                                                                                                                                   

in hub in eyelet

Introcan 3

Table  4  .    Surface  area  (in2)  and  biofilm  count  (CFU)  

ViaValve™  Safety  I.V.  Catheter  

BD  Insyte™  Autoguard™  BC  Shielded  I.V.  Catheter  

B.Braun  Introcan  Safety®  3  catheter  

 Surflash®  Plus  Safety  I.V.  catheter    

 Surflash®  Plus  Safety  I.V.  catheter    

(5)

Figure  3.    Scanning  electron  microscope  image  of  the  white   arm  within  the  flow  path.    The  large  aggregates  of  spherical   objects  Indicate  biofilm  formaWon  by  Staph  aureus.  

CONCLUSIONS DISCUSSION

This  project  was  funded  by  Smiths  Medical  Inc.  

This  informaJon  was  provided  under  a  Montana  State  University  TesJng  Services   Agreement  and  Ryder  Science,  Inc.  and  is  not  intended  to  endorse  or  recommend  any   product  or  service                                                                                                                                                                                                                    February  2015  

As  shown  in  the  illustraJons  and  Table  1,  the,  BD   Insyte™  Autoguard™  BC  Shielded  I.V.  Catheter  and   the  B.Braun  Introcan  Safety®  3  catheters  have   more  complex  flow  paths  than  the  ViaValve™  

Safety  I.V.  catheter,  as  well  as  higher  internal   surface  areas.  Flow  path  irregulariJes,  in   parJcular  areas  that  receive  minimal  fluid  flow,   are  areas  that  promote  bacterial  a^achment  and   biofilm  formaJon.  This  is  analogous  to  dead-­‐legs   and  surface  irregulariJes  in  high  purity  water   systems,  which  serve  as  reservoirs  for  biofilm   accumulaJon  and  contaminaJon  of  the  system.  

Figure  5.  This  three-­‐dimensional  reconstrucWon  of  confocal   scanning  laser  microscope  images  shows  biofilm  growth  on   a  catheter  septum.  The  biofilm  was  stained  with  the  LIVE/

DEAD®  BacLight™  Bacterial  Viability  Kit  (Life  Technologies   CorporaWon,  Carlsbad,  CA).  Live  bacteria  appear  green.  

Figure  4.    Scanning  electron  microscopic  image  of  the   intraluminal  surface  of  a  sleeve  within  the  catheter  hub.      

The  large  aggregates  of  spherical  objects  Indicate  biofilm   formaWon  by  Staph  aureus.  

 

RESULTS The  log  sum  CFU  for  biofilm  formed  on  all  internal  fluid  

pathway  surfaces  within  Smiths  Medical  ViaValve™  Safety  I.V.  

catheter  was  3.84,  BD  Insyte™  Autoguard™  BC  Shielded  I.V.  

Catheter  was  4.02,  B.Braun  Introcan  Safety®  3  catheter  was   4.90  Surflash®Plus  Safety  IV  catheter  was  5.43.  

 

Biofilm  was  observed  on  SEM  on  internal  component  surfaces   with  higher  CFU  counts  

Figure  4.    Scanning  electron  microscopic  image  of  the   intraluminal  surface  of  a  catheter  tubing.    The  small     groups  of  spherical  objects  are  S.  aureus  cells.    

Figure  4.    Scanning  electron  microscopic  image  showing   Staph  aureus  biofilm  on  the  inner  surface  of  a  sleeve  within   a  catheter  hub.  

Figure  2.    Scanning  electron  microscopic  image  of  the  inside  of  a   catheter  hub.    The  small  groups  of  spherical  objects  are  S.  aureus   cells.    

There  are  differences  in  biofilm  formaJon  among   the  devices  with  higher  internal  surface  areas  and   volume.    These  differences  increase  the  potenJal   risk  for  transfer  of  bacteria  into  the  bloodstream   among  the  different  blood  control  valved  PIVC   designs.  

 

StaJsJcal  analyses  that  considered  the  surface   area  and  volume  of  each  catheter  indicated  that   there  were  significantly  greater  mean  log  bacterial   densiJes  in  Flush  1  and  Flush  2  as  either  surface   area  or  volume  was  increased.    

 

ViaValve™  Safety  I.V.  catheters  had  staJsJcally   significantly  fewer  bacteria  in  flush  counts   compared  to  both  Autoguard™  BC,  Introcan®  

Safety  3,  and  the  Surflash  Plus®  catheters  .  It  also   had  fewer  bacteria  on  internal  surfaces  as  well  as  a   smaller  internal  surface  area  and  less  complex  fluid   path.  These  differences  may  reduce  the  potenJal   risk  for  bacterial  transfer  and  bloodstream   infecJon.  

   

Referensi

Dokumen terkait