Sibbitt eAppendix Page 1 of 4

Fig. E-1

Photographs showing a mechanical patient-safety syringe (Avanca RPD) fitted with a safety needle with a rotating sheath (Eclipse). A, The components of the mechanical syringe including an injection plunger, an aspiration plunger, a pulley system connecting the two plungers so that the plungers reciprocate (alternate), a functional barrel for the injection plunger, and an accessory barrel for the aspiration plunger. Fluid is only contained within the functional barrel with the needle fitting. B, To aspirate, the thumb depresses the aspiration plunger. C, To inject, the thumb depresses the injection plunger.

(2)

Sibbitt eAppendix Page 2 of 4

Fig. E-2

Photographs showing the Procedur-SF safety syringe, which is a digital retractable mechanical syringe that reciprocates with two plungers, one an aspiration plunger and the other an

injection plunger. A, To inactivate the needle, the injection plunger is fully depressed, with the thumb locking the plunger into the needle fitting. B and C, The aspiration plunger then is

depressed with the thumb (B), which retracts the needle within the syringe barrel, shielding the needle (C).

(3)

Sibbitt eAppendix Page 3 of 4

TABLE E-1 Safety-Engineered Syringes and Needles

Safety Device Manufacturer

Type of Anti- Needlestick Device

FDA Cleared

Health-care Worker

Safety

Patient Safety

Changeable Needle

Type of Changeable Needle

Can Use Specialty Needles

Can Use 2- in (5.1-cm)

Needles

Integra Becton

Dickinson (BD)

Retractable syringe, spring driven

Yes Yes No Yes Proprietary No No

Procedur-SF Avanca Medical

Devices.

Retractable syringe, digitally driven

Yes Yes Yes Yes Conventional

needles

Yes Yes

RPD Safety Syringe Avanca Medical Devices

Safety syringe with shielded needle, digitally driven

Yes Yes Yes Yes Any Luer Fitting

Safety Needle

Yes Yes

Baksnap Duopross

Meditech

Retractable syringe, manually driven

Yes Yes No Yes Conventional

needles

Yes Yes

Safepro Invirosnap Inviro Retractable syringe, manually driven

Yes Yes No Yes Conventional

needles

Yes Yes

VanishPoint Retractable

Technologies

Retractable syringe, spring driven

Yes Yes No No Fixed needle only No No

Eclipse Safety Needle Becton Dickinson

Retractable syringe, spring driven

Yes Yes No Yes Proprietary No No

SafetyGlide Safety Needle Becton Dickinson

Shielded needle, digitally driven

Yes Yes No Yes Proprietary No No

SurGuard Safety Needle Terumo Shielded needle, digitally driven

Yes Yes No Yes Proprietary No No

Monoject Magellan Safety Needle

Kendall (Covidien)

Shielded needle, digitally driven

Yes Yes No Yes Proprietary No No

Safety-Lok Safety Syringe Becton Dickinson

Syringe with safety shield, manually driven

Yes Yes No Yes Conventional

needles

Yes Yes

Monoject Safety Syringe Kendall (Covidien)

Syringe with safety shield, manually driven

Yes Yes No Yes Conventional

needles

Yes Yes

Digitally Activated Shielded (DAS) Syringe

Science &

Technology Corporation @ UNM

Syringe with safety shield, digitally driven

In process

Yes No Yes Conventional

needles

Yes Yes

(4)

Sibbitt eAppendix Page 4 of 4

TABLE E-2 Clinical Characteristics of Safety-Injection Devices

Safety Device

Requires Two Hands to Inactivate

Must Change Holding

Finger Position to

Inactivate

Can Be Prematurely Activated While Needle in Patient

Can Use the One-Needle Two-Syringe

Technique

Splatter of Contaminant When Activated

in Tissues

Splatter of Contaminant When Activated

Outside Body

Safety Feature Can Be Reliably

Activated with Needle in Joint

Capsule

Cost per 100 Devices (US

dollars)

Integra No No Yes Yes Minimal Minimal No (spring too

weak)

$130

Procedur-SF No No Yes Yes Minimal Moderate Yes $150

RPD Safety Syringe No Yes No Yes Minimal Minimal No $150

Baksnap Yes Yes Yes Yes Minimal Moderate Yes $34

Safepro Invirosnap Yes Yes Yes Yes Minimal Moderate Yes $50

VanishPoint No No Yes No Minimal Moderate No (spring too

weak)

$135

Eclipse Safety Needle No Yes No Yes Cannot activate

in tissue

Minimal No $40

SafetyGlide Safety Needle

No Yes No Yes Minimal Minimal Yes (but

difficult)

$40

SurGuard Safety Needle No Yes No Yes Cannot activate

in tissue

Minimal No $30

Monoject Magellan Safety Needle

No Yes No Yes Minimal Minimal Yes (but

difficult)

$40 Safety-Lok Safety

Syringe

Yes Yes No Yes Minimal Minimal Yes $40

Monoject Safety Syringe Yes Yes No Yes Minimal Minimal Yes $40

Digitally Activated Shielded (DAS) Syringe

No No No Yes Minimal Minimal Yes $40

Referensi

Dokumen terkait

[r]

₪ Apakah sistem pembayaran SPP yang baru akan dapat diterima.

E-1 Average zone of bacterial growth inhibition left and antimicrobial concentration right for atmospheric Atm; solid circles and vacuum-mixed Vac; open circles Cemex Genta.. The

OUTCOMES AFTER TREATMENT OF UNSTABLE ANKLE FRACTURES WITH METALLIC VERSUS BIODEGRADABLE IMPLANTS http://dx.doi.org/10.2106/JBJS.K.01221 Page 2 of 3 TABLE E-1 Radiographic Criteria

†1 = bedridden or confined to wheelchair, 2 = sedentary minimum capacity for walking or other activity, 3 = semi-sedentary white-collar job, bench work, light housekeeping, 4 = light

Patients undergoing revision arthroplasty were stratified first according to the type of stem cemented [CPT; Zimmer, Warsaw, Indiana] or uncemented [ZMR; Zimmer] and then by whether

E-3 The anterior center-edge angle ACEA was measured on false-profile radiographs by measuring the angle that subtended a vertical plumb line through the center of the head and a line

Quatman eAppendix Page 2 of 2 treatment, and osteochondral lesions resulting from osteonecrosis related to chemotherapy or steroid use • Studies that described synthesized results