Use of Freestyle Libre Pro TM Flash Glucose Monitoring System in Different Clinical Situations at a Diabetes Centre

Bangalore Srinivasan Ratna

1

, Radhakrishnan Subashini

2

, Ranjit Unnikrishnan

3

, Viswanathan Mohan

4

of flash glucose monitoring. It combines inputs from multiple days of glucose data and collapses them into a 24-hour modal day period. It allows quick analysis of poor glycemic control and glycemic variability and can be used as a patient education tool. It consists of three parts: a disposable sensor, a hand-held reader and a proprietary software. The sensor is placed on the back of upper arm with the help of an applicator (Figure 1). The sensor has a small, flexible 4mm-long tip inserted just under the skin;

it can be worn up to 14 days. Its amperometric electrochemical sensor measures glucose readings in the 40 to 500 mg/dl range every 15 minutes during the entire wear period and stores them. Using near field communication technology, the sensor data is downloaded into the reader. While the sensor is

1Fellow, Dr. Mohan’s Diabetes Specialities Centre, Chennai, Tamil Nadu; 2Senior Biostatistician, Madras Diabetes Research Foundation, Chennai, Tamil Nadu; 3 Vice Chairman, Dr. Mohan’s Diabetes Specialities Centre and Madras Diabetes Research Foundation, Chennai, Tamil Nadu; 4President, Madras Diabetes Research Foundation and Chairman, Dr. Mohan’s Diabetes Specialities Centre, Chennai, Tamil Nadu

Received: 20.04.2016; Revised; 22.03.2017; Accepted: 30.03.2017

O R I G I N A L A R T I C L E

Abstract

Flash Glucose Monitoring (FGM) is a newly introduced Glucose Monitoring system. FGM provides both graphical representation of qualitative and quantitative changes in glucose levels occurring over 14-day period collapsed into single modal day graph as well as day by day graphs.

In this report, we present our experience with FGM in different clinical situations. In our experience FGM is an extremely useful clinical tool for detection of glucose variability, hyperglycemia and hypoglycemia. It can also be used as an effective educational tool when the data from FGM is shared with the patient. This helps in motivating the patient in achieving better glycemic control.

Editorial Viewpoint

• A m b u l a t o r y g l u c o s e monitoring is a novel method for monitoring.

• It can be used to adjust drug dose.

• It can be useful to motivate the patient to achieve good glycemic control.

Introduction

A

dequate control of blood glucose levels largely depends upon frequent monitoring of control both at the clinic as well as at home by the patient. Self monitoring of blood glucose (SMBG), introduced in the 1980s, is the most widely used method of all the glucose monitoring techniques. It is most useful in patients on multiple insulin doses or on continuous subcutaneous insulin pump. It may also be helpful for treatment or self management for patients using less frequent insulin injections or on non-insulin therapies.1 SMBG however only gives point of test blood glucose measurements without predicting the direction or rate of change of blood glucose or a continuous monitoring. Moreover, the need for frequent painful finger pricks, issues with calibration and accuracy of the glucose meter as well as the cost of strips contribute

to poor patient compliance. As the concept of glycemic variability and its importance evolved, techniques to measure glycemic variability h a v e s i m u l t a n e o u s l y b e e n developed.2,3 An example is the Continuous Glucose Monitoring System (CGMS) that has been in use for several years.4 We have earlier reported on our initial experience with CGMS at our centre.5 The limitations of the currently available CGMS devices include the need for multiple finger stick calibrations, pain at sensor insertion site, expense and the relatively short life of the sensor (up to 7 days only).6

Flash Glucose Monitoring (FGM) is a newly introduced method

(2)

Fig. 1 : FGM showing sensor and reader

Fig. 2A: FGM in a normal  person

Fig. 2B: MAGE on day 2 (Case 1)

0 20 40 60 80 100 120 140 160

00:14 00:59 01:44 02:29 03:14 03:59 04:44 05:29 06:14 06:59 07:44 08:29 09:14 09:59 10:44 11:29 12:14 12:59 13:44 14:29 15:14 15:59 16:44 17:29 18:14 18:59 19:44 20:29 21:14 21:59 22:44 23:29 Series1 109

76

121

91 151

87 111

85 132141

97 122

105

Day 2

Glucose(mg/dl)

Hours

86 MAGE – 37 mg/dl

MAGE Σnλ if λ > ν

λ = each blood glucose increase or decrease (nadir-peak or peak nadir) n = number of observations ν = 1 SD of mean glucose for 24-hr period

Fig. 3A: Type 2 diabetes of short duration

Fig. 3B: MAGE on day 2 (Case 2)

Day 2 300

250 200 150 100 50 0

229

77

239

85 255

Glucose mg/dl 112

MAGE - 150 mg/dl

Hours

0:13 1:13 2:13 3:13 4:13 5:13 6:13 7:13 8:13 9:13 10:13 11:13 12:13 13:13 14:13 15:13 16:13 17:13 18:13 19:13 20:13 21:13 22:13 23:13

single-use only, the reader can be used to collect data from multiple sensors. The report is generated by connecting the reader to a computer in which the software has been installed. The graphical data provided by FGM includes mean of glucose values at different point of day, a median curve representing the 50th percentile and also the 75th percentile and 25th percentile curves defining the Inter Quartile Range (IQR). The width of the band between the 10th and

90th percentile curve is an index of the glucose excursion (variability).

The glycemic trends for individual days can also be obtained in a graphical way. For the purpose of this paper, we have also manually calculated the Mean Amplitude of Glycemic Expression (MAGE) as the arithmetic mean of differences between consecutive peaks and nadirs of the glucose readings. The MAGE was calculated by taking 24-hour glucose values of Day 2 and we compared this with 24-hour glucose readings of Day 13, Days 1 and 14 were not considered, as they usually have only incomplete readings depending on time of application and removal.

The aim of this paper is to show the use of FGM in a few clinical situations. Specifically, we report on cases illustrating the use of FGM in different types of diabetes.

Case 1 (FGM in a Normal Person)

This was a 61 year old person with normal glucose tolerance in whom FGM was applied for a few days to see the normal glycemic variability. His fasting plasma glucose (FPG) was 96 mg/dl, post prandial plasma glucose (PPPG) 128 mg/dl and HbA1c, 5.6%. The MAGE on Day 2 was 37mg/dl with the lowest recorded glucose reading of 74 mg/dl and highest of 151 mg/dl during the days when he wore the FGM (Figures 2A and 2B).

Case 2 (Type 2 diabetes of short duration)  

This was a 44 year old male patient with type 2 diabetes of 3 years duration, on Tab. Glimepiride 2 mg two times a day. His FPG was 162 mg/dl, PPPG 296 mg/dl and

14 May 2015 - 18 May 2015 (5 days)

14 May 2015 - 18 May 2015

Glucose

00:00 06:00 12:00 18:00 00:00

00:00 06:00 12:00 18:00 00:00 Average Glucose

94 80% 20% 0%

mg/dL

97 84% 14% 2%

mg/dL

102 100% 0% 0%

mg/dL

107 100% 0% 0%

mg/dL

98 100% 0% 0%

mg/dL

101 94% 5% 1%

mg/dL

Time in TargetTime Below Target Time Above

Target mg/dL350

14080 0

14080 0

14080 0

14080 0

14080 0 mg/dL350

mg/dL350

mg/dL350

mg/dL350

Daily Glucose Summary

Thu14 May

Fri15 May

Sat 16 May

Sun17 May

Mon 18 May

Pro Pro

7 January 2016 - 21 January 2016 (15 days)

Daily Glucose Summary 7 January 2016 - 21 January 2016 (15 days)Daily Glucose Summary

Glucose Glucose

Thu7 Jan

Fri15 Jan

Sat 16 Jan

Sun 17 Jan

Mon 18 Jan

Tue 19 Jan

Wed20 Jan

Thu21 Jan Fri8 Jan

Sat9 Jan

Sun10 Jan

Mon11 Jan

Tue12 Jan

Wed13 Jan

Thu 14 Jan

00:00 06:00 12:00 18:00 00:00

00:00 06:00 12:00 18:00 00:00

00:00 06:00 12:00 18:00 00:00 mg/dL200

14080 0 mg/dL200

14080 0 mg/dL200 14080 0 mg/dL200

14080 0 mg/dL200

14080 0 200 mg/dL 14080 0 mg/dL200 14080 0 200 mg/dL 14080 0

Average GlucoseTime in TargetTime Below Target Time Above

Target 138 56% 0% 44%

mg/dL

165 32% 2% 66%

mg/dL

193 12% 4% 84%

mg/dL

183 24% 0% 76%

mg/dL

162 38% 2% 60%

mg/dL

144 49% 0% 51%

mg/dL

146 53% 0% 47%

mg/dL

145 49% 0% 51%

mg/dL

mg/dL200

14080 0 mg/dL200

14080 0 mg/dL200 140 80 0 mg/dL200 140 80 0 mg/dL200

14080 0 mg/dL200

14080 0 mg/dL200

14080

000:00 06:00 12:00 18:00 00:00 Average GlucoseTime in TargetTime Below

TargetTime Above Target

150 42%4% 54%

mg/dL

149 49%6% 45%

mg/dL

143 46%1% 53%

mg/dL

155 47%0% 53%

mg/dL

158 50%0% 50%

mg/dL

117 81%0% 19%

mg/dL

114 90%6% 4%

mg/dL

153 46%2% 52%

mg/dL 7 January 2016 - 21 January 2016

(3)

Fig. 4A: Long duration of type 2 diabetes Fig. 3C: MAGE on day 13 (Case 2)

0 20 40 60 80 100 120 140 160 180 200

00:13 01:13 02:13 03:13 04:13 05:13 06:13 07:13 08:13 09:13 10:13 11:13 12:13 13:13 14:13 15:13 16:13 17:13 18:13 19:13 20:13 21:13 22:13 23:13 Serie…

80

135

106 90 176

Day 13

188

MAGE – 74 mg/dl

Fig. 4B: MAGE on day 2 (Case 3)

Fig. 4C: MAGE on day 13 (Case 3)

0 50 100 150 200 250 300 350

00:00 00:45 01:30 02:15 03:00 03:45 04:30 05:15 06:00 06:45 07:30 08:15 09:00 09:45 10:30 11:15 12:00 12:45 13:30 14:15 15:00 15:45 16:30 17:15 18:00 18:45 19:30 20:15 21:00 21:45 22:30 23:15 Series1

93

307

97 187

98 Day 2

MAGE – 120 mg/dl

Hours

Glucose mg/dl

0 50 100 150 200 250 300 350

00:00 00:45 01:30 02:15 03:00 03:45 04:30 05:15 06:00 06:45 07:30 08:15 09:00 09:45 10:30 11:15 12:00 12:45 13:30 14:15 15:00 15:45 16:30 17:15 18:00 18:45 19:30 20:15 21:00 21:45 22:30 23:15 Series1

301

129 226

136 Day 13

MAGE – 158 mg/dl

Hours

Glucose mg/dl

HbA1c, 8.2%. He was started on Inj. Glargine 10 units at bedtime along with tab Glimepiride 2 mg + Metformin 500 mg two times a day in view of high blood glucose levels. His AGP (Figure 3A) shows high blood glucose levels during the first week, which improved during second week with no hypoglycemic episodes. The MAGE was 150 mg/

dl on Day 2 which improved to 74 mg/dl on Day 13 (Figure 3B and 3C). The dose of glargine was later reduced and eventually withdrawn. In this case AGP helped us to analyse the glycemic control after initiation of insulin and also helped to document improvement in glycemic variability.

Case 3 (Type 2 diabetes of long duration)

This was a 55 year old male patient with type 2 diabetes of 20 years’ duration with diabetic retinopathy, diabetic nephropathy,

systemic hypertension, ischaemic heart disease, dyslipidemia and hypothyroidism. His FPG was 227 mg/ dl, PPPG 236 mg/dl and HbA1c, 8.6%. He was on Inj Aspart 16U-8U-4U plus inj Glargine 28U at bedtime, along with Tab Gliclazide 30mg once a day and Tab Linagliptin 5 mg once a day.

FGM was initiated for achieving a better glycemic control in view of the comorbidities especially organ related complications. FGM (Figure 4A) showed uncontrolled diabetes at several points of the day during the two week period. On Day 2 (Figure 4B) MAGE was 120 mg/dl.

On Day 13 (Figure 4C) MAGE was still 158 mg/dl. Further adjustments were made in insulin doses based on sensor readings. The lab reports post treatment change were FPG 122 mg/dl, PPPG 150mg/dl and HbA1c of 7.7%. A second FGM sensor was initiated to know if the improved blood sugars were stable and within the expected range. The

second FGM report (Figure 4D) shows much better control with less glycemic variability shown by reduction in MAGE to 83 mg/dl on Day 2 (Figure 4E) to 86 mg/dl on Day 13 (Figure 4F). In this patient, AGP, along with adjustment of dosage of drugs enabled us to achieve better glycemic control with a reduction in HbA1c level by 0.9%.

Case 4 (Gestational Diabetes Mellitus)

This was a 26 year primigravida diagnosed with gestational diabetes mellitus (GDM) in the second trimester of pregnancy. Her FPG was 125 mg/dl, PPPG-229 mg/dl and HbA1c, 5.8 mg/dl. She was started on Inj Biphasic insulin Aspart 8U-0-6U. FGM was initiated to monitor glycemic control during pregnancy. FGM showed (Figure 5A) high postprandial blood glucose during the day time, with few low blood glucose levels early morning during initial first week

Glucose

Glucose 169 14% 0% 86%

mg/dL

143 68% 0% 32%

mg/dL

152 41% 0% 59%

mg/dL

141 40% 2% 58%

mg/dL

141 49% 5% 46%

mg/dL

157 22% 1% 77%

mg/dL

171 18% 0% 82%

mg/dL

159 26% 0% 74%

mg/dL

Average GlucoseTime in TargetTime Below Target Time Above

Target

149 42% 0% 58%

mg/dL

168 23% 0% 77%

mg/dL

203 19% 1% 80%

mg/dL

207 28% 0% 72%

mg/dL

171 30% 3% 67%

mg/dL

111 100% 0% 0%

mg/dL

163 34% 1% 65%

mg/dL 7 January 2016 - 21 January 2016 Tue26 May

Wed27 May

Thu 28 May

Fri 29 May

Sat30 May

Sun 31 May

Mon 1 Jun

Tue 2 Jun

00:00 06:00 12:00 18:00 00:00 mg/dL350

14080 0 mg/dL350 140 80 0 350 mg/dL

14080 0 mg/dL350

14080 0 350 mg/dL

14080 0 350 mg/dL

14080 0 mg/dL350

14080 0 mg/dL350

14080 0

AverageGlucoseTime in TargetTime Below Target Time Above Target

Wed 3 Jun

Thu 4 Jun

Fri 5 Jun

Sat 6 Jun

Sun 7 Jun

Mon8 Jun

00:00 06:00 12:00 18:00 00:00 350

mg/dL

140 80 0 350 mg/dL

140 80 0 mg/dL350

140 80 0 mg/dL350

140 80 0 mg/dL350

140 80 0 mg/dL350

140 80

000:00 06:00 12:00 18:00 00:00 00:00 06:00 12:00 18:00 00:00

(4)

Fig. 4D: Long duration of type 2 diabetes: A second FGM sensor

Fig. 4E: MAGE on day 2 while wearing second sensor (Case 3)

0 20 40 60 80 100 120 140 160 180

00:00 00:45 01:30 02:15 03:00 03:45 04:30 05:15 06:00 06:45 07:30 08:15 09:00 09:45 10:30 11:15 12:00 12:45 13:30 14:15 15:00 15:45 16:30 17:15 18:00 18:45 19:30 20:15 21:00 21:45 22:30 23:15 Series1 74

169 162

55 111

Hours

MAGE – 83 mg/dl Day 2

Glucose mg/dl

Fig. 4F: MAGE on day 13 while wearing second sensor (Case 3)

Fig. 5A: Gestational diabetes mellitus

Hours Day 13

0 50 100 150 200 250

00:00 00:45 01:30 02:15 03:00 03:45 04:30 05:15 06:00 06:45 07:30 08:15 09:00 09:45 10:30 11:15 12:00 12:45 13:30 14:15 15:00 15:45 16:30 17:15 18:00 18:45 19:30 20:15 21:00 21:45 22:30 23:15 Series1 84

212

40

200

129 158

Glucose mg/dl

MAGE – 86 mg/dl

period. During the second week of sensor use, there was significant improvement in the post prandial blood glucose levels with a few hypoglycaemic episodes persisting during early morning hours. The MAGE of 108 mg/dl on Day 2 (Figure 5B), decreased to 59 mg/dl on Day 13 (Figure 5C). The morning dose of insulin was increased to 10 U for further control of postprandial spikes, with reduction in the night dose in view of early morning hypoglycemia. Thus FGM enabled us to achieve strict glycemic control in pregnancy and detect the asymptomatic hypoglycemia which helped in insulin dose adjustment.

Case 5 (Type 1 Diabetes)

This was a 10 year old boy with Type 1 DM of 5 months duration.

His blood glucose values were FPG 162 mg/dl, PPPG 174 mg/dl and HbA1c of 9.7%. He was on Inj Aspart 3U-3U-3U along with Inj

Glargine 8U at bedtime. FGM was initiated to see why his HbA1c was so high, in spite of relatively acceptable fasting and postprandial glucose levels. His FGM (Figure 6A) showed high blood sugar levels at several points during the day with occasional hypoglycemic episodes.

The MAGE on Day 2 was 88 mg/

dl (Figure 6B) and 284 mg/dl on Day 13 (Figure 6C). This type of profile is characteristic of type 1 diabetes. The dose of insulin was adjusted to Inj Aspart 3U-2U-2U and Inj. Glargine 10 U. In this case, the FGM enabled us to detect wide fluctuations in the blood glucose levels and to improve the diabetes control by adjusting the insulin doses.

Case 6 (Fibrocalculous Pancreatic Diabetes)

This was a 55 year patient with fibrocalculous pancreatic diabetes of 18 years’ duration. His current

medications were Inj Biphasic Isophane insulin 22U-0-8U. His FPG was 99 mg/dl, PPPG 252 m g / d l a n d H b A 1 c , 7 % . F G M was initiated to study the blood sugar fluctuations. FGM showed high blood glucose levels during postprandial periods, with a few hypoglycemic episodes during early morning hours (Figure 7A).

Even though HbA1c was 7%, AGP showed high glycemic variability with MAGE of 120 mg/dl on Day 2 (Figure 7B) and 93 mg/dl on Day 13 (Figure 7C). In this case, the FGM enabled us to visualize the day to day variability of blood glucose even when the patient’s HbA1c was apparently under control and helped us to make adjustments in the dosage of insulin.

Discussion

In this paper we have presented 5 different clinical types of diabetes in which FGM was initiated and also 1 normal individual. The first case shows a normal glucose tolerant subject where there is

Glucose

00:00 06:00 12:00 18:00 00:00

Glucose

00:00 06:00 12:00 18:00 00:00

Wed26 Aug Thu

3 Sep

Fri4 Sep

Sat 5 Sep

Sun 6 Sep

Mon 7 Sep

Tue 8 Sep Thu27 Aug

Fri28 Aug

Sat 29 Aug

Sun30 Aug

Mon31 Aug

Tue1 Sep

Wed2 Sep 350 mg/dL 180 70 0

350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350

mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70

000:00 06:00 12:00 18:00 00:00 Average GlucoseTime in TargetTime Below Target Time Above Target 127 100% 0% 0%

mg/dL

102 94% 6% 0%

mg/dL

106 67% 31% 2%

mg/dL

144 68% 0% 32%

mg/dL

128 94% 4% 2%

mg/dL

113 89% 9% 2%

mg/dL

131 83% 7% 10%

mg/dL

127 75% 7% 18%

mg/dL

00:00 06:00 12:00 18:00 00:00 26 August 2015 - 8 September 2015

Average GlucoseTime in TargetTime Below

TargetTime Above Target

153 64% 3% 33%

mg/dL

138 76% 5% 19%

mg/dL

133 92% 6% 2%

mg/dL

146 81% 0% 19%

mg/dL

141 83% 4% 13%

mg/dL

103 72% 24% 4%

mg/dL

128 80% 8% 12%

mg/dL

Glucose

00:00 06:00 12:00 18:00 00:00

00:00 06:00 12:00 18:00 00:00 Thu

6 Aug Fri

14 Aug

Sat 15 Aug

Sun 16 Aug

Mon17 Aug

Tue 18 Aug

Wed19 Aug

Thu 20 Aug Fri7 Aug

Sat 8 Aug

Sun 9 Aug

Mon 10 Aug

Tue 11 Aug

Wed 12 Aug

Thu 13 Aug

350 mg/dL 140 80 0 350 mg/dL 140 80 0 350 mg/dL 140 80 0 350 mg/dL 140 80 0 350 mg/dL 140 80 0 350 mg/dL 140 80 0 mg/dL350 140 80 0 350 mg/dL 140 80 0

Glucose

00:00 06:00 12:00 18:00 00:00

00:00 06:00 12:00 18:00 00:00 mg/dL350

140 80 0 350 mg/dL 140 80 0 350 mg/dL 140 80 0 mg/dL350 140 80 0 350 mg/dL 140 80 0 350 mg/dL 140 80 0 350 mg/dL 140 80 0 Average

GlucoseTime in TargetTime Below Target

Time Above Target

101 38%38%24%

mg/dL

99 47%35%18%

110 57%19%24%

mg/dL

mg/dL

112 59%16%25%

mg/dL

128 65% 1% 34%

mg/dL

125 69% 1% 30%

mg/dL

119 73% 0% 27%

mg/dL

115 57%18%25%

mg/dL 107 55%25%20%

mg/dL

49 6% 94%0%

mg/dL

89 51%46%3%

mg/dL

75 46%53%1%

mg/dL

84 32%60%8%

mg/dL

92 43%48%9%

mg/dL

116 74% 5%21%

mg/dL

133 60% 0%40%

mg/dL Average GlucoseTime in TargetTime Below

TargetTime Above Target

(5)

but also by the oxidative stress generated by glycemic variability.12 The FGM provides an easy visual analysis of presence of glycemic variability. Glycemic variability is said to present when glucose values are widely spread, for example, when the IQR and 10th and 90th percentile curves cover a wider area. Asymptomatic hypoglycemia can be also detected.

In this report, we calculated the intraday glycemic variability by manually calculating the MAGE.13 By comparing MAGE on Day 2 with that on Day 13 we could analyse the effect of the treatment prescribed over the 2 week duration and make adjustments to the prescription given at the clinic 2 weeks earlier.

In our experience, FGM helps us to easily visualize whether patient blood glucose level is within the target range. When FGM report is communicated to the patient, they can understand the glucose levels better and become more involved in diabetes management.

This ultimately helps to reduce the practically very little fluctuation

of glucose levels. The second case shows how, in a recent onset T2DM patient, after starting treatment, the MAGE dramatically decreased and the control of diabetes improved.

Till such technology became available, physicians would only get a snapshot of the improvement based on occasional SMBG’s that are done. Case 3 illustrates greater fluctuations in a T2DM patient with longer duration of diabetes. Case 5 shows the marked fluctuations in a case of type 1 diabetes illustrating the ‘brittle diabetes’ in this patient.

In contrast, Case 4, a patient with GDM shows how mild the diabetes is, and how quickly GDM responds to treatment. The case with FCPD demonstrates more fluctuations in glucose levels like the type 1 patient.

The FGM is a very useful tool to learn about the glycemic status of the patient over a 14 day period after initiating or altering a patient’s antidiabetic drug regimen. As it provides day to day information about blood glucose levels for 2 weeks after a clinic visit, this is

most useful to adjust dosage of drugs. It also helps us to educate and motivate the patients to achieve good glycemic control.

Mazze et al7 first introduced U n i v e r s a l S o f t w a r e r e p o r t (Ambulatory Glucose Profile for systematic presentation of SMBG data) and this was further developed by the same group.8 An expert panel of diabetes specialists has given recommendations for standardizing glucose reporting and analysis of FGM to optimize c l i n i c a l d e c i s i o n m a k i n g .9 A study by European diabetologists supports the use of FGM for glucose data analysis and to take treatment decisions.10 FGM is considered as a valuable tool in the case of patients with type 1 diabetes and type 2 diabetes who are on insulin to identify poor glycemic control, to detect hypoglycemia and to study glycemic variability.11

It is now believed that the risk of complications of diabetes is not solely determined by exposure to sustained chronic hyperglycemia Fig. 5B: MAGE on day 2 (Case 4)

Fig. 5C: MAGE on day 13 (Case 4)

0 20 40 60 80 100 120 140 160 180 200

00:08 00:53 01:38 02:23 03:08 03:53 04:38 05:23 06:08 06:53 07:38 08:23 09:08 09:53 10:38 11:23 12:08 12:53 13:38 14:23 15:08 15:53 16:38 17:23 18:08 18:53 19:38 20:23 21:08 21:53 22:38 23:23 Series1

46

180

40 173

76 Day 2

139

183

93

MAGE – 108 mg/dl

Hours

Glucose mg/dl

Day 13

45 77

56

144 102

138 Figure 5C :

MAGE – 59 mg/dl

Hours

Glucose mg/dl

Fig. 6A: Type 1 diabetes

Fig. 6B: MAGE on day 2 (Case 5)

Day 2

177

111 71 Figure 6B

MAGE – 88 mg/dl

Hours

Glucose mg/dl

Glucose

Sat 29 Aug

Sun30 Aug

Mon 31 Aug

Tue 1 Sep

Wed2 Sep

Thu3 Sep

Fri4 Sep

Sat5 Sep

00:00 06:00 12:00 18:00 00:00

00:00 06:00 12:00 18:00 00:00

00:00 06:00 12:00 18:00 00:00 00:00 06:00 12:00 18:00 00:00

350 mg/dL 180 70 0 350 mg/dL 180 70 0

350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0

Average GlucoseTime in TargetTime Below

Target Time Above Target

Average GlucoseTime in TargetTime Below TargetTime Above Target

144 100% 0% 0%

285 8% 0% 92%

mg/dL

157 66% 2% 32%

mg/dL

204 37% 0% 63%

mg/dL

222 28% 0% 72%

mg/dL

249 15% 0% 85%

mg/dL

212 19% 0% 81%

mg/dL

248 12% 0% 88%

mg/dL

204 36% 0% 64%

mg/dL

Sun 6 Sep

Mon 7 Sep

Tue8 Sep

Wed 9 Sep

Thu 10 Sep

Fri11 Sep

Sat12 Sep

29 August 2015 - 12 September 2015 350

mg/dL 180 70 0 mg/dL350 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 350 mg/dL 180 70 0 Glucose

mg/dL

204 42% 0% 58%

mg/dL

206 27% 0% 73%

mg/dL

212 46% 0% 54%

mg/dL

198 33% 2% 65%

mg/dL

215 43% 0% 57%

mg/dL

188 43% 2% 55%

mg/dL

213 34% 0% 66%

mg/dL

(6)

9. Bergenstal RM, Ahmann AJ, Bailey T, et al. Recommendations for standardizing glucose reporting and analysis to optimize clinical decision making in diabetes: the Ambulatory Glucose Profile (AGP). Diabetes Technol Ther 2013; 15:198-211.

10. Matthaei S. Assessing the value of the Ambulatory Glucose Profile in clinical practice. Br J Diabetes Vasc Dis 2014; 14:148- 152.

11. Matthaei S, DeAlaiz RA, Bosi E, et al.

Consensus recommendations for the use of Ambulatory Glucose Profile in clinical practice. Br J Diabetes Vasc Dis 2014; 14:153- 157.

12. Monnier L, Colette C. Glycemic variability:

should we and can we prevent it? Diabetes Care 2008; 31 (suppl 2):S150-4.

13. Monnier L, Colette C, Owens DR. Glycemic variability: the third component of the dysglycemia in diabetes. Is it important?

How to measure it? J Diabetes Sci Technol 2008; 2:1094-100.

fluctuations of glucose readings and to improve HbA1c values.

We therefore feel that the FGM is an excellent new tool in the physician’s armamentarium to assess response to therapy. This paper is an illustration of different scenarios where the Freestyle Libre ProTM FGM system can be used in clinical practice.

References

1. American Diabetes Association. Standards of medical care in diabetes—2015. Diabetes Care 2015; 38 (suppl 1):S1-S93.

2. Radbard D, Bailey T, Jovanovic L, et al.

Improved quality of glycemic control and reduced glycemic variability with use of continuous glucose monitoring. Diabetes Technology Ther 2009; 11:717-23.

3. Qu Y, Jacober S, Zhang Q, et al. Rate of hypoglycemia in insulin - treated Patients with Type 2 diabetes can be predicted from glycemic variability data. Diabetes Technol

Ther 2012; 14:1008-12.

4. Beck RW, Tamborlane WV, Bergenstal RM, et al. T1D Exchange Clinic Network. The T1D Exchange clinic registry. J Clin Endocrinol Metab 2012; 97:4383-9.

5. Vidya K, Sudhir R, Mohan V. Continuous glucose monitoring system - useful but expensive tool in management of diabetes.

J Assoc Physicians India 2004; 52:587-590.

6. Vaddiraju S, Burgess DJ, Tomazos I, et al. Technologies for continuous glucose monitoring: current problems and future promises. J Diabetes Sci Technol 2010;

4:1540-62.

7. Mazze RS, Lucido D, Langer O, et al.

Ambulatory glucose profile: representation of verified self-monitored blood glucose data. Diabetes Care 1987; 10:111-7.

8. Mazze R, Strock E , Wesley D, et al.

Characterizing glucose exposure for individuals with normal glucose tolerance using Continuous Glucose Monitoring and Ambulatory Glucose Profile analysis.

Diabetes Technology Ther 2008; 10:149-159.

Fig. 6C: MAGE on day 13 (Case 5)

Fig. 7A: Fibrocalculous pancreatic diabetes

Day 13

0 50 100 150 200 250 300 350 400 450

00:05 00:50 01:35 02:20 03:05 03:50 04:35 05:20 06:05 06:50 07:35 08:20 09:05 09:50 10:35 11:20 12:05 12:50 13:35 14:20 15:05 15:50 16:35 17:20 18:05 18:50 19:35 20:20 21:05 21:50 22:35 23:20 Series1

76

401

99 341

MAGE – 284 mg/dl

Hours

Glucose mg/dl

mg/dL350 14080 0 mg/dL350 140 800 350 mg/dL 140 800 mg/dL350 14080 0 mg/dL350 140 800 mg/dL350 14080 0

mg/dL350 140 80 0 Glucose

00:00 06:00 12:00 18:00 00:00

00:00 06:00 12:00 18:00 00:00 mg/dL350

14080 0 Average

Glucose

107 52% 29% 19%

mg/dL Time in TargetTime Below

Target Time Above Target

Average GlucoseTime in TargetTime Below

Target Time Above Target

127 40% 18% 42%

mg/dL

163 39% 2% 59%

mg/dL

188 28% 0% 72%

mg/dL

177 24% 3% 73%

mg/dL

186 5% 0% 95%

mg/dL

180 24% 15% 61%

mg/dL

173 39% 0% 61%

mg/dL

180 30% 0% 70%

mg/dL

159 35% 8% 57%

mg/dL 156 40% 10% 50%

mg/dL

147 46% 1% 53%

mg/dL

157 28% 16% 56%

mg/dL

149 58% 0% 42%

mg/dL

135 60% 5% 35%

mg/dL

137 30% 27% 43%

mg/dL Thu21 Jan

Fri29 Jan

Sat30 Jan

Sun 31 Jan

Mon 1 Feb

Tue2 Feb

Wed 3 Feb

Thu4 Feb

21 January 2016 - 4 February 2016 Fri

22 Jan

Sat23 Jan

Sun24 Jan

Mon25 Jan

Tue26 Jan

Wed27 Jan

Thu 28 Jan

350 mg/dL

14080 0

00:00 06:00 12:00 18:00 00:00 350

mg/dL

14080 0 350 mg/dL 14080 0 350 mg/dL

14080 0 350 mg/dL

14080 0 350 mg/dL 14080 0

00:00 06:00 12:00 18:00 00:00 350

mg/dL 140 80 0

Fig. 7B: MAGE on day 2 (Case 6)

Fig. 7C: MAGE on day 13 (Case 6)

0 50 100 150 200 250 300

00:14 00:59 01:44 02:29 03:14 03:59 04:44 05:29 06:14 06:59 07:44 08:29 09:14 09:59 10:44 11:29 12:14 12:59 13:44 14:29 15:14 15:59 16:44 17:29 18:14 18:59 19:44 20:29 21:14 21:59 22:44 23:29 Series1 183

90 213

182

87 Day 2

MAGE – 120 mg/dl

Hours

Glucose mg/dl

40

0 50 100 150 200 250 300 350

00:14 00:59 01:44 02:29 03:14 03:59 04:44 05:29 06:14 06:59 07:44 08:29 09:14 09:59 10:44 11:29 12:14 12:59 13:44 14:29 15:14 15:59 16:44 17:29 18:14 18:59 19:44 20:29 21:14 21:59 22:44 23:29 Series1

165 Day 13

302

141 236

165 235 162

93

MAGE – 93 mg/dl

Hours

Glucose mg/dl