Towards remote monitoring of Parkinson ’ s disease tremor using wearable motion capture systems

Mehdi Delrobaei

a,*

, Sara Memar

b,1

, Marcus Pieterman

b,1

, Tyler W. Stratton

c,d

, Kenneth McIsaac

e

, Mandar Jog

b,f,2

aCenter for Research and Technology (CREATECH), Faculty of Electrical Engineering, K. N. Toosi University of Technology, Tehran, Iran

b Lawson Health Research Institute, London, ON, Canada

c Laboratory of Medicine and Pathobiology, University of Toronto, Toronto, ON, Canada

d St. Michael’s Hospital, Toronto, ON, Canada

eDepartment of Electrical and Computer Engineering, Western University, London, ON, Canada

f Department of Clinical Neurological Sciences, Western University, London, ON, Canada

A R T I C L E I N F O

Keywords:

Inertial sensing Wearable technologies Remote monitoring Whole-body tremor Parkinson’s disease

A B S T R A C T

The management of movement disorders is shifting from a centralized-clinical assessment towards remote monitoring and individualized therapy. While a variety of treatment options are available, ranging from phar- maceutical drugs to invasive neuromodulation, the clinical effects are inconsistent and often poorly measured.

For instance, the lack of remote monitoring has been a major limitation to optimize therapeutic interventions for patients with Parkinson’s Disease (PD). In this work, we focus on the assessment of full-body tremor as the most recognized PD symptom. Forty PD and twenty two healthy participants were recruited. The main assessment tool was an inertial measurement unit (IMU)-based motion capture system to quantify full-body tremor and to se- parate tremor-dominant from non-tremor-dominant PD patients as well as from healthy controls. We developed a new measure and evaluated its clinical utility by correlating the results with the Unified Parkinson’s Disease Rating Scale (UPDRS) scores as the gold standard. Significant correlation was observed between the UPDRS and the tremor severity scores for the selected tasks. The results suggest that it is feasible and clinically meaningful to utilize the suggested objective tremor score for the assessment of PD patients. Furthermore, this portable as- sessment tool could potentially be used in the home environment to monitor PD tremor and facilitate optimizing therapeutic interventions.

1. Introduction

The healthcare system is experiencing a paradigm shift from stan- dard clinical procedures to individualized assessment and treatment.

This change is accompanied by emerging technologies such as wear- ables and the Internet of Things (IoT)[1]. In this new paradigm, phy- sicians can quantify the dynamics of disease progress, predict health risks, tailor individualized therapeutic protocols, and in doing so, ef- fectively reduce costs that are burdening health care systems globally [2]. Treatment of movement disorders, especially those related to Parkinson’s disease (PD), could greatly benefit from such efficient and affordable approaches[1,3-6].

Tremor is the most recognized symptom of PD, characterized by

involuntary and approximately rhythmic shaking of a limb, head, or the entire body[7–10]. PD tremor is not life-threatening, but it often re- duces quality of life by affecting patients ability to perform certain daily living activities. Traditionally, tremor has been assessed using human observer-based PD rating scales, primarily the Unified Parkinson’s Disease Rating Scale (UPDRS)[11]. Recently, advances in technology have allowed for more quantitative measures of tremor. Various sensor systems have been developed to objectively measure PD tremor. These sensor systems include: electromyography (EMG), accelerometers, gy- roscopes, goniometers, and optical motion capture systems. However, the main drawback of these methods is that they are generally limited to the quantification of tremor severity for an individual limb, failing to provide a comprehensive, full-body assessment of tremor[7,9,12-15].

https://doi.org/10.1016/j.jns.2017.11.004

Received 16 August 2017; Received in revised form 10 October 2017; Accepted 2 November 2017

This study was supported in part by a grant from the Academic Medical Organization of Southwestern Ontario (AMOSO).

*Corresponding author.

1Equally contributed as the second author.

2The authors have no conflicts of interests.

E-mail addresses:[email protected](M. Delrobaei),[email protected](S. Memar),[email protected](M. Pieterman),[email protected](T.W. Stratton), [email protected](K. McIsaac),[email protected](M. Jog).

Available online 08 November 2017

0022-510X/ © 2017 Elsevier B.V. All rights reserved.

MARK

(2)

The vast number of tremor-measuring systems offer a wide range of assessment mechanisms. Surface EMG technique involvesfixing elec- trode sensors on the skin which measure muscular contraction [3,12,16]. Accelerometers function by combining linear acceleration data, which may include gravity and additive noise, to measure the amplitude of oscillations of individual body segments[3,13,14]. Lastly, gyroscopes are commonly used to measure angular velocity[5,14,15].

These types of sensors are often attached in close proximity to body joints in order to measure rate of change, and consequently, tremor amplitude [3]. Joint angles can be directly measured using goni- ometers, which can be placed across a body joint to provide measures of rapid movements[3,9,18]. Numerous commercially available devices (APDM[19], REMPARK[20], Kinesia[21], PERFORM[5], Motus[22], Lift Labs[5], Physilog[23]) are able to track tremor in a single joint or body part. Although the tracking is available for longer periods, it does not provide a comprehensive picture of tremor in the patient.

Recently, camera-based systems (optical 3D motion trackers) have been widely used to assess movement disorders, including tremor. To perform this assessment, multiple cameras are set up to detect markers (either active or passive) placed on body parts of participants, and compute joint movements [3,9,24,25]. However, the camera-based technology’s elaborate set-up renders the recording system stationary, restricting its use to a controlled lab environment[26]. Inertial mea- surement units (IMUs) combine accelerometers, gyroscopes, and mag- netometers (which measure the local magneticfield vector) to offer a novel approach in quantifying movement disorder symptoms. The sig- nificance of IMU sensors is that they can be serially attached to body parts to give a full-body description of body motion[27]. IMUs have allowed for the development of wearable technologies that are accu- rate, portable, easy-to-use, and can efficiently assess a variety of movement disorders[20,21].

In early stages of disease, PD tremor often begins with occasional tremor in thefingers, eventually progressing towards involvement of the entire arm and in some cases, the entire body. Therefore, full-body assessment of tremor becomes critical as the disease progresses. Such assessment aids clinicians in developing a more comprehensive treat- ment protocol. Whole-body tremor is a topic less explored in the lit- erature, compared to the local assessment of tremor. Hence, thefirst motivation of this research is to reveal the importance whole-body tremor and select the best data collection and analysis approach. This research is also motivated by the idea of remote motor symptom monitoring for follow-up of PD patients. At times, it may be un- reasonable to frequently ask a PD patient to make it to the clinic due to the weather, distance, or physical inability. Remote patient monitoring would reduce the need for patients to come to clinic for their tremor to be assessed[20].

In this framework, a set of clinically relevant standardized tasks could be defined for the patients to accomplish while wearing a safe and reliable data collection (i.e., motion capture) system. Information from the data collection system could then be sent directly to their physician.

This method needs to be simple enough to set up in the home en- vironment. Therefore, the main purpose of this study is to: (1) in- vestigate the applicability of using an IMU-based full-body motion capture system in PD participants to track the biomechanics of tremor in all body segments; (2) develop standard motor tasks and algorithms that can detect, segment, and comprehensively output a clinically meaningful tremor score; (3) evaluate the clinical utility of the kine- matic measure by correlating this new score with the UPDRS scores during the proposed standard tasks; (4) demonstrate the ability of this method to clearly separate tremor dominant from non-tremor dominant PD, as well as from healthy participants.

Thefinal goal of this study is to introduce an objective framework in which whole-body tremor severity could be clinically quantified and analyzed reliably following remote monitoring of PD patients.

Additionally, we hope to apply a similar framework to aid in the op- timization of more invasive treatments for PD such as deep brain

stimulation (DBS), wherein setting adjustments for the implanted de- vice often require months offine-tuning.

2. Methods 2.1. Participants

Forty PD participants and twenty two healthy controls participated in the study. The PD participants were recruited from the Movement Disorders Center in London Health Sciences Center (London, ON, Canada). Participant demographics are listed inTable 1. In this table, LED denotes levodopa equivalency dosage, a measure which combines the contribution made by each PD medication into a single dosage, based on a standardized formula[28].

Inclusion criteria for PD participants were: (1) idiopathic Parkinson’s disease, (2) males or females of 45 to 85 years of age, (3) have been on stable doses of anti-Parkinson medication, including any levodopa preparation, (4) no dementia or psychiatric abnormalities, (5) able to walk without any walking aids, (6) able to provide informed consent. Control participants were healthy and age-matched with no dementia or psychiatric abnormalities. The study was approved by the Human Subjects Research Ethics Board (REB # 107253) at Western University (London, ON, Canada) and all participants provided their written informed consent prior to participation.

2.2. Assessment tools

Tremor severity was measured using UPDRS as well as a full body, IMU-based, wearable motion capture system (IGS-180–Synertial Ltd., UK). This system includes 17 embedded IMUs and the placement of each sensor is illustrated inFig. 1. Each IMU includes three accel- erometers, three gyroscopes, and three magnetometers that measure linear accelerations, angular velocities, and magnetic forces in all di- rections (3-axis each). The relative position and orientation between adjacent sensor units enable a fusion software (using quaternion method[40,41]—developed commercially available by Inertial Labs Inc., Virginia, USA) to identify body joint angles. The fusion software is implemented on a main processing unit (MPU) and communicates wirelessly to a receiver linked to a personal computer. Data acquisition was performed at 60 Hz sampling rate using IGS-Bio software Version 2.56 (Synertial Ltd., UK), configured for full-body human motion. The complete list of measured joint angles can be seen in the Appendix.

The system is also capable of generating a real-time 3-D animation based on the recorded joint angle data. The animation can be easily replayed at any time from the recorded data. A snapshot of the ani- mation is illustrated inFig. 2(c).

A specific, minimal set of instructions were given to all participants.

These instructions were standardized and simple enough that partici- pants could don the suit system with no or minimal spousal assistance.

This was done in an effort to judge the technology’s ease of use.

Throughout this manuscript, the upper-extremity UPDRS means the sum of the UPDRS-Part III tremor scores (either rest or postural) for right arm and left arm. The upper-extremity UPDRS score for both arms together ranges from 0 = Normal to 8 = Severe. Hence, total UPDRS Table 1

Participants demographics.

Controls (n=22) PD (n=22)

Age (years), mean (SD) 64.90 (3.91) 65.28 (7.49)

Males, n (%) 11 (50%) 34 (85%)

PD duration (years), mean (SD) 8.8 (4.33)

L-Dopa duration(yrs), mean (SD) 7.1 (4.36)

LED*(mg/day), mean (SD) 1082 (473)

* Levodopa equivalency dosage; calculation for LED was based on a standardized formula from literature by Tomlinson et al.[28].

(3)

means the sum of UPDRS-Part III rest tremor scores for head, right arm, left arm, right leg, and left leg. Total UPDRS ranges from 0 = Normal to 20 = Severe. It is noted that the UPDRS-Part III does not include the assessment of full-body postural tremor. While the UPDRS includes chin/jaw tremor, the kinematic recordings do not measure this tremor.

However, head tremor is measured and included in the analysis.

2.3. Clinical tasks and procedure

Tremor can be assessed during a variety of different tasks; however, for remote monitoring, the selected task needs to be simple, clinically valid, and long enough for data analysis. In clinic, tremor is mainly assessed while sitting at rest (referred to as rest tremor) and sitting while performing a posture (referred to as postural tremor)—Fig. 2.

All participants were asked to perform the following standard tasks

twice while seated:

(1) Rest: while sitting, the participants rested both of their forearms on the arms of a chair with hands hanging loosely offthe edge. The participants were asked to hold this position for 20seconds s.

(2) Posture: while sitting, participants fully extended their arms for- ward with hands pronated at shoulder height level and asked to hold this position for 20seconds s.

The PD participants were assessed OFF and ON levodopa. In this study, a participant was considered OFF levodopa when it has been 12 or more hours since their last dose of levodopa medication has been taken. A participant was considered ON levodopa at approximately 45–60minutes min after taking their normal dose of levodopa.

2.4. Tremor severity score (TSS) calculation

A tremor severity score (TSS) was calculated at each body joint so tremor severity could be segmented based on the limb of interest. To calculate this score, signals containing angular displacement of all body joints were obtained from the motion capture system. The signals were band-passfiltered from 2 Hz to 20 Hz to eliminate non-tremor move- ments. The root-mean-square of eachfiltered signal (associated with each joint) was then considered as the TSS in that body joint. The root- mean-square of all joints associated with each body parts’(head, right arm, left arm, right leg, and left leg) TSS was calculated and called segmental TSS. Finally, the sum of all segmental TSS values were cal- culated and denoted as full-body TSS. Therefore, the total TSS (in de- grees) can be calculated as

=

=

TSS RMS F[ ( )]J

i n

Hz i 1

2 20

(1) whereJiis theithith body joint movement, n (= 47, see Appendix) is the maximum number of joint movements involved in the calculation, F2−20Hz(⋅) is the function whichfilters the signal from 2 Hz to 20 Hz, andRMS[⋅] is a function to calculate the root-mean-square of all points forming the signal. This method was developed based on a review of the relevant literature (see sectionSection 4). Hence, it is always possible to either report a single full-body tremor severity score or present the decomposition of tremor severity per body segment or even per single joint movement (e.g., wristflexion-extension, ulnar-radial, and prona- tion-supination–See Appendix).

2.5. Validation and statistical analysis

In order to verify the effectiveness of our method, we conducted Fig. 1. Body positioning of each IMU sensor.

(a) (b) (c)

Fig. 2. Clinical standard tasks to assess tremor at (a) rest, (b) posture, while wearing Synertial’s IGS-180 wearable motion capture system, (c) a snapshot of the generated animation.

(4)

relevant statistical analyses to mainly investigate the correlation be- tween the UPDRS (as the gold standard) and the calculated tremor se- verity score results. The PD participants outcomes were also compared between OFF and ON medication states as well as to the control par- ticipants. Either parametric or non-parametric statistical tests were se- lected following Shapiro-Wilks test of normality. The results are re- ported with 95% confidence interval. The statistical analyses were carried out with SPSS software, version 22.0 (SPSS Inc., IL, USA).

3. Results

3.1. Upper-extremities rest TSS compared with UPDRS

In order to conduct a more thorough comparison, the PD partici- pants were initially divided into two groups called tremor-dominant and non-tremor-dominant. This separation was made based on asses- sing PD participants OFF-medication and while sitting at rest using the UPDRS-Part III (motor examination). Any PD participant who had a total UPDRS score higher than two was marked tremor-dominant PD and the remainder of the PD participants were marked non-tremor- dominant PD (Table 2).

Figs. 3–5show the results of the kinematic assessment compared with the UPDRS scores.Fig. 3illustrates the upper-extremity TSS cal- culated for all participants while sitting at rest. It also compares the results with the sum of the upper extremity UPDRS scores collected during the same task. As can be seen, while sitting at rest, the TSS for tremor-dominant PD participants in the OFF state was significantly higher than the non-tremor-dominant PD in their OFF state (p

< 0.001), the tremor-dominant PD participants in their ON state (p= 0.034), and the healthy controls (p= 0.001). Furthermore, the TSS of the tremor-dominant PD participants in their ON state was still sig- nificantly higher than the healthy controls (p < 0.039). The same pattern of change was observed in the UPDRS upper-extremity results.

3.2. Full-body rest TSS compared with UPDRS

Fig. 4 presents the full-body TSS during sitting at rest task and compares the results with the total UPDRS. While sitting at rest, the TSS for tremor-dominant PD participants in their OFF state was significantly higher than the non-tremor-dominant PD in their OFF state (p

< 0.001), as well as the tremor-dominant PD participants in their ON

state (p= 0.047), and the healthy controls (p= 0.003). The TSS of the non-tremor-dominant PD participants in their ON state was also sig- nificantly higher than the healthy controls (p = 0.014). The same pattern of change was observed in the total UPDRS scores.

3.3. Upper-extremity posture TSS compared with UPDRS

Fig. 5presents the results associated with the upper-extremity TSS while the participants were performing the posture task. The TSS of upper-extremity for tremor-dominant PD participants in their OFF state was significantly higher than the non-tremor-dominant PD in their OFF state (p < 0.001), as well as the healthy controls (p= 0.008). The TSS of the non-tremor-dominant PD participants in their ON state was also significantly higher than the healthy controls (p= 0.001). The same pattern of change was observed in the total UPDRS scores.

3.4. Correlation analysis

To further validate ourfindings, both Pearson product moment and Spearman rank-order correlations were calculated to identify both the linear and the monotonic relationships between the kinematically measured tremor (TSS) and the UPDRS results.Table 3summarizes the correlation analysis results.

The average time to put on the suit took no more than 3 to Table 2

Total UPDRS-Part III scores for PD participants OFF-medication and sitting at rest; the data were used to group PD participants into dominant and non-dominant tremor.

Tremor dominant Tremor non-dominant

Patient ID Total UPDRS Patient ID Total UPDRS

PD-01 8 PD-03 0

PD-02 4 PD-04 0

PD-06 7 PD-05 2

PD-09 4 PD-07 1

PD-12 7 PD-08 1

PD-13 5 PD-10 1

PD-14 5 PD-11 1

PD-15 4 PD-18 0

PD-16 7 PD-19 2

PD-17 3 PD-21 0

PD-20 4 PD-24 1

PD-22 4 PD-26 0

PD-23 3 PD-28 0

PD-25 4 PD-29 2

PD-27 7 PD-30 0

PD-31 8 PD-33 2

PD-32 8 PD-35 2

PD-34 3 PD-36 2

PD-38 8 PD-37 0

PD-39 1

PD-40 2

Fig. 3. The upper-extremities TSS for all participants while sitting at rest and the equivalent upper extremities UPDRS scores.*Significancep< 0.05. Error bars represent the standard errors.

(5)

4minutes min. Although the data was not specifically collected, most patients were noted to be able to put on the suit themselves. All patients kept the suit on for the entirety of the recording session with no dis- comfort reported. An animation showing the actual performance of the patient is generated for further clinical evaluation (Fig. 2(c)). These animations can also be used by the clinician to rate the clinical state visually in addition to the kinematic sensor data.

4. Discussion

4.1. Limitations of UPDRS

The current standard of care in the management of movement dis- orders associated with PD includes subjective evaluation using the clinical gold standards such as the UPDRS[29]. In our previous work, we illustrated the subjective nature of the UPDRS motor score by re- vealing a weak correlation between four trained raters in scoring a randomized subset of PD participants[33].

Tremor in PD can occur in any body part. While upper extremity tremor is more visible, lower extremity tremor can occur with sig- nificant prevalence[29,30]. This is the reason why the UPDRS requires the measurement of tremor in all body segments. However, there are several limitations to the use of the UPDRS in this scenario. One of those limitations being that the UPDRS takes a single visual assessment

of tremor. The severity of tremor in PD, however, varies over periods and longer assessment times can quantify this better.

Also, a change from rest to posture can produce a silent period termed latency where there is no tremor. This is followed by a re- emergence of the tremor which can be missed in visual assessments.

Additionally, the measurement of tremor in the lower extremities is visually more difficult than in the upper extremities. Thus, the occur- rence of tremor in all body parts, variability over time, latency, and difficulty in lower extremity visual evaluation are all reasons why a multi-sensor whole body quantitative tool is required.

Fig. 4. The full-body TSS for all participants while sitting at rest and the equivalent total UPDRS scores.*Significancep< 0.05. Error bars represent the standard errors.

Fig. 5. Upper-extremity TSS for all participants while holding a posture and the equivalent upper extremities UPDRS scores.*Significancep< 0.05. Error bars represent the standard errors.

Table 3

The linear (Pearson correlation) and the monotonic (Spearman correlation) relationships between the TSS and the UPDRS results.

Pearson corr. Spearman corr.

r p ρ p

Rest OFF Upper ex. 0.741 0.000 0.738 0.000

Full body 0.737 0.000 0.731 0.000

ON Upper ex. 0.770 0.000 0.765 0.000

Full body 0.650 0.000 0.602 0.001

Posture OFF Upper ex. 0.523 0.001 0.643 0.000

ON Upper ex. 0.549 0.000 0.552 0.000

(6)

4.2. Emerging technologies

No reproducible assessment tool has been widely accepted to be considered as the objective gold standard of tremor assessment.

Kinematically recorded movements have a much higher resolution for measuring movements. Although not all aspects of such recordings are necessarily clinically useful, the reliability of an objective measure is inherent, over a subjective visual tool. Indeed, the UPDRS scores are dependent on the amplitude of the tremor and yet this judgement is made visually. Tools that actually measure this amplitude would therefore be justified as being more consistent. Consequently, similar to other such studies [18,31,32,34,35], this study is also lacking the comparison of the proposed high-resolution system with a widely ac- cepted kinematic tool.

Measuring tremor in one extremity is limiting and technologies that are designed to perform such assessment are mostly unable to provide a complete picture of the presence and extent of tremor. The current commercially-available technologies in the area of PD tremor assess- ment are generally lacking such comprehensive capability. They may not be (1) portable[24,25], (2) multi-sensor[19,21], (3) easily worn by patients, or (4) able to transfer data such that an animation can be produced for further visual evaluation.

We initially investigated if a portable full-body motion capture suit is useful for the measurement of whole body tremor in PD patients. The multi-sensor system recorded tremor from all the body segments (limbs, trunk and head) simultaneously. Thus, quantitative and objective measurements could be carried out at the same time point.

In addition, recordings were carried out over a longer duration than the typical visual assessment. This allowed better averaging of tremor over all body segments. Since sensors were placed across all segments, individual segmental tremor was recorded and then summed to provide an overall tremor score. This dealt with the limitations discussed above of single limb and lower limb assessments, as well as tremor onset la- tency.

4.3. Development of TSS measure

Two clinically applicable tasks were selected based upon the ease of administration and what is clinically useful in normal practice [23].

The sitting-at-rest task was intended to track resting tremor which is part of the UPDRS and measured simultaneously in all the segments.

The posture task in the upper extremities was designed to capture ac- tion tremor.

Although PD tremor is classically a rest tremor, our prior work has shown that postural tremor is significant in PD[7,18,31]. The duration of this task was designed to include the re-emergent tremor. This is important as such postural tremor can be a significant component of the overall PD tremor and be responsible for an impairment of the quality of life of patients as it affects task performance. Thus, quantification of postural tremor in PD is as, if not more, important than rest tremor.

The method of calculating the tremor severity score - TSS was de- veloped based on reviewing recent successful studies found in the lit- erature [4-6,13-15,17,32]. For instance, Powell et al. [4]considered raw tremor acceleration waveforms and performed a time frequency analysis. A group of Gaussian bandpass filters, centred at 3 Hz, 4 Hz, 5 Hz, and 6 Hz with equal bandwidth of 1 Hz were then designed to calculate tremor severity.

Rigas et al.[6]also used accelerometric data and analyzed them using both a low-pass FIRfilter with cut-offfrequency of 3 Hz as well as a band-pass FIRfilter with frequencies from 3 to 12 Hz to extract fea- tures corresponding to PD tremor severity. The frequency band was selected according to methods suggested by[7,18]. Tracking below this band implies dyskinesia (uncontrolled twisting or jerking movements), a common side effect of levodopa seen in PD patients.

4.4. Evaluation of TSS measure

The proposed TSS was different between the healthy control and PD participants. As expected, the overall TSS was higher across all mea- sures in PD. A small amount of tremor was recorded by the sensors in the control group as they are more sensitive than the visual assessment of the rater. Further, since the n value was large in this dataset, we could categorize patients into either tremor-dominant or non-dominant groups based on their total UPDRS tremor scores.

This separation showed the granularity of the kinematic system’s ability to further separate out differences in tremor quantity between these groups of participants. Once separated into groups, the correla- tion between the UPDRS and the kinematic rating was robust. This relationship is shown in both the linear and monotonic correlational analysis. The correlation for postural tremor is however only moderate.

One possible reason for this is the fact that in PD, postural tremor can be delayed by a latency of up to 5seconds s after a change of position of the upper limbs from rest (Fig. 2(a)) to arms outstretched (Fig. 2(b)). If the postural visual rating is done immediately after this change in po- sition, and latency occurs, then it might be scored a 0 or 1 (i.e. little to no tremor). The kinematic analysis technique measures tremor for a longer duration and this re-emergent tremor after the latency is more likely to be captured.

4.5. Logistics

Minimal assistance was given to the patients in putting on the suit.

For each task, the patients were instructed with minimum details. This was done to determine if the participants would be able to set up the experiments in their home environment. The ease and speed of donning the suit with or without help is an important aspect of the technology- human interface. Therefore, although not specifically tested in this study, we anticipate that human instructions can reliably be replaced with either an animation or some audio cue based on a predetermined routine.

The relative ease of use of the technology encourages its use in the remote or home based measurement scenarios. The home-based appli- cation of the proposed technology can make it a platform for remote assessment and be used for optimizing therapeutic medical and surgical interventions. The importance and applicability of these approaches has have already been highlighted in the literature[36–38]. For instance, MercuryLive, a web-based integrated platform has recently been de- veloped to enable clinicians to analyse wearable sensor data and in- teract remotely with PD patients in the home setting [36]. Sanchez- Ferro et al.[39]presented a systematic review on the new methods for the assessment of Parkinson’s disease, including tremor.

While the home-based application of the proposed technology is encouraged in this research work, the system is simple enough to be used in the clinic to provide clinicians with objective measures which do not hinder their standard practice. The technology based objective measurements can be used to provide scores for tremor to the clinician.

If needed, the animation can also be used to perform a visual as- sessment without having to have an actual video of the patient. The animation can be created locally from de-identified raw data thereby protecting the confidentiality of the data during transmission from a remote site or in the clinic. As the system we used is wireless, this then becomes a standardized, portable comprehensive assessment tool for clinic and remote utilization. The system then allows the physician to have “smart” data that could not only expedite the assessment and improve access for patients but also help tailor therapy directly to the patient’s needs.

The proposed wearable technology in conjunction with the existing web-based platforms could assist in expanding the scope of PD tele- medicine. Such future technologies would effectively enable the clin- icians to make informed and rapid decisions regarding the state of the patient and then optimize the necessary therapeutic intervention.

(7)

Preliminary evidence also exists suggesting that wearable sensors could facilitate the process of optimizing stimulator settings in patients with Parkinson’s disease undergoing DBS[37].

5. Conclusion

The results showed that a simple measure defined as the full-body tremor severity score or TSS can help in assessing PD tremor using such multi-sensor wearable technology. The TSS correlated well with scores obtained from a gold standard clinical scale, the UPDRS. This quanti- tative approach reliably separated tremor-dominant PD participants versus non-tremor-dominant and healthy participants. The findings suggest that such portable wearable motion capture systems can be used for automatically detecting and reporting the severity of full-body PD tremor.

If the suggested method is implemented and applied, an accurate animation developed from the sensor data could be remotely generated and considered along with the calculated TSS. In conclusion, the ani- mation combined with the calculated TSS provides physicians with a comprehensive objective assessment of their patients, allowing them to

make medication adjustments accordingly. The physicians can review the actual movements with no need to transfer and watch the actual video of the patient.

An ideal movement disorders remote monitoring system would potentially include either wearable sensors or sensors embedded in the environment to collect, analyse, and transfer relevant data. Such a system could provide clinicians with real-time reporting of the severity of multiple symptoms to optimize both the dosing and timing of a suggested medical intervention. This work was an attempt, foremost, to introduce a new tool and analysis approach for tremor as a prototype to pave the way for further developments.

Acknowledgement

The authors appreciate the participation of Dr. Andrew Parrent, Dr.

Matthew Hebb, and Dr. Keith MacDougall (Western University, London, ON, Canada). We also acknowledge the efforts of Mr. Greydon Gilmore (Lawson Health Research Institute, London, ON, Canada) and Dr.

Fariborz Rahimi (Bonab University, East Azerbaijan, Iran).

Appendix A Table 4

The list of joint angles, included in the assessment.

Body part Segment Motion

Head Head Flexion/extension

Lateral tilt Axial rotation

Trunk Right clavicle Axial rotation

Depression/elevation Retraction/protraction

Left clavicle Axial rotation

Depression/elevation Retraction/protraction

Thorax Flexion/extension

Lateralflexion Rotation

Pelvis Flexion/extension

Lateralflexion Rotation

Right arm Right wrist Flexion/extension

Ulnar/radial Pronation/supination

Right elbow Flexion/extension

Pronation/supination

Right shoulder Flexion/extension

Abduction/adduction Rotation

Left arm Left wrist Flexion/extension

Ulnar/radial Pronation/supination

Left elbow Flexion/extension

Pronation/supination

Left shoulder Flexion/extension

Abduction/adduction Rotation

(8)

Right leg Right hip Flexion/extension Abduction/adduction Rotation

Right knee Flexion/extension

Rotation

Right ankle Flexion/extension

Inversion/eversion Rotation

Left leg Left hip Flexion/extension

Abduction/adduction Rotation

Left knee Flexion/extension

Rotation

Left ankle Flexion/extension

Inversion/eversion Rotation

References

[1] C.F. Pasluosta, H. Gassner, J. Winkler, J. Klucken, B.M. Eskofier, An emerging era in the management of Parkinson’s disease: wearable technologies and the internet of things, IEEE Health Info. 19 (6) (2015) 1873–1881.

[2] W.-L. Liao, F.-J. Tsai, Personalized medicine: a paradigm shift in healthcare, BioMedicine 3 (2) (2013) 66–72.

[3] G. Grimaldi, M. Manto, Neurological tremor: sensors, signal processing and emerging applications, Sensors (Basel). 10 (2) (Jan. 2010) 1399–1422.

[4] H.C. Powell, M.A. Hanson, J. Lach, On-body inertial sensing and signal processing for clinical assessment of tremor, IEEE Trans. Biomed. Eng. 3 (2) (2009) 108–116.

[5] A.T. Tzallas, M.G. Tsipouras, G. Rigas, D.G. Tsalikakis, E.C. Karvounis, M.

Chondrogiorgi, F. Psomadellis, J. Cancela, M. Pastorino, M.T.A. Waldmeyer, S.

Konitsiotis, D.I. Fotiadis, PERFORM: a system for monitoring, assessment and management of patients with Parkinson’s disease. Sens. (Basel). 14 (11) (Jan. 2014) 21329–21357.

[6] G. Rigas, A.T. Tzallas, M.G. Tsipouras, P. Bougia, E.E. Tripoliti, D. Baga, D.I.

Fotiadis, S.G. Tsouli, S. Konitsiotis, Assessment of tremor activity in the Parkinson s disease using a set of wearable sensors, IEEE Trans. Inf. Technol. Biomed. 16 (3) (2012) 478–487.

[7] F. Rahimi, C. Bee, D. Debicki, A.C. Roberts, P. Bapat, M. Jog, Effectiveness of boNT A in Parkinson’s disease upper limb tremor management. Can. J. Neurol. Sci.

40 (5) (2013) 663–669.

[8] A. Hassan, S.S. Wu, P. Schmidt, I.A. Malaty, Y.F. Dai, J.M. Miyasaki, What are the issues facing Parkinson’s disease patients at ten years of disease and beyond?:

data from the NPF-QII study, Parkinsonism Relat. Disord. 18 (2012) S10–S14.

[9] G. Grimaldi, M. Manto, Tremor from Pathogenesis to Treatment, Morgan and Claypool, San Rafael, CA USA, 2008.

[10] L. Cleeves, L.J. Findley, M. Gresty, Assessment of rest tremor in Parkinson’s disease, Adv. Neurol. 45 (1986) 349–352.

[11] C.G. Goetz, B.C. Tilley, et al., Movement Disorder Society-Sponsored Revision of the Unified Parkinson’s Disease Rating Scale (MDS-UPDRS): scale presentation and clinimetric testing results, Mov. Disord. 23 (15) (2008) 2129–2170.

[12] M.J. Zwarts, G. Drost, D.F. Stegeman, Recent progress in the diagnostic use of surface EMG for neurological diseases. J. Electromyogr. Kinesiol. 10 (5) (2000) 287–291.

[13] S. Spieker, V. Ströle, A. Sailer, A. Boose, J. Dichgans, Validity of long-term electromyography in the quantification of tremor, Mov. Disord. 12 (6) (1997) 985–991.

[14] F. Foerster, M. Smeja, Joint amplitude and frequency analysis of tremor activity, Electromyogr. Clin. Neurophysiol. 39 (1) (1999) 11–19.

[15] A. Salarian, H. Russmann, C. Wider, P.R. Burkhard, F.J.G. Vingerhoets, K.

Aminian, Quantification of tremor and bradykinesia in Parkinson’s disease using a novel ambulatory monitoring system. IEEE Trans. Biomed. Eng. 54 (2) (Feb. 2007) 313–322.

[16] R.D. Fusfeld, Classification of the electromyogram by a pattern-recognition method, Med. Biol. Eng. Comput. 20 (July) (1982) 496–500.

[17] J.A. Gallego, E. Rocon, J.O. Roa, J.C. Moreno, J.L. Pons, Real-time estimation of pathological tremor parameters from gyroscope data, Sens. (Basel). 10 (3) (Jan.

2010) 2129–2149.

[18] F. Rahimi, C. Bee, A. South, D. Debicki, M. Jog, Variability of hand tremor in rest and in posture a pilot study, Proc. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. EMBS (2011) 470–473.

[19] T.O. Mera, M.A. Burack, J.P. Giuffrida, Quantitative assessment of levodopa- induced dyskinesia using automated motion sensing technology, Proc. Annu. Int.

Conf. IEEE Eng. Med. Biol. Soc. EMBS (2012) 154–157.

[20] A. Sama, C. Perez-Lopez, J. Romagosa, D. Rodriguez-Martin, A. Catala, J.

Cabestany, D.a. Perez-Martinez, A. Rodriguez-Molinero, Dyskinesia and motor state detection in Parkinson’s disease patients with a single movement sensor, Proc.

Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. EMBS (June) (2012) 1194–1197.

[21] G. Mostile, J.P. Giuffrida, O.R. Adam, A. Davidson, J. Jankovic, Correlation between Kinesia system assessments and clinical tremor scores in patients with essential tremor, Mov. Disord. 25 (12) (Sep. 2010) 1938–1943.

[22] M.M. Koop, A. Andrzejewski, B.C. Hill, G. Heit, H.M. Bronte-Stewart,

Improvement in a quantitative measure of bradykinesia after microelectrode re- cording in patients with Parkinson’s disease during deep brain stimulation surgery, Mov. Disord. 21 (5) (May 2006) 673–678.

[23] N. Alam, B. Johnson, J. Gendreau, K. Tavakolian, C. Combs, R. Fazel-rezai, A.E.

Setup, tremor quantification of Parkinson’s disease - a pilot study, Proc. IEEE Int.

Conf. Electro. Info. Tech. (2016) 755–759.

[24] K.L. Schoffer, V. Patterson, S.J. Read, R.D. Henderson, J.D. Pandian, J.D.

O’Sullivan, Guidelines forfilming digital camera video clips for the assessment of gait and movement disorders by teleneurology, J. Telemed. Telecare 11 (7) (2005) 368–371.

[25] Z. Uhríková, O. Šprdlík, M. Hoskovcová, A. Komárek, O. Ulmanová, V. Hlaváč, C.D. Nugent, E. Ruzicka, Validation of a new tool for automatic assessment of tremor frequency from video recordings, J. Neurosci. Methods 198 (1) (may 2011) 110–113.

[26] F. Rahimi, C. Bee, C. Duval, P. Boissy, R. Edwards, M. JogEMAP Group, Using ecological whole body kinematics to evaluate effects of medication adjustment in parkinson’s disease, J. Parkinsons. Dis. 4 (4) (2014) 617–627.

[27] F. Rahimi, C. Duval, M. Jog, C. Bee, A. South, M. Jog, R. Edwards, P. Boissy, Capturing whole-body mobility of patients with Parkinson disease using inertial motion sensors: expected challenges and rewards, Conf. Proc. IEEE Eng. Med. Biol.

Soc. 2011 (Jan. 2011) 5833–5838.

[28] C.L. Tomlinson, R. Stowe, S. Patel, C. Rick, R. Gray, C.E. Clarke, Systematic review of levodopa dose equivalency reporting in Parkinson’s disease, Mov. Disord.

25 (15) (Nov. 2010) 2649–2653.

[29] J. Jankovic, Parkinson’s disease: clinical features and diagnosis. J. Neurol.

Neurosurg. Psychiatry 79 (4) (Apr. 2008) 368–376.

[30] A. Burleigh, F. Horak, et al., Levodopa reduces muscle tone and lower extremity tremor in Parkinson’s disease, Can. J. Neurol. Sci. / J. Can. des Sci. Neurol. 22 (4) (Nov. 1995) 280–285.

[31] F. Rahimi, O. Samotus, J. Lee, M. Jog, Effective management of upper limb Parkinsonian tremor by IncobotulinumtoxinA injections using sensor-based bio- mechanical patterns. Tremor, and Other Hyperkin Mov. 5 (2015).

[32] K. Niazmand, K. Tonn, A. Kalaras, U.M. Fietzek, J.H. Mehrkens, T.C. Lueth, Quantitative evaluation of Parkinson’s disease using sensor based smart glove, Comput. Med. Syst. (CBMS), 2011 24th Int. Symp. 2011, pp. 1–8.

[33] M. Delrobaei, S. Tran, G. Gilmore, K. McIsaac, M. Jog, Characterization of multi- joint upper limb movements in a single task to assess bradykinesia, J. Neurol. Sci.

368 (Sep. 2016) 337–342.

[34] N.H. Ghassemi, F. Marxreiter, C.F. Pasluosta, P. Kugler, J. Schlachetzki, Combined accelerometer and EMG analysis to differentiate essential tremor from Parkinson’s disease, IEEE Conf. Eng. in Med. and Bio. (2016) 672–675.

[35] B.T. Cole, P. Ozdemir, S.H. Nawab, Dynamic SVM detection of tremor and dys- kinesia during unscripted and unconstrained activities, Conf. Proc. IEEE Eng. Med.

Biol. Soc. (2012) 4927–4930.

[36] B. Chen, S. Patel, T. Buckley, R. Rednic, D.J. Mcclure, L. Shih, D. Tarsy, M.

Welsh, P. Bonato, A web-based system for home monitoring of patients with Parkinson’s disease using wearable sensors, IEEE Trans Biomed Eng. 58 (3) (2011) 831–836.

[37] S. Patel, C. Mancinelli, R. Hughes, A. Dalton, L. Shih, P. Bonato, Optimizing deep brain stimulation settings using wearable sensing technology, 4th Int. IEEE/EMBS Conf. Neural Eng. NER’09 2009, 2009, pp. 6–9.

[38] A. Sánchez-Ferro, W. Maetzler, Advances in sensor and wearable technologies for Parkinson’s disease, Mov. Disord. 31 (9) (Sep. 2016) 1257-1257.

[39] A. Sánchez-Ferro, M. Elshehabi, C. Godinho, D. Salkovic, M.A. Hobert, J.

Domingos, J.M. van Uem, J.J. Ferreira, W. Maetzler, New methods for the as- sessment of Parkinson’s disease (2005 to 2015): a systematic review, Mov. Disord.

31 (9) (Sep. 2016) 1283–1292.

[40] H.P. Nguyen, F. Ayachi, C. Lavigne-Pelletier, M. Blamoutier, F. Rahimi, P.

Boissy, M. Jog, C. Duval, Auto detection and segmentation of physical activities during a Timed-Up-and-Go (TUG) task in healthy older adults using multiple in- ertial sensors. J. Neuroeng. Rehabil. 12 (1) (Jan. 2015) 36.

[41] A.M. Sabatini, Quaternion-based extended Kalmanfilter for determining orienta- tion by inertial and magnetic sensing. IEEE Trans. Biomed. Eng. 53 (7) (Jul. 2006) 1346–1356.

Referensi

Dokumen terkait

This study therefore came up with autonomy strategies that can assist the leadership of the hospital to harvest to their full potential of nurses hence

Tujuan dari penelitian ini membuktikan pengaruh massage frirage terhadap peningkatan ROM (Range Of Motion) gangguan cedera lutut pada atlet basket UKM UNNES dan PPLP

The significant influence of Fatimid dynasty authority on the development of Islamic education is very evident in the efforts of the caliphs in building educational institutions, even

The choice of such videos simplifies the visual detection of the motion occurrence and limits it to pre-determined values that can be easily compared with the motion detected by the

Example of ‘My Science World’ section Chapter 8 Force and Motion has the highest number of examples in daily life applications which consists of 17 examples in comparison with Chapter

see Fig.3 : Installation of 9 DOF motion sensors and three dimensional motion analysis maker to the trans-tibial prosthesis for track collaborative research Available Facilities and

The objective of this study is to examine the content validity of Al-Qalb as one of the dimensions to be incorporated in the institutional leadership competencies framework and measure

This really portrays the image of Kelantan as the cradle of Malay culture because most of the traditional Malay cultures can be found there Mohd, 2011.Overall, the subjects of the study