1 Microbalance Max capacity: minimal 6,1 g 1 unit Readability: 0.001 mg/ 1 µg
Repeatibilty : maksimal 1 µg Linearity : maksimal 4 µg
Minimum sample weght : ≤2 mg Include: IQ/OQ, training, sertifikat IQ/PQ
Meja timbangan anti getar, Non magnetic material
2 Atomic Absorption Spectrofotometer Basic : 1 unit
Analysis methods : Dual Atomizer (Flame & Graphite Furnace) Optics : Optical double beam
Wavelength range : 185 to 900 nm Detector : photomultiplier tube
background correction : - BGC-SR (high-speed self-reversal method) - BGC - D2 (D2 lamp method)
Monochromator : Czerny turner/Ebert-Fastie HC lamps : minimal 6 lamps
measurement mode : flame continous method, flame micro sampling method, furnace method, flame emission method
Flame : 1 set
Burner : titanium /air-cooled burner Nebulizer : Platinum-iridium capillary Safety measures : Yes
Auto Sampler : 1 set
Repeatability : max 1%RSD (20 µL) Rinse water bottle : Yes
Functions : Zero point detection, auto rinse, random acces, auto mixing
Harga Satuan Harga Total SPESIFIKASI PENGADAAN ALAT LABORATORIUM
No. Nama Barang
Jumlah Spesifikasi
Nozzle rinse : Solvent rinse aspiration method (Flame) and Solvent Rinse discharge method (Furnace)
Computer and Printer : Intel core i5,RAM minimal 4GB, windows 7 professional minimal 32-bit, HDD minimal 500GB, Intel Graphics Media accelerator, USB Optical Mouse, USB Standard Value
Keyboard, Media Card Reader 1 set
Complete with :
Mn, Sn, Sb, Ag (with Certificate of Analysis ) 20 set
Hollow cathode Lamps for Pb, Hg, Cd, Zn, As, Al, Mg, Ca, Na, K, Cu, Fe, Cr, Co, Ni, Se, Mn,
FURNACE ANALYSIS (PPB LEVEL) :
Heating temperature range : Ambient to 3000 °C 1 unit
Safety measures : Yes
Heating Control system : Yes Complete with:
Pyrocoated Graphite Tube min 5 set
High Density Graphite Tube min 5 set
Cooling water circulator 1 set
Transformer 1 set
Cooler conector kit 1 set
Regulator Assy 1 set
Include in :Warranty min a year, Instalation and training
3 Ultrasonic - degasser degas function for the efficient degassing of the cleaning liquid and for laboratory purposes 2 unit Mains voltage (Vac) : 220-240 V
Tank internal : minimal 505 x 300 x 200 mm (WHD) Basket internal : Yes
Max. filling volume tank : minimal 28 L Material tank & casing : Stainless steel Carrying handles : Yes
Include :
Guarantee min a year Instalation and training
4 Automatic Developing Chamber 1 unit
- Fully automatic development of 20x10 cm and 10x10 cm TLC/HPTLC plates
- the option "Humidity Control" allows reproducible chromatography at defined activity of the layer - chamber saturation is established at the same time (time controlled)
- Control of time for chamber saturation, pre-conditioning and chromatography
- control of solvent introduction for chamber saturation, pre-conditioning and chromatography - recording of detected developing distance and time
- recording of temperature and humidity
- the desired activity of the stationery phase is established within 2 to 5 minutes and the on-line detection
- The opening of the chamber during manual development as well as all other human and environmental factors have been eliminated in Automatic Developing Chamber
Electrical connections : 100-240 V AC 50/60 Hz, 20 W Accessories :
Option "Humidity Control" provides for control of layer activity by controlling the relative humidity in the developing chamber. This is done by means of salt solution (salt solution not included)
Include in: Guarantee min a year, instalation and training. Compatible with "WinCATS-planar chromatography Manager"
5 UPS 5 KVA Kapasitas 5000 VA / minimal 3250 watt 2 unit
Frequency : minimal 45 maksimal 65 Hz Voltage : 220/230 V
Efficiency : minimal 85% Back Up Time : minimal 10 min
Include: Guarantee min a year, instalation
6 Refrigerator Refrigerator Type : Top Mount
capasity iso gross minimal 560 (ℓ/CU.FT) 1 unit
Dimension (WHD) : minimal 755x1777x707 mm frezer compartment :
Multi Air flow : Yes Temperred glass : Yes
7 Incubator 1 unit
Textured Stainless Steel Casing fully insulated stainless steel doors with compression door lock
inner glass doors rear zinc-plated steel on lockable castors Interior - Heating Concept
easy-to-clean interior,made of stainless steel stainless steel grids
Textured stainless steel casing (WHD) : minimal 585 x 787 x 514 mm Interior - Heating Concept (WHD) : minimal 400 x 400 x 330 mm Volume minmal 53 L
Temperature Range
5°C above ambient up to +80°C Voltage / Power Rating
230 V (+/- 10 %), 50/60 Hz Include:
Original operation manual Certificate of origin Certificate of Calibration
Sub Total Pajak 10 % Total