Dok. 01.02.100.04.12.0678
2 HPLC with auto & manual injector, UV-VIS Detector with manual loop 500µl *) Unit 1
Specifications :
- System Pump : Binary and Quaternary
- Flow rate setting range : 0.001mL/min. to 10 mL/min.
- Flow rate accuray :
- Flow rate precission : ≤ . %
- Number of solvent mixed : minimal 4
- Mixing - concentration accuracy : ≤ . % Detector uv-vis
- Light source : D2 lamp
- Wavelength range : 190-700 nm or wider
- Wavelength accuracy : ± 1 nm
- Wavelength precision : ± 0,1 nm
- Function : Dual Wavelenght
- Linearity : 2.5 AU
Column Oven
- Temperature setting range :
- Temperatur control precision : ≤ , o C
- Coulumn storage : min 4 columns
Autosampler
- Injection Precision : < 0,5% or better
- Injection volume : 0,1-100 uL
- Number of sample vials : minimal 100 vial@±1-2 ml
- Sample carryover :
Manual injector
Lampiran Addendum Spesifikasi Teknis
PENGADAAN ALAT LABORATORIUM BALAI BESAR POM di BANJARMASIN RINCIAN SPESIFIKASI TEKNIS PENGADAAN ALAT TAHUN 2012
No. SPESIFIKASI SATUAN JUMLAH
≤ %
4°C to 80°C or wider
Number of connectable units : min 8 units
Computer
Include :
a. Maintenance Kit for Pump b. Auto rinsing kit for Pump c. Mixer SUS
d. Sample vial, 1 - 2ml (100pc/ pk) min 2 pkg e. Micro syringe 100 uL 1 pcs
f. C18 Coulumn 4.6x250 mm, 5 um min 2 pcs g. C 8 Coulumn 4.6x250 mm, 5 um min 2 pcs h. C18 Coulumn 4.6x150 mm, 5 um min 1 pcs i. Check valve inlet min 2 pcs
j. HUB W/ Cable k. Tools Kit l UPS 5 KVA m. Reservoir Tray
n. Sample loop 500µl for manual ijnector o. Instalasi
p. Training
q. Original manual book r. Factory cerificate s. Guarantee min 1 year
3 HPLC with autosampler, UV-VIS Detector *) Unit 2
Specifications :
Min Intel Core i3-2100 Processor, 4GB DDR3, Windows 7 Professional 32-bit, HDD 500GB, DVD RW, Intel Graphics Media Accelerator (GMA) X4500HD,intel G-41 chipset, USB Optical Mouse , USB Keyboard, , Monitor LCD 18,50", Laser colour printer, USB 2 port
Pompa
- System Pump : Quaternary
- Flow rate setting range : 0.000 mL/min. to 10.000 mL/min. (0.050 to 5.000 mL/ min typical)
- Flow rate accuray :
- Flow rate precission : ≤ . %
- Gradient for easy method development : min 10 gradient curves function
- Number of solvent mixed : minimal 4
- Mixing - concentration accuracy : ≤ . % Detector UV-Vis
- Light source : D2 lamp
- Wavelength range : 190-700 nm or wider
- Wavelength accuracy : ± 1 nm
- Wavelength precision : ± 0,1 nm
- Flow cell design :
- Function : Dual Wavelenght
- Linearity : 2.5 AU
- Lamp optimization :
Column Oven
- Temperature setting range :
- Coulumn storage : min 2 columns
Autosampler
- Injection volume : 0,1-100 uL
- Number of sample vials : minimal 100 vial@±1-2 ml
- Sample carryover :
- Injection Precision : < 0,5% or better
Computer
≤ %
maximizes energy throughtput, eliminates refractive index effect, ensuring the highest sensitivity while maintaining optimal spectral performance
Designed to automatically maximize signal to noise in both visible and UV spectrum with the use of the single D2 lamp, extends the useful life of the lamp for consisten result overtime
20°C to 65°C or wider
Include :
a. Tool kit
b. C18 Coulumn 4.6x250 mm, 5 um min 2 pcs c. C18 Guard Coulumn 4.6x20 mm, 5 um min 2 pcs d. Sample/standard filtration, Ø 13 mm 0.45 µm e. Sample vial, 1 - 2ml (100pc/ pk) min 1 pkg f UPS 2 KVA
g. Instalasi h. Training
i. Original manual book and software j. Factory cerificate
k. Guarantee min 1 year
4 Aquadestillata (Glass-Destilation) *) unit 1
Spesification
a. Capacity : min. 8 L/ hour
b. Cooling water requirement : ± 144 L/ hour
c. Conductivity : ± 2.2 µs/cm at 20°C
d. Evaporator material : borosilicate glass
e. Condensor material : borosilicate glass
f. Heating element material : Quartz sheathing
g. Low water cut off : in condensor, auto power cut off
h. Energy saving : through distillation of heated cooling water
i. Distillate withdrawl : Right hand side
j. Cooling water outlet : Ø 12.7 mm
k. Impurity detector : Automatic
l. Housing : Galvanized steel, powder coated with epoxy resin
m. Cooling water pressure : > 3-7 bar
Min Intel Core i3-2100 Processor, 4GB DDR3, Windows 7 Professional 32-bit, HDD 500GB, DVD RW, Intel Graphics Media Accelerator (GMA) X4500HD,intel G-41 chipset, USB Optical Mouse , USB Keyboard, , Monitor LCD 18,50", Laser colour printer, USB 2 port
n Dimension (WxDxH) : max 650 x 365 x 390 mm
o. Power input : 400 V/ 50-60 Hz
p. Electrical consumption : max. 6.0 kW
q. Another features : - Electronic impurity detector by automatic water change - Water saving through automatic water cut-off
- Sterillisation of the condensers by steam - CO2 degassing through outlet in the condenser
Included of
- Wall or bench mounting - Instalation
- Guarantee min 1 year
- Original operational manual book
5 Aqua reverse Osmosis System *) unit 1
Spesification
a. Nominal permeate flow : min 16 L/h
b. Product water volume : 160 to 320 L/day
b. Inlet feed water : tap water
c. Ionic content rejection : > 96 %
d. Organics, particulate : > 99 %
microorganism rejection
e. Dimension (HxWxD) : max. 45.7 x 25.5 x 31.5 cm
f. Operating weight : max 15 kg
g. Built in reservoir volume : 6 L
h. Electrical feed voltage : 220 - 240 V i. Tap (feed) water connection : 1/2" Gaz M j. Tap (feed) water pressure : 0.5 - 6 bar
k Water quality : type 3
Accessories/catridge used
- RO membrane catridge (2 pcs) - Sanitization chlorine tablet (2 pack)
- Vent filter for PE reservoir (2 pcs) - Pretreatment kit (2 pcs)
- 100 L PE reservoir (1 pc) - 1200 L tank (1 pc) - Water pump (1 pc) Included of
- Installation
- Guarantee min. 1 year
- Original operational manual book
10 Gas Chromatography (GC) with detector FID, ECD *) Unit 1
Specifications : Column Oven
- Temperature range : + 4o to ⁰ C
- Cool down rate :
- Programmability : Up to 20
- Retention time repeatibility : ≤ , % for -deca e Injection Port
a. Split/Splitless Injection Unit ( 2 set )
- Temperature range : Room temperature +10 o C to 400 o C or better - Pressure setting range : 0 - 900 kPa or better
- Split ratio setting range : 1-8000 or better
- Total Flow setting range : 0 -1200 mL/min or better
- Programmable steps : 7
b. Injection Unit for Widebore Column
- Temperature range : Room temperature +10°C to 400°C or better
- Pressure setting range :
- Total Flow setting range :
- Programmable steps :
Detectors
˂ i utes fro °C to °C
Four controlledr detectors can be installed simulatenously and individually temperature - Flame Ionization Detector (FID)
Specification :
a. Temperature range : to 400°C or higher
b. Minimum detection quantity : 1.5 pgC/s
c. Dynamic range : 107
or greater
- Electron Capture Detector (ECD) Specification :
a. Temperature range : to 350°C or higher
b. Minimum detection quantity : 6 fg/s
c. Dynamic range : 104
or greater
Autoinjector / Autosampler
a. Sample Injection Methode : Liquid sample injection via special Microsyringe.
b. Number of Samples : Minimum 100 vial @1 - 2 mL
c. Number of sample injections : 1 - 99 injections per sample.
Data Processing and Data Analysis Accessories :
- Carrier Gas Supply pipe min 2 pcs - Gas filter, molecular sieve
- He Gas + bottle - Pressure Regulator He - H2 Gas Supply Pipe - H2 Gas + bottle - Pressure Regulator H2 - Oil-less air compressor Computer
- Regulator filter
- Air filter for air compressor (silica gel) - Joint Adapter
- Air supply pipe, 2 pcs
- 10-µL microsyringe min 2 set
- Capillary Column Q-Plot, 30m, 0.53mm ID, 0.25um min 1 pcs - Capillary Column DB-1, 30m, 0.25mm ID, 0.25um min 1 pcs - Capillary Column Stabilwax, 30m, 0.25mm ID, 0.25um min 1 pcs - Capillary Column DB-5, 30m, 0.25mm ID, 0.25um min 1 pcs - Capillary Cutter
- ECD Control Unit - ECD Detector Unit - N2 Gas + bottle for ECD - Presure Regulator N2 for ECD - Oxygen Trap for ECD
- AUX Temperature Control Unit - RS-232 Cable
- 1 - 2 ml Sample Vial Set with white cap & septum, 100 pcs/set min 5 set - UPS 5 KVA
- Instalation - Training
- Guarantee min. 1 year - Original operation manual - Factory Certificate
- Tool Kit
- Training to BAPETEN to get ECD license
16 Trinocular Microscope Unit 1
Spesifications :
- Viewing Heads
- Eyepieces
Paired DIN KHW 10x, FN - Objective Changer
Smooth operating, ball bearing mounted, quadruple nosepiece - Objectives
S. Flat Achromats, SM Planachromats, Planachromats, or phase contrast - Stage
- Condensers
- Illuminator
Built in 6 V 300 W, variable intensity Koehler illumination for apertures up to 1.25 Head : Trinoculer
Eyepieces : KHW 10 x FN 20
Objectives : SM Plan Brightfield 4 x, 10 x, S40 x, S 100 x oil Mech stages : Attached
Condensers : Achromatic NA 1.25 with blue filter Illumination : Koehler 6 V 30 W halogen
Accessories
- USB 2.0 Camera 1.5 megapixel camera
Samples : 15 fps at full 1440x1080 resolution and 60 fds at 640x480
Available with standard binocular heads or trinocular heads for photomicrography, 30 o inclined eyetubes with adjustable
diopter settings and interpupilarrydistance, 360 o rotatable
Attached mechanical stage : 120 mm x 140 mm surface with 75 mm (X) by 40 mm (Y) movement, graduated with vernier readings to 0.1 mm, drop down right or left handed controls
Flat top stage : 171 mm x 140 mm surface with 80 mm (X) by 52 mm (Y) movement, graduated with vernier readings to 0.1 mm, detachable speciment quide, drop down right of handed controls.
Standard models have either an Achromatic NA 1.25 condenser with iris diaphragm and filter tray for brighfiel work, simple NA 1.25 slider phase condenser, or an NA 1.25 Zernike phase condenser with filter tray and independently centerable annuli for phase contrast. Cardioid type NA 1.25 darkfield condenser for high power ( 100x objective only) darkfield work is optional
High speed USB 2.0 interface eliminates a framegrabber and facilitates ease of instalation on both laptop and dekstop computers. Infinity Capture software application included.
- Mount adapter with 0.5 x lens - Personal computer
Included
- Guarantee 1 year - Instalation - Training
- Original operation Manual
21 Safety Cabinet (Yellow) For Flammable Unit 2
Specification :
a. Material :
b. Construction :
c. Load capacity : 90 gallon/ 159 kg
d. Complience with : OSHA and NFPA standards, and most are FM approved
e. Safety device : - dual vents with flame arresters
- paddle handle with slip resistant - fail safe closing mechanism.
f. Dimensions : H x W x D : min 1651 x 1092 x 864 mm
g. No. of shelves / door : 2/2
Included :
- Instalation
- Guarantee min. 1 year
Compact 3-point stainless steel bullet self-latching system and welded construction
- Shut and latch automatically when fusible link melt at 165°F (74°C) under fire conditions
- double-walled steel with 38 mm of insulating air space for fire resistance
Galvanized steel min. 51 mm, fully painted inside-outside with a durable lead free epoxy/ polyester powder paint for chemical resistance.
- Original operation manual
25 4.0 Vertical Auto-desiccator cabinets Unit 1
Protecting valuable humidity-sensitive items, such as reagents. It's full automatic system. electronic dehumidification system that automatically regenerates the desiccant and reduces the relative humidity to about 25% within 8 hours A hygrometer dial in the door allows for easy monitoring of relative humidity.
Spesification
a. Material :
b. Range of humidity control :
c. Power supply :
d. Internal Volume :
e. Dimension (HxWxD) :
f. Color : Clear
g. Accessories :
- 3 perforated shelves that are removable
h. Include
- Instalation
- Guarantee min. 1 year - Original operation manual
27 2.0 Auto desiccator cabinets Unit 3
A hygrometer dial in the door allows for easy monitoring of relative humidity. Spesification
a. Material :
b. Range of humidity control :
c. Power supply :
d. Internal Volume :
e. Dimension (HxWxD) :
f. Color : Clear
Co-polyester plastic up to 25% RH 230 V AC 1.9 cu feet
max 51.8 x 34.1 x 41.4 cm
31.4 x 34.1 x 41.4 cm
- 2 latches and loops for lock or tamper evident seal
Protecting valuable humidity-sensitive items, such as reagents. It's full automatic system. electronic dehumidification systemthat automatically regenerates the desiccant and reduces the relative humidity to about 25% within 8 hours
g. Accessories :
- 2 perforated shelves that are removable h. Include
- Instalation
- Guarantee min. 1 year - Original operation manual