• Tidak ada hasil yang ditemukan

ART Ivanna K.Timotius Heart rate Full text

N/A
N/A
Protected

Academic year: 2017

Membagikan "ART Ivanna K.Timotius Heart rate Full text"

Copied!
11
0
0

Teks penuh

(1)

HEARTRA TE COMPliT4 710N JMPLEMElVTE/) ON f1ELJ) PROliRAMMABLE GATE ARRAY BOARD

Ivanna K. Timot/11.\

HEART RATE COMPUTATION IMPLEMENTED ON FIELD

PROGRAMMABLE GATE ARRAY BOARD

Ivanna K. Timotius

Department of Electronic Engineering, Satya Wac ana Christi5tn University, Salatiga,

i vanna_ timotius@yahoo. com

Abstract

Heart rate is the most basic information of heart activities. This paper aims to

describe a possible real time system to detect electrocardiogram (ECG) signaL count the

R-R interval of the ECG signal, and compute the heart rate. This project is implemented successfully using field programmable gate array (FPGA) board.

Keywords: Heart rate. ECG. FPGA.

1. INTRODUCTION

The electrocardiogram (ECG) is a technique of recording bioelectric currents

generated by the heart. This technique produces an ECG signal, which is a biological

signal that shows the electrical activities of the heart as shown in Fig. I. This signal is used to observe some basic information of the heart condition and determine whether the

heart monitored is suftering from any abnormalities.

R , _ _ _ R-R Interval ;_;_;_;..;.:..::_:c._:;; _ _ _ R ,

MセMセセ@

I I

as

Fig. 1. ECG signal

The most basic information monitored from ECG signal is heart rate. Heatt rate can

be computed using the R-R interval of the ECG signal This paper aims to make a real

time system to detect the ECG signaL count the R-R interval, and compute the heart rate.

[image:1.1053.36.475.170.623.2]
(2)

Techne Jurnal Ilmiah Elektrott:knika Vol. l) No l April 20 I 0 Hal 1 . 11

(VHDL) on an Altera cyclone school boy field programmable gate array (FPGA) board

to compute the heart rate.

The block diagram of the system is illustrated in Fig. 2. In the ECG block, a circuit

is designed to obtain the ECG signal from the human body. The output of this ECG block is the input to the FPGA board. The processes to be done within the FPGA board include

tilt: analog lo lhgltai セZオョカエZiセiuh@ v\i.JC j vf the EL \.J signal, i:hc digvmhm iu alt<:i:>LHc ihc heart rate, as well as the digital to analog conversion (DAC) of the signal which marks

the R points. An oscilloscope is used to display the output

Fig. 2. The block diagram of the system

2. ECG CIRCUIT

In general, biological signals are very small. An ECG signal is approximately O.l to 6 mV [1]. Therefore amplification is needed in order to monitor them better. Fig. 3 and Fig. 4 show the block diagram and the circuit diagram of an ECG amplifier. The system

starts with taking the signal from the body by connecting self-adhesive electrodes to

significant part of the body Then, the signal is amplified using an instrumentation amplifier. After that, it goes through a band pass titter that eliminates noise of the input

signal.

Differential セᄋ@ Instrument

Input

f

i

Amplifier

High Pass

Filter

Low Pass

Filter

Fig. 3. ECG amplifier block diagram

ECG

Signal

HEART RAn: cumpcQセT@ TTON lMPLE\cJEVTED ON FJELD proHOraNセimable@ OATE ARRA Y BOAR/) Ivanna K Timofius

An instrumentation amplifier is used to attenuate common biological signal from

the positive and negative inputs produced by not-heart body organs and amplify the ditTerence between the inputs produced by heart. An instrumentation amplifier having a

high common mode rejection ratio (CMRR) is needed for this application. The

instrumentation amplifier circuit used in the セカウエ・ュ@ is shown in Fig. 5 The gain can be

computed using Eq I. To control the CMRR, a variable resistance was placed between

the reference and instnnnentation amp I itier [I ] .

'"

"' • .i Ol

1

11

+\2'W

Q

I'

MMセ@

セ@_t

- _!__

llt

"'

"'

••

+i2<1

Zキセェ@ IT ᄋBMセ@

I

" _ .. ,.,\.,---·---·

-·-Fig. 4. ECG circuit diagram

MセM ..

.,__ ______ セMjᄋOBカG|NMMセセ⦅ャ@ __ NN⦅⦅セ|[|NG|LMMMMMMMMMMMGBBG@

li. loti

4'

I

""'"'

6 ·1211

RIO

C'l

·12vc I

14 _OJ.Al•U I Fill

ャᄋOᄋセMMMlセLLLLセM

·-r '

l--

,y

,,+1-

'1.

\h ECG

Fig. 5. Instmmentation amplifier circuit diagram

(l)

The band pass filter is made out of cascaded high and low pass filter. Since the

[image:2.1054.50.469.219.446.2] [image:2.1054.45.480.363.650.2]
(3)

Techne Jurnal Ilmiah Elektrott:knika Vol. l) No l April 20 I 0 Hal 1 . 11

(VHDL) on an Altera cyclone school boy field programmable gate array (FPGA) board

to compute the heart rate.

The block diagram of the system is illustrated in Fig. 2. In the ECG block, a circuit

is designed to obtain the ECG signal from the human body. The output of this ECG block is the input to the FPGA board. The processes to be done within the FPGA board include

tilt: analog lo lhgltai セZオョカエZiセiuh@ v\i.JC j vf the EL \.J signal, i:hc digvmhm iu alt<:i:>LHc ihc heart rate, as well as the digital to analog conversion (DAC) of the signal which marks

the R points. An oscilloscope is used to display the output

Fig. 2. The block diagram of the system

2. ECG CIRCUIT

In general, biological signals are very small. An ECG signal is approximately O.l to 6 mV [1]. Therefore amplification is needed in order to monitor them better. Fig. 3 and Fig. 4 show the block diagram and the circuit diagram of an ECG amplifier. The system

starts with taking the signal from the body by connecting self-adhesive electrodes to

significant part of the body Then, the signal is amplified using an instrumentation amplifier. After that, it goes through a band pass titter that eliminates noise of the input

signal.

Differential セᄋ@ Instrument

Input

f

i

Amplifier

High Pass

Filter

Low Pass

Filter

Fig. 3. ECG amplifier block diagram

ECG

Signal

HEART RAn: cumpcQセT@ TTON lMPLE\cJEVTED ON FJELD proHOraNセimable@ OATE ARRA Y BOAR/) Ivanna K Timofius

An instrumentation amplifier is used to attenuate common biological signal from

the positive and negative inputs produced by not-heart body organs and amplify the ditTerence between the inputs produced by heart. An instrumentation amplifier having a

high common mode rejection ratio (CMRR) is needed for this application. The

instrumentation amplifier circuit used in the セカウエ・ュ@ is shown in Fig. 5 The gain can be

computed using Eq I. To control the CMRR, a variable resistance was placed between

the reference and instnnnentation amp I itier [I ] .

'"

"' • .i Ol

1

11

+\2'W

Q

I'

MMセ@

セ@_t

- _!__

llt

"'

"'

••

+i2<1

Zキセェ@ IT ᄋBMセ@

I

" _ .. ,.,\.,---·---·

-·-Fig. 4. ECG circuit diagram

MセM ..

.,__ ______ セMjᄋOBカG|NMMセセ⦅ャ@ __ NN⦅⦅セ|[|NG|LMMMMMMMMMMMGBBG@

li. loti

4'

I

""'"'

6 ·1211

RIO

C'l

·12vc I

14 _OJ.Al•U I Fill

ャᄋOᄋセMMMlセLLLLセM

·-r '

l--

,y

,,+1-

'1.

\h ECG

Fig. 5. Instmmentation amplifier circuit diagram

(l)

The band pass filter is made out of cascaded high and low pass filter. Since the

[image:3.1054.51.484.200.606.2]
(4)

Technc Jurna1 Ilnuah Ekktroteknika Vol. lJ No. 1 Aprii201U Hal l- 11

high pass filter is 0.5 Hz and the cutoff frequency of the low pass filter is 100Hz. Fig. 6

and Fig. 7 show the connection of a simple first order inverting active high pass filter

with a unity gain and a simple first order inverting active low pass filter with a unity gain

[2]. Eq. 2 is used to compute for the resistance and capacitance value of the high pass and

low pass filter

(2)

where

a,

is the gain, R, is the resistance value, ('1 is the capacitance value. and}; is the

cutoff frequency. Since the filters are unity gain, the value of R2 IS equal to that of R1 [2].

Fig. 6. First order inverting active high pass filter

Fig. 7. First order inverting active low pass t1lter

Fig. 8 shows the ECG signal generated from the ECG amplifier. Since the ECG

signal of each person differ in strength, the amplifier is made such that the gain of the

instrumentation amplifier may be adjusted to conform to the required signal strength by

the FPGA module.

;/

i

Fig. 8. Resulted ECG signal

HEART/lr1JE ( U/t!IPUTA 1101\. /MPLEilJENTED lJJV FJELD proャOram[セᄋiable@ GATE ARRAY BOARD

Ivanna K. Tnnotius

DI. SYSTEM IMPLEMENTED ON FPGA

The output of the ECG circuit is then teed to a FPGA board. The digital algorithm

implemented on the FPGA aims to point out the R points and calculate the real time hea11

rate. The approach implemented on the FPGA board sta11s with a moving average filter to

rC'Ill(We the hiuh frequcncv nnise Then it followed hv R point finding and the heart riltt>

calculation The detail digital algorithm is shown at Fig. 9

YES

ADC_DATA

..

Moving Average

ᄋᄋセMNQNNNNNNᄋMMᄋᄋᄋᄋMMᄋャ@

(S Points)

---.. ---セBG@ セBBGBM

---1

ADC_D

セ^OPP_@

Gセ@

"-y

!

NO

/A"-._

セtィセZォゥ_^@

Gセ@

NO

セMMセセMG@

Freq Counting Renew t11e Heart Rat& Data

ADC_OO \]HRPPI Qセ@ COUNT= 0

Fig. 9. Digital system flowchart

First, the system clears all register and makes a welcome display on the liquid

[image:4.1054.54.483.225.684.2]
(5)

Technc Jurna1 Ilnuah Ekktroteknika Vol. lJ No. 1 Aprii201U Hal l- 11

high pass filter is 0.5 Hz and the cutoff frequency of the low pass filter is 100Hz. Fig. 6

and Fig. 7 show the connection of a simple first order inverting active high pass filter

with a unity gain and a simple first order inverting active low pass filter with a unity gain

[2]. Eq. 2 is used to compute for the resistance and capacitance value of the high pass and

low pass filter

(2)

where

a,

is the gain, R, is the resistance value, ('1 is the capacitance value. and}; is the

cutoff frequency. Since the filters are unity gain, the value of R2 IS equal to that of R1 [2].

Fig. 6. First order inverting active high pass filter

Fig. 7. First order inverting active low pass t1lter

Fig. 8 shows the ECG signal generated from the ECG amplifier. Since the ECG

signal of each person differ in strength, the amplifier is made such that the gain of the

instrumentation amplifier may be adjusted to conform to the required signal strength by

the FPGA module.

;/

i

Fig. 8. Resulted ECG signal

HEART/lr1JE ( U/t!IPUTA 1101\. /MPLEilJENTED lJJV FJELD proャOram[セᄋiable@ GATE ARRAY BOARD

Ivanna K. Tnnotius

DI. SYSTEM IMPLEMENTED ON FPGA

The output of the ECG circuit is then teed to a FPGA board. The digital algorithm

implemented on the FPGA aims to point out the R points and calculate the real time hea11

rate. The approach implemented on the FPGA board sta11s with a moving average filter to

rC'Ill(We the hiuh frequcncv nnise Then it followed hv R point finding and the heart riltt>

calculation The detail digital algorithm is shown at Fig. 9

YES

ADC_DATA

..

Moving Average

ᄋᄋセMNQNNNNNNᄋMMᄋᄋᄋᄋMMᄋャ@

(S Points)

---.. ---セBG@ セBBGBM

---1

ADC_D

セ^OPP_@

Gセ@

"-y

!

NO

/A"-._

セtィセZォゥ_^@

Gセ@

NO

セMMセセMG@

Freq Counting Renew t11e Heart Rat& Data

ADC_OO \]HRPPI Qセ@ COUNT= 0

Fig. 9. Digital system flowchart

First, the system clears all register and makes a welcome display on the liquid

[image:5.1054.74.449.201.646.2]
(6)

Techne Jurnallltmah Ekktrotckmka Vol. Y No. l Apnl2010 Hal i 11

frequency 2 kHz. and conducts an eight point moving average filtering to remove the high

frequency noise Theoretically, the moving average filtering is made by adding the last 8

data, then dividing the result by 8. However, a divider is avoided to make the system simple. Therefore, we used three right shifts to replace the divider. The moving average system is explained in Fig 10.

ADC_DAT11

<::::§C_AVR_CNT--0?:;>

MMMMNセMᄋMM

NO

_...-!._,

セ@

--- -AVR -

cセ[セQ_^@

.-·

NO

YES ADC BUF1 <= adcセdata@ AOC_AVR_CNT++

I

VES ᄋセN@ ADC BUF2 <= ADC DATA ADC_AVR_ CNT++

ADC 0 <= ADC BUF1 + ADC BUF2 + .. + ADC BUF8

Fig. I 0. Moving average flowchart

After conducting moving average filtering, the algorithm wants to locate the R

points. In the implemented system, a point is considered as an R point if the point is not excessively close to the previous R point, the point is above cettain threshold, and the

point is a maximum point.

The interval between two R points is called by R-R interval. This R-R interval is

counted with the help ofvariable 'count'. After conducting a moving average, the system

waits until variable ·count' is higher than 500 samples. This approach is implemented to

avoid calculation based on excessively close maximum points and to make the overall system faster. Even though the overall ECG information ranged from 0.5 to 100Hz [1],

the human heartbeat rarely goes above 4 Hz or 240 bpm Therefore, we can expect that

;i

there is no ョセキ@ R point in ECG signal within ','4 second (500 samples) after the previous

R point Note that if we expect to count a heartbeat faster than 240 bpm, the constant 500 should be lowered.

'

11£:-tRT RA 1E ( 'OlvJl>(T4110N IMPLE/1-JENTEl> ON FIEU> PROURAMM4.RLE (1:4 TE

ARRAY BOARD

Ivanna K Ti moti us

If the count is already higher than 500 samples, the system checks if the data concerned is above the threshold_ After the concerned data is above certain point, the

system checks whether it is a maximum point_ The applied method for conducting a

maximum point calculation is explained in Fig. ll. A point is considered as a maximum point if this point is the highest point among its 24 neighbor samples The common

approach used to tind a maximum point is called gradient method Howevec the approach applied is has a little different with the gradient method. The gradient method

takes the centre point of Fig. 12a as a maximum point, but do not take the centre point of

Fig. l2b as a maximum point The applied method takes both centre points of Fig. I2a

and Fig. 12b as a maximum point. The applied method was chosen to minimize the effect ofthe high frequency noise.

·.r

---_ ... セMM

'·'-'

....

--

----.

ᄋᄋaサI」⦅PQS\adc⦅セセj@ .. >

セ@·--.--·

MMMMセMN@ --...

ADC D13<AOC

6?

·.>

セ@ セ@ ____ ...

」ッオイMNセエ\セッ[ᄋ@

ADC_DO \セ@ 12'b0

.. ヲイNャAヲLAセou{ャャゥAャ⦅A[AN@

I

l

.. , ADC D23<«ADC 024 ADC_D24<=ADC_O

[image:6.1054.59.487.200.555.2]
(7)

Techne Jurnallltmah Ekktrotckmka Vol. Y No. l Apnl2010 Hal i 11

frequency 2 kHz. and conducts an eight point moving average filtering to remove the high

frequency noise Theoretically, the moving average filtering is made by adding the last 8

data, then dividing the result by 8. However, a divider is avoided to make the system simple. Therefore, we used three right shifts to replace the divider. The moving average system is explained in Fig 10.

ADC_DAT11

<::::§C_AVR_CNT--0?:;>

MMMMNセMᄋMM

NO

_...-!._,

セ@

--- -AVR -

cセ[セQ_^@

.-·

NO

YES ADC BUF1 <= adcセdata@ AOC_AVR_CNT++

I

VES ᄋセN@ ADC BUF2 <= ADC DATA ADC_AVR_ CNT++

ADC 0 <= ADC BUF1 + ADC BUF2 + .. + ADC BUF8

Fig. I 0. Moving average flowchart

After conducting moving average filtering, the algorithm wants to locate the R

points. In the implemented system, a point is considered as an R point if the point is not excessively close to the previous R point, the point is above cettain threshold, and the

point is a maximum point.

The interval between two R points is called by R-R interval. This R-R interval is

counted with the help ofvariable 'count'. After conducting a moving average, the system

waits until variable ·count' is higher than 500 samples. This approach is implemented to

avoid calculation based on excessively close maximum points and to make the overall system faster. Even though the overall ECG information ranged from 0.5 to 100Hz [1],

the human heartbeat rarely goes above 4 Hz or 240 bpm Therefore, we can expect that

;i

there is no ョセキ@ R point in ECG signal within ','4 second (500 samples) after the previous

R point Note that if we expect to count a heartbeat faster than 240 bpm, the constant 500 should be lowered.

'

11£:-tRT RA 1E ( 'OlvJl>(T4110N IMPLE/1-JENTEl> ON FIEU> PROURAMM4.RLE (1:4 TE

ARRAY BOARD

Ivanna K Ti moti us

If the count is already higher than 500 samples, the system checks if the data concerned is above the threshold_ After the concerned data is above certain point, the

system checks whether it is a maximum point_ The applied method for conducting a

maximum point calculation is explained in Fig. ll. A point is considered as a maximum point if this point is the highest point among its 24 neighbor samples The common

approach used to tind a maximum point is called gradient method Howevec the approach applied is has a little different with the gradient method. The gradient method

takes the centre point of Fig. 12a as a maximum point, but do not take the centre point of

Fig. l2b as a maximum point The applied method takes both centre points of Fig. I2a

and Fig. 12b as a maximum point. The applied method was chosen to minimize the effect ofthe high frequency noise.

·.r

---_ ... セMM

'·'-'

....

--

----.

ᄋᄋaサI」⦅PQS\adc⦅セセj@ .. >

セ@·--.--·

MMMMセMN@ --...

ADC D13<AOC

6?

·.>

セ@ セ@ ____ ...

」ッオイMNセエ\セッ[ᄋ@

ADC_DO \セ@ 12'b0

.. ヲイNャAヲLAセou{ャャゥAャ⦅A[AN@

I

l

.. , ADC D23<«ADC 024 ADC_D24<=ADC_O

[image:7.1054.60.464.122.655.2]
(8)

Techne Jurnal1lmmh Eh!ktrotekmka VoL Y No. l Apnl 20 l 0 Hal 1 I I

till+>(nl

(a I

GudiPnt lV[Ptlto!l: xfll- x(O) • 0 xf::!:] - x:f:\) · 0

xf31- x(.!) 0 xf-l)-x(3)' 0

illt+xful

(bl

AJIJdieod Meothod:

x(O) , x[2] x[ll·x(1) x(3J x(ll x(41· x(2]

Fig. 12 Two different maximum point methods

After a point is considered as a maximum point, the system puts some value at the DAC output and calculates the heart rate. This DAC output is used to see whether all R points are poim out corredly. The heart rate calculation is based on the \·ariable 'count'

which indicates the R-R interval. Given that the ADC sampling frequency is 2 kHz, by

using this variable we can count the heart rate (per minutes) using the following equation:

Hem1 Rate (/min) Sampling: Freq (/seer 60

Ll)Ulll

120000

Count

(3)

For displaying the computed heart rate into seven-segment display, first, we need to determine the hundreds, tenth, and ones value of the heart rate. The method for

conducting the frequency-hundreds is explained in Fig. 13, the analogous method is used

for the tenth and ones. The results are sent to the seven-segment display, which is controlled by a finite state machine (FSM) After the computation is done, the variable

count starts from zero again, for the next R-R interval. Then, the data buffer is renewed

for the next moving average and maximum point calculation.

'

I

HEART RATE COllc/PUJATUJN IMPLEME.'VT£[) OI\' FIELD PROUR4MMABLE (i4 TE ARR4Y BlJAR/)

lvonna K Timol!l!S

FREQ

NO

NO

NO

v

I

BC0100<«'1 000"

I •

FS •. FRE010-<=FRE0-800 ,

BC0100<="0000' FREQ1D<=FREQ

セM FR£.()10

Fig. 13. Frequency-hundreds approach

IV. RESULTS AND DISCUSSION

To test the algorithm on the FPGA, we used sine waves from the signal generator

as the FPGA input. First, we used a 2 Hz signal. The result can be seen at Fig. 14 where

output of the DAC pointed out the maximum points correctly and the seven-segment display showed the heart rate 125 bpm.

Fig. 14. 2Hz input, display: 125

The highest heart rate that this system can handle is 4 Hz or 240 bpm The results

can be seen at Fig. 15. This limitation is due to the approach that the minimum R-R

[image:8.1054.183.365.69.198.2]
(9)

Techne Jurnal1lmmh Eh!ktrotekmka VoL Y No. l Apnl 20 l 0 Hal 1 I I

till+>(nl

(a I

GudiPnt lV[Ptlto!l: xfll- x(O) • 0 xf::!:] - x:f:\) · 0

xf31- x(.!) 0 xf-l)-x(3)' 0

illt+xful

(bl

AJIJdieod Meothod:

x(O) , x[2] x[ll·x(1) x(3J x(ll x(41· x(2]

Fig. 12 Two different maximum point methods

After a point is considered as a maximum point, the system puts some value at the DAC output and calculates the heart rate. This DAC output is used to see whether all R points are poim out corredly. The heart rate calculation is based on the \·ariable 'count'

which indicates the R-R interval. Given that the ADC sampling frequency is 2 kHz, by

using this variable we can count the heart rate (per minutes) using the following equation:

Hem1 Rate (/min) Sampling: Freq (/seer 60

Ll)Ulll

120000

Count

(3)

For displaying the computed heart rate into seven-segment display, first, we need to determine the hundreds, tenth, and ones value of the heart rate. The method for

conducting the frequency-hundreds is explained in Fig. 13, the analogous method is used

for the tenth and ones. The results are sent to the seven-segment display, which is controlled by a finite state machine (FSM) After the computation is done, the variable

count starts from zero again, for the next R-R interval. Then, the data buffer is renewed

for the next moving average and maximum point calculation.

'

I

HEART RATE COllc/PUJATUJN IMPLEME.'VT£[) OI\' FIELD PROUR4MMABLE (i4 TE ARR4Y BlJAR/)

lvonna K Timol!l!S

FREQ

NO

NO

NO

v

I

BC0100<«'1 000"

I •

FS •. FRE010-<=FRE0-800 ,

BC0100<="0000' FREQ1D<=FREQ

セM FR£.()10

Fig. 13. Frequency-hundreds approach

IV. RESULTS AND DISCUSSION

To test the algorithm on the FPGA, we used sine waves from the signal generator

as the FPGA input. First, we used a 2 Hz signal. The result can be seen at Fig. 14 where

output of the DAC pointed out the maximum points correctly and the seven-segment display showed the heart rate 125 bpm.

Fig. 14. 2Hz input, display: 125

The highest heart rate that this system can handle is 4 Hz or 240 bpm The results

can be seen at Fig. 15. This limitation is due to the approach that the minimum R-R

[image:9.1054.48.480.60.638.2]
(10)

Jurnal Ill inh EJ J I - I J

be detected, as shown in Fig. 15b. Therefore, the heart rate display is lower than the

actuaJ frequency. For a signal above lOO Hz, no maximum point is detected as shown in

Fig. L Sc. This happened because the actuaJ maximum points are not the highest among its

24 neighbor sa mpl es.

W

M

セ@

Fig. 15. (a) Hz input. display: 23 J (b) J 0 Hz input, display: 194 (c) 100Hz input

In the implemented system, a signal is consider as a maxi mum point if the signal

magnitude is higher than 80 mV as shown in Fig. 16a. If the signal is lower than 80 mV,

no maximum point is detected as shown in Fig. 16b.

(a) (b)

Fig.

o.

(')2Hz, J 00 mV input, display: 123 (b) 2Hz, 80 mV input, displa : 26

T

es

1

o ll yst • we connect the ECG circuit to uman and the result is

show in Fig. 7. As sho\ in the figure, all R p ints re detected co rr I . Tl e di pia

shown in the seven-segment display indicates the real time heart rate.

'

'

HEART RATE 0 IPUTA.TION l fPLEMEN1 DON FIE D PROGRAM ,(]JL£ GATE ARRA BOARD

1 anna K Timotius

Fig. 17. Detected R points of a ECG signal

Thi paper describ done os ible method to detect electrocardiogram (ECG) i nal.

count the R-R interval of the ECG signal, and compute the heart rate. In the fu ture, an

algorithm to detect the other points and give some information regarding the heart rate

ariability (HRV)

will

compl t this research.

ACKNOW L EDG E:MENT

The author would to acknowledge Wei-Chih Hu, Ph.D . (Chung Yuan Christian

University, Taiwan) fo r lending me FPGA board and electronic components, Patricia

Angela Abu for t he heJp in understanding FPGA and Quartus software, Kimberly Jane

Uy for making the ECG circuit, and many other friends in their help.

REFERENCES

I. "Basic prin iples of design an ECG amplifier - sensitivity otentials]; frequency

response; common mode ejection ratio," http:// perfusion com.au

2. "Active Filter Design Techni ues, Literature Number SLOA088," Texas Instrument

3. Altera Cyclone choolboy nuaJ

[image:10.1051.14.502.36.712.2]
(11)

Jurnal Ill inh EJ J I - I J

be detected, as shown in Fig. 15b. Therefore, the heart rate display is lower than the

actuaJ frequency. For a signal above lOO Hz, no maximum point is detected as shown in

Fig. L Sc. This happened because the actuaJ maximum points are not the highest among its

24 neighbor sa mpl es.

W

M

セ@

Fig. 15. (a) Hz input. display: 23 J (b) J 0 Hz input, display: 194 (c) 100Hz input

In the implemented system, a signal is consider as a maxi mum point if the signal

magnitude is higher than 80 mV as shown in Fig. 16a. If the signal is lower than 80 mV,

no maximum point is detected as shown in Fig. 16b.

(a) (b)

Fig.

o.

(')2Hz, J 00 mV input, display: 123 (b) 2Hz, 80 mV input, displa : 26

T

es

1

o ll yst • we connect the ECG circuit to uman and the result is

show in Fig. 7. As sho\ in the figure, all R p ints re detected co rr I . Tl e di pia

shown in the seven-segment display indicates the real time heart rate.

'

'

HEART RATE 0 IPUTA.TION l fPLEMEN1 DON FIE D PROGRAM ,(]JL£ GATE ARRA BOARD

1 anna K Timotius

Fig. 17. Detected R points of a ECG signal

Thi paper describ done os ible method to detect electrocardiogram (ECG) i nal.

count the R-R interval of the ECG signal, and compute the heart rate. In the fu ture, an

algorithm to detect the other points and give some information regarding the heart rate

ariability (HRV)

will

compl t this research.

ACKNOW L EDG E:MENT

The author would to acknowledge Wei-Chih Hu, Ph.D . (Chung Yuan Christian

University, Taiwan) fo r lending me FPGA board and electronic components, Patricia

Angela Abu for t he heJp in understanding FPGA and Quartus software, Kimberly Jane

Uy for making the ECG circuit, and many other friends in their help.

REFERENCES

I. "Basic prin iples of design an ECG amplifier - sensitivity otentials]; frequency

response; common mode ejection ratio," http:// perfusion com.au

2. "Active Filter Design Techni ues, Literature Number SLOA088," Texas Instrument

3. Altera Cyclone choolboy nuaJ

[image:11.1051.5.495.29.767.2]

Gambar

Fig. 1. ECG signal
Fig. 2. The block diagram of the system
Fig. 5. Instmmentation amplifier circuit diagram
Fig. 8. Resulted ECG signal
+7

Referensi

Dokumen terkait

Teknik ini peneliti lakukan bersama teman-teman yang juga melakukan penelitian selama proses pengumpulan data agar data yang diperoleh sesuai dengan masalah yang

bahwa berdasarkan pertimbangan sebagaimana dimaksud pada huruf a dan b, perlu menetapkan Peraturan Menteri Keuangan tentang Modalitas Penurunan Tarif Bea Masuk Dalam Rangka

Demikian Pengumuman Pemenang Pelelangan ini dibuat dan ditandatangani pada hari, tanggal dan bulan sebagaimana tersebut di atas untuk dipergunakan sebagaimana

[r]

Selain memiliki banyak sungai, wilayah Kalimantan Barat juga banyak memiliki pulau besar dan kecil yang tersebar di sepanjang selat karimata dan laut natuna yang

Dengan pertimbangan ingin meningkatkan jenjang pendidikan sampai tingkat sarjana (S1), pada tahun 1981 ATN Malang dikembangkan menjadi Institut Teknologi Nasional

Sehingga studi yang digunakan dalam penelitian ini adalah studi deskriptif kualitatif untuk menggambarkan persepsi masyarakat Surabaya sesuai dengan proses penerimaan

Sehingga umumnya investor yang akan menanamkan modalnya dalam suatu perusahaan akan melakukan analisa terhadap kinerja keuangan perusahaan dengan melihat faktor