• Tidak ada hasil yang ditemukan

Microfluidic Technologies for High Throughput

N/A
N/A
Protected

Academic year: 2023

Membagikan "Microfluidic Technologies for High Throughput"

Copied!
180
0
0

Teks penuh

Finally, I would like to thank all the graduate and postdoctoral students who worked with me for their help in several situations where I just didn't have the technical expertise to go far on my own. Several platforms have emerged in the past few years to partition or subdivide chemical libraries, including 3456-well high density.

Organization 3

Introduction 5

For fabrication processes where the polymer is cast or injection molded, Tg is a very important parameter. As the temperature rises above Tg, it is in a viscoelastic state and can easily flow12.

Specific Polymer Properties 8

Silicone-based Polymers 8

The Young's modulus of PDMS depends on the degree of cross-linking between linear silicone chains. At this point, the wafer has a positive photoresist pattern of the channels on the surface.

Figure 2.2: Peroxide-based mechanism for PDMS cross-linking
Figure 2.2: Peroxide-based mechanism for PDMS cross-linking

Diene-based Polymers 19

This problem was resolved by lowering the catalyst/monomer ratio of the mixture from 1:24,000 to. No swelling was observed for water and all alkanes (C7 - C16) in bulk tests where solvent was pipetted onto the surfaces of the copolymers.

Figure 2.6: Ruthenium-based Grubb
Figure 2.6: Ruthenium-based Grubb's catalyst for ROMP reactions

Polyurethane-based Polymers 25

However, the linear nature of elastomeric chemical bonds makes it more susceptible to the uptake of water and other hydrophilic compounds. Curing in the presence of oxygen (Ebecryl 270 with 1% Irgacure catalyst (Ciba-Geigy)) resulted in a tacky, uncured film on the top surface of the.

Figure 2.9: Urethane linkage diagram formed by the addition polymerization of a  diisocyanate and a dialcohol group
Figure 2.9: Urethane linkage diagram formed by the addition polymerization of a diisocyanate and a dialcohol group

Conclusions 31

Using this technology, we created Ebecryl 270 microfluidic devices consisting of a thick layer (~5 mm) of molded channels connected to a #1 coverslip coated with a thin layer (~20 µm) of Ebecryl 270.

Introduction 33

When thinking about a generic design for microfluidic HTS devices, two different processing schemes become apparent: serial and parallel. Mechanically, throughput depends on factors such as flow rate, sample concentration, and detector acquisition time.

Microencapsulation 35

Bulk Emulsions 37

All preparations contained a heterogeneous size distribution of micelles, with the largest size distribution observed in the mineral oil-based preparation. Problems encountered during the experiment that contributed to the polydispersity included the evaporation of the decane in the continuous phase (~25% of the total volume over 10 minutes at 4°C) and the roughness of the surface of the capillary point.

Figure 3.2: Light microscope images (40x, NA 1.2) of reverse micelles in (a) heptane,  (b) decane, and (c) light mineral oil (from left to right)
Figure 3.2: Light microscope images (40x, NA 1.2) of reverse micelles in (a) heptane, (b) decane, and (c) light mineral oil (from left to right)

Microfluidic Crossflow 43

By lowering the surface tension between oil and water, surfactant forced the oil up into the water channel (Figure 3.6), disrupting the pressure balance of the system. In an effort to increase the frequency of droplet generation in the devices, pumping the water into the oil phase at the T-junction was explored. When the water stream closest to the oil inlet entered the T-junction, it blocked the oil flow to the outlet.

Figure 3.5: First microfluidic crossflow  chip design (top). Channel widths had a  single fixed valve on different prototype  masks ranging from ~50 - 100 µm
Figure 3.5: First microfluidic crossflow chip design (top). Channel widths had a single fixed valve on different prototype masks ranging from ~50 - 100 µm

Microfluidic Crossflow: Theory and Fluid Mechanics 56

Droplet formation is accomplished by high shear forces generated at the leading edge of the water perpendicular to the oil flow, generating picoliter-scale droplets. The self-organization of the inverted vesicles depends on the differential pressure between the water and oil surfactant phases, with higher relative water pressures driving the formation of increasingly complex droplet arrays (Figure 3.17). At water pressures that slightly exceed the oil pressure (Pw>Po), the packing density of the droplets in the 60 µm channel increases.

Figure 3.16: Reverse vesicles in square channels. Photomicrographs show the transition  from the 30 µm wide channel to the 60 µm wide channel
Figure 3.16: Reverse vesicles in square channels. Photomicrographs show the transition from the 30 µm wide channel to the 60 µm wide channel

A Variation on a Theme: In-line Droplet Generation 65

The pattern of droplet formation in the central cavity is complex, depending on the channel geometry, the relative fluid pressures, and the viscosities of the oil and water phases. At higher λ values ​​in the in-line microfluidic device, wire retraction appears to dominate and propagation of the Rayleigh instability is not observed. In an in-line microfluidic device using n-hexane as the discontinuous phase and dimethylformamide as the bulk phase, λ=0.3, monodisperse droplets are continuously produced at the end of the hexane flow with a regular periodicity (Figure 3.21).

Figure 3.19: In-line microfluidic droplet  generating device design.
Figure 3.19: In-line microfluidic droplet generating device design.

Droplet Sorting 70

Optical Valves 71

Sorting efficiency was both measured by looking at the PMT output values ​​of the output material in the collection pits. Excess air trapped in the chip as a result of the filling process is outgassed through the elastomeric silicone. Motorized stage components (DynaOptics Motion) move the microfluidic chip mounted in the stage under the objective lens of the.

Figure 3.23: Central board layout of optical valve system. The output beam from the  808nm TO3 is shaped by the aspheric lens, while the pair of achromat lenses minimizes  beam expansion
Figure 3.23: Central board layout of optical valve system. The output beam from the 808nm TO3 is shaped by the aspheric lens, while the pair of achromat lenses minimizes beam expansion

Multilayer Soft Lithography: Elastomeric Valves 78

Detection: Development of a Fluorescence-Activated Droplet Sorter 85

The first sorting devices used electroosmotic flow, in which the net fluid flow in the channel is driven by the migration of the positive ions in response to an electric field. As a functional demonstration of the device, the fluorescence-activated droplet sorter was used to sort 2-3 µm diameter droplets filled with either green (fluorescence) or red (R-phycoerythrin) fluorescent dyes. To sort the mixed droplet population, a detection region ~100 µm upstream of the T-junction near the device outlets was set up (Figure 3.33).

Figure 3.30: Microfluidic droplet sorter.
Figure 3.30: Microfluidic droplet sorter.

Conclusions 92

Visual inspection of the output wells was performed under a 40x objective (0.65NA) on an inverted microscope (Olympus IX50) equipped with a mercury lamp. Of the thousands of sorted droplets in each well, only a small fraction (~1-2%) was of the wrong dye type. With the development of the fluorescence-activated droplet sorter, an integrated microfluidic screening system emerged.

Introduction 94

The first report of enzymatic activity in inverted vesicles was described in 1992 using polyphenol oxidase87, where Vmax was comparable to that in water and twice that shown in inverted micelles88. More recently, Tawfik and Griffiths reported the use of inverted vesicles for the transcription/translation of HaeIII methyltransferase, using substrate-labeled genetic material to specifically isolate coding DNA strands from a tenfold excess of genes encoding another enzyme38. In this section, the development and characterization of a model enzyme system in inverted vesicles generated by microfluidic crossflow using recombinant p-nitrobenzyl (pNB) esterase expressed in E.

Principle and Design 95

Under monodisperse droplet generation conditions, the PMT-based detector system can be used not only to compare endpoint activity between individual droplets at a fixed position in the outflow channel, but also to obtain single-cell kinetic data for the enzyme. The length and width of the channel where the droplets emerge from the cross-flow junction were varied to create a long exit path that served as an incubation chamber for the encapsulated enzyme and substrate. With adequate monodisperse droplet spacing at the crossflow junction, water droplets in mineral oil/2% range psi w/o-s) traveled as a fully coherent stream from the crossflow junction to the outflow well (Figure 4.2).

Figure 4.1: Microfluidic  channel layout in a
Figure 4.1: Microfluidic channel layout in a

Model Enzyme System: P-Nitrobenzyl Esterase in E. coli 97

The culture temperature was then raised to 42°C for an additional 2.5 hours to induce synthesis of the recombinant esterase protein. After induction, the cells were spun down at 735 x g, washed twice in PBS pH 7.0 to remove residual external enzyme, and resuspended in ice-cold PBS in preparation for the on-chip assay. A 0.5 mM stock of the substrate, fluorescein diacetate (FDA), was prepared by dissolving 2 mg of FDA in 10 ml of acetone.

Figure 4.3: The conversion of fluorescein diacetate to fluorescein (λ ex  = 490 nm,   λ em (max)  = 518 nm)
Figure 4.3: The conversion of fluorescein diacetate to fluorescein (λ ex = 490 nm, λ em (max) = 518 nm)

Encapsulation and Assay Mechanics 99

When the inlet pressure of the substrate and cell solutions was balanced, very little mixing was observed between the streams under the laminar flow conditions present in the microfluidic device. As the points were cut at the junction, the cells were encapsulated with the substrate mixture (Figure 4.4). Spots containing no cells are non-fluorescent while spots containing one or more cells expressing the enzyme have converted the substrate to fluorescent product.

Figure 4.4: Encapsulated bacteria in  droplets generated by microfluidic  crossflow.
Figure 4.4: Encapsulated bacteria in droplets generated by microfluidic crossflow.

Droplet Assay Troubleshooting 101

Flow Balance 102

Once cross-contaminated, background fluorescence levels in the droplets increased dramatically as the substrate was rapidly converted to product prior to encapsulation at the cross-flow junction. Micron-sized bacterial cells were difficult to see at the low cell concentration used in the assay, making it extremely difficult to balance the aqueous inputs at the cross-flow junction. Modifications to the channel geometry were made at the cross-flow junction to increase the backflow resistance between the two aqueous streams.

Figure 4.6: Flow balance at the
Figure 4.6: Flow balance at the

Cell Adhesion to PDMS 104

Bovine serum albumin (BSA) and the nonionic detergent Tween 20 were investigated as water-soluble blocking agents in a diluted mixture of cells. The inhibitory effects of the blocking agents were confirmed in mass spectrofluorometer tests (Table 4.1). HCl-treated Sylgard devices were constructed by injecting a solution of dilute HCl, pH 2.7, into two-layer cross-flow serpentine devices that filled all channels.

Table 4.1: Surfactant and additive effects on whole cell E. coli expressing recombinant  pNB esterase
Table 4.1: Surfactant and additive effects on whole cell E. coli expressing recombinant pNB esterase

Substrate Autohydrolysis 107

The hydrolyzed resorufin product also adhered to the input channel of the substrate, staining it with a bright fluorescent orange residue.

Conclusions 108

Introduction 109

Microfluidic Large Scale Integration 110

These integrated microfluidic networks were used to fabricate a microfluidic analogue of the comparator and a microfluidic memory storage device whose behavior is similar to random access memory (RAM). Microfluidic systems have been used to demonstrate a diverse range of biological applications, including biomolecular separations. In this chapter, we describe microfluidic multiplexing technology and show how it can be used to fabricate silicon devices with thousands of valves and hundreds of individually addressable reaction chambers.

Microfluidic Multiplexors 112

To load the sample into the array area, a set of barrier valves are closed to isolate the array from the rest of the flow channel network. Vertical valves pictured are part of the non-actuated array sandwich valve (not visible in part A for simplicity). An illustration of the multiplexer addressing process is shown in Figure 5.5, where every other row of the matrix is ​​purged sequentially.

Figure 5.2: 1024 well serpentine chip schematic.
Figure 5.2: 1024 well serpentine chip schematic.

Microfluidic Memory Storage Device 118

Individual chamber addressing is achieved via flow channels that run parallel to the sample chambers and use pressurized fluid under the control of the row and column multiplexers to flush the chamber contents to the exit. This device adds a significant level of complexity to previous microfluidic conduits, because there are levels of control: the multiplexors operate valve actuation conduits, which in turn operate the valves themselves. To demonstrate the functionality of the microfluidic memory storage device, we loaded the central memory storage chambers of each row with dye (2.4 mM bromophenol blue in 0.1 M sodium citrate buffer, pH 7.2) and proceeded to purify individual chambers .

Figure 5.7: Mechanics of a single chamber purge within a single row of the microfluidic  memory chip
Figure 5.7: Mechanics of a single chamber purge within a single row of the microfluidic memory chip

Microfluidic Comparator 121

We used this chip as a microfluidic comparator that measures the expression of a specific enzyme using a fluorogenic substrate system to provide an amplified output signal in the form of a fluorescent product. An electronic comparator circuit is designed to provide a large output signal when the input signal exceeds a reference threshold value. The amplified output signal was dependent only on the number of CCP-expressing cells in an individual chamber (Figure 5.13).

Figure 5.10: Comparator chip mechanics in chambers of single column. Elastomeric  valves enable each of the 256 chamber on the chip can be independently
Figure 5.10: Comparator chip mechanics in chambers of single column. Elastomeric valves enable each of the 256 chamber on the chip can be independently

Conclusions 126

Introduction 127

In Vitro Protein Synthesis in Picoliter Volumes 128

The picoliter-sized compartments can be used as reaction vesicles for in vitro transcription/translation of protein libraries or to monitor protein expression and activity at the level of single cells. A strong fluorescent green signal was observed in the chip containing the eGFP DNA construct (Figure 6.1), while no signal was observed with the negative control chip. Although the detection of eGFP using a bacterial extraction system in the described experiment was not sensitive to only one molecule of DNA/compartment, it is not unreasonable to believe that this goal can be achieved in the near future with an array-based microfluidic device . .

Figure 6.1: eGFP transcription/translation in the array-based 1024 well serpentine  microfluidic device with ~2 x 10 5  eGFP DNA constructs/ 80 pL well (10 second  exposure, SBIG-7 cooled CCD camera)
Figure 6.1: eGFP transcription/translation in the array-based 1024 well serpentine microfluidic device with ~2 x 10 5 eGFP DNA constructs/ 80 pL well (10 second exposure, SBIG-7 cooled CCD camera)

Detection Systems for Array-Based Microfluidic Chips 130

1024-well serpentine array chips were used to analyze the enzymatic activity of the wild-type CCP and mutant libraries at the single-cell level. The histogram plot of the background-subtracted wild-type and mutant CCP single-cell activity levels in the microarray chip (Figure 6.8) is comparable to the bulk microtitre. Wild-type single cell CCP assays (background subtracted) in divided single compartments (~300 pL) of the comparator microfluidic chip showed similar C.V.

Figure 6.2: XY scanner stage layout. Motorized stage components (DynaOptics Motion)   move the microfluidic chip mounted in the stage under the objective lens of the
Figure 6.2: XY scanner stage layout. Motorized stage components (DynaOptics Motion) move the microfluidic chip mounted in the stage under the objective lens of the

Cytochrome c Peroxidase: Array-based Enzyme Library Screening 135

CCP Expression System 136

A set of primers was designed to clone the gene (forward: 5' - CGG CGA CAT ATG AAA ACG CTC GTT CAT G - 3'; reverse: 5' - GAG CTC GAA TTC CTA TAA ACC TTG TTC CTC TAA - 3'). of the original vector and introduce unique restriction sites Nde I and Eco RI (underlined). Using a set of primers, the gene was amplified from the pT7CCP vector by PCR (polymerase chain reaction), restricted, and ligated into the multiple cloning site of the expression vector, pET26b+ (Novagen). The polymerase is under the control of the lacUVA5 promoter, and expression is induced by the chemical isopropylthiogalactoside (IPTG).

Random Mutagenesis: Library Construction 137

A plasmid library was isolated from resuspended cells using a commercial kit (Qiagen, Qiaprep Spin Kit) and used to transform E.

CCP Library Assay: Bulk vs. Microfluidic 138

An activity profile was plotted comparing wild-type CCP and mutant enzyme cultures (Figure 6.4). Statistical analysis of the two populations revealed a rather narrow distribution of the activity profile for the wild-type CCP enzyme, while the mutant population had a broad distribution with a significant fraction of inactive clones. Statistical analysis of the serpentine chip array assays revealed activity profiles for wild-type CCP enzyme and the mutant library that were quite different from the respective microtiter plate profiles (Table 6.1).

Figure 6.4: Activity profile of wild-type cytochrome c peroxidase (CCP) single colonies  vs
Figure 6.4: Activity profile of wild-type cytochrome c peroxidase (CCP) single colonies vs

Conclusions 146

The variability of the expression level depends not only on the amino acid sequence of the recombinant enzyme, but also on the content of the enzyme at the individual cell level and the stage of the cell in its growth cycle. However, array-based chips that can rapidly screen thousands of bacteria can identify "super" mutants in a large library of mutagenized recombinant proteins that may never be found using techniques that sample only a small fraction of the library's diversity. While much research is still in the proof-of-concept stage, this technology has the potential to transform chemical analysis in the same way that miniaturization transformed computing, making tools smaller, more integrated, cheaper and much faster130. .

Table 6.3: Microfluidic comparator chip assay (1 hour incubation at 22°C; Substrate -  100 µm Amplex Red, 880µm H 2 O 2  in PBS pH 7.2)
Table 6.3: Microfluidic comparator chip assay (1 hour incubation at 22°C; Substrate - 100 µm Amplex Red, 880µm H 2 O 2 in PBS pH 7.2)

Structure of repeating PDMS subunits. 8

Peroxide-based mechanism for PDMS cross-linking. 9

Platinum-based mechanism for PDMS cross-linking. 9

Fabrication of one-layer PDMS devices fabricated from a wet-etched 12

Dicyclopentadiene polymerization mechanism. 20

Clamping apparatus for solvent testing in polyDCDP microfluidic device. 22

Urethane linkage diagram formed by the addition polymerization of a 26

Light microscope images of reverse micelles in various oils. 39

Basic stages of crossflow membrane emulsification. 41

Experimental setup for capillary-based crossflow (w/o) emulsification 42

Gambar

Figure 2.3: Platinum-based mechanism for PDMS cross-linking
Figure 2.6: Ruthenium-based Grubb's catalyst for  ROMP reactions
Figure 2.8: Clamping apparatus for  solvent testing in polyDCDP  microfluidic device.
Figure 2.9: Urethane linkage diagram formed by the addition polymerization of a  diisocyanate and a dialcohol group
+7

Referensi

Dokumen terkait

This examination has long been performed by using X-ray, but is now increasingly being replaced by CT-scan The objective of this study is to assess the radiation doses