Development of insulated isothermal PCR for rapid on-site malaria detection Abstract

Background Detection of Plasmodium spp. is sometimes inconvenient especially in rural areas that are distant from a laboratory. In this study a portable diagnostic test of Plasmodium spp. was developed using insulated isothermal polymerase chain reaction (iiPCR) as an alternative approach to improve this situation. Methods A pair of universal primers and probe were designed to amplify and detect gene encoding 18S small sub- unit rRNA of Plasmodium spp using iiPCR method in a portable device, POCKIT™. The efficiency and detection limit of the assay were evaluated using quantitative real-time polymerase chain reaction (qPCR) approach before being subjected to testing in POCKIT™. Detection results of POCKIT™ were displayed as ‘+’, ‘−’ or ‘?’ based on the fluorescence ratio after/before reaction. A total of 55 and 35 samples from malaria patients and healthy subjects, respectively, were screened to evaluate the feasibility of this newly designed iiPCR assay. Results The iiPCR assay allowed the detection of various species of Plasmodium, including those infecting humans (Plasmodium falciparum, P.

vivax, P. knowlesi, P. malariae, P. ovale), monkeys, birds, and rodents. Efficiency of the assay achieved 96.9 % while the lower detection limit was ≥100 copies of plasmodial DNA. Specificity of the assay was assured as it could not detect human, bacterial and other parasitic DNA. Among the 55 clinical samples tested, 47 (85.4 %) of them were detected as positive by POCKIT™. Four (7.3 %) samples with fluorescence ratio after/before reaction of <1.2 were reported as negative while another four (7.3 %) were ambiguously detected as they had fluorescence ratios between 1.2 and 1.3. The fluorescence ratio was not found to be associated with the copy number of plasmodial DNA. This approach can only be considered as a qualitative method. Conclusions The portable iiPCR system may serve as an alternative approach for preliminary screening of malaria in endemic rural areas. The system may also be useful for detecting animal malaria in the field. Although it is not as quantitative as qPCR method, it is comparatively fast and easy to handle. It is believed that the POCKIT-iiPCR assay is able to achieve 100

% sensitivity if increased amount of DNA from each sample is used. The iiPCR assay can

(2)

also be upgraded in future to detect multiple Plasmodium spp. at the same time by designing the specific primers and probes.

Referensi

Dokumen terkait

coli 42 2.2 Designation, selection, and determination of the optimal conditions of TaqMan-based real-time PCR 43 2.2.1 Primers, probe designation and selection 43 2.2.2 Determination

https://doi.org/10.1007/s00203-020-01903-1 ORIGINAL PAPER Challenging TaqMan probe‑based real‑time PCR and loop‑mediated isothermal amplification LAMP: the two sensitive molecular

However, in this research we developed a quantitative probe based one-step rRT-PCR assay for detection and viral load determination of the CCHF virus in suspected blood samples of the

In addition, the potential of directly using proteinase K for RT-LAMP reaction with different volumes, i.e., 1, 3, and 5 μL, was also evaluated as shown in Figure 11A6, Figure