• Tidak ada hasil yang ditemukan

ALAT UKUR KAPASITAS VITAL PAKSA (KVP) BERBASIS MICROCONTROLLER ATMega16

N/A
N/A
Protected

Academic year: 2017

Membagikan "ALAT UKUR KAPASITAS VITAL PAKSA (KVP) BERBASIS MICROCONTROLLER ATMega16"

Copied!
2
0
0

Teks penuh

(1)

ALAT UKUR KAPASITAS VITAL PAKSA (KVP)

BERBASIS

MICROCONTROLLER

ATMega16

Prio Agung Pamuji

1

, Nur Hudha Wijaya

2

, Heri Purwoko

3

Program Studi Teknik Elektromedik, Program Vokasi

Universitas Muhammadiyah Yogyakarta

JL. Lingkar Selatan, Tamantirto, Kasihan Bantul Yogyakarta 55183

Email :

[email protected], [email protected]

INTISARI

Volume dan kapasitas paru-paru manusia dipengaruhi oleh beberapa faktor, diantaranya: usia, jenis kelamin, faktor penyakit dan aktifitas seseorang. Ukuran badan seseorang dapat mempengaruhi kapasitas paru-paru, dimana seseorang dengan tubuh besar memiliki paru-paru yang besar. Kapasitas paru-paru ditentukan juga oleh luas permukaan tubuh untuk pertukaran gas. Pada penelitian ini sensor tekanan gas MPX5100GP diaplikasikan dalam alat ukur kapasitas vital paksa (KVP) paru-paru, microcontroller ATMega16 sebagai pengendali utama dan LCD sebagai tampilannya. Keunggulan dari sensor tekanan gas MPX5100GP yaitu hanya dengan tiupan kecil saja sudah mempengaruhi tegangan keluarannya. Peneliti melakukan perbandingan antara alat yang dibuat dengan spirometry udara. Sinyal respirasi yang diberikan pada sensor akan di konversi menjadi nilai kapasitas vital paksa (KVP) dalam bentuk angka digital dan ditampilkan pada layar LCD karakter 2x16. Setelah melalui proses perbandingan, kemudian alat digunakan untuk menguji kapasitas vital paru-paru seseorang. Berdasarkan penelitian yang dilakukan didapatkan data pengukuran yang berubah-ubah, hal ini disebabkan karena setiap pengukuran yang dilakukan, tekanan yang diberikan oleh relawan selalu berbeda. Setelah melalui proses perbandingan alat dan analisis data, didapatkan kesalahan pengukuran tertinggi alat ukur kapasitas vital paksa (KVP) paru-paru sebesar 3.8% dan kesalahan rata-rata terendah sebesar 0.01%.

(2)

MEASUREMENT TOOLS OF FORCE VITAL CAPACITY (FVC)

BASED ON MICROCONTROLLER ATMega16

Prio Agung Pamuji

1

,

Nur Hudha Wijaya

2

,

Heri Purwoko

3

Program Studi Teknik Elektromedik, Program Vokasi

Universitas Muhammadiyah Yogyakarta

JL. Lingkar Selatan, Tamantirto, Kasihan Bantul Yogyakarta 55183

Email :

[email protected], [email protected]

ABSTRACT

The volume and capacity of human lungs is influenced by several factors, including: age, gender, disease factors and activities of a person. Body size of someone can affect the volume of the lungs, where a person with a large body has a large lung. The volume of the lungs is also determined by the body surface area for gas exchange. In this study the MPX5100GP gas pressure sensor was applied in measurement tool of lung capacity force using microcontroller ATMega16 as main controller and LCD as its appearance. The advantages of the MPX5100GP gas pressure sensor is that only a small blow will affecting the output voltage. Researchers do a comparison between the tools that created by researchers with Spirometry air. The respiratory signal given to the sensor will convert become the value of vital capacity of the lungs force in the form of digital numbers and displayed on the LCD screen using 2x16 character. After going through the process of comparison, then the tool is used to test the vital capacity of a person's lungs force. Based on the research, the data obtained is changing, this is because every measurment that did by the researcher, the pressure given by someone are the different. After the researcher comparing the tools and data analyzed the researcher get the highest error measurement tools of lung capacity force is 3.8% and the lowest error is 0.01%

Referensi

Dokumen terkait

Berdasarkan hasil penelitian dan pembahasan, makadapat disimpulkan sebagai berikut: Nilai kapasitas vital paru dan kapasitas vital paksa pada qori’ lebih tinggi daripada non

Tujuan penelitian ini adalah untuk mengetahui hubungan antara indeks massa tubuh dan volume ekspirasi paksa detik 1 (VEP1) / kapasitas vital paksa (KVP) pada pasien PPOK stabil

Kesimpulan : Terdapat hubungan yang bermakna antara Indeks Massa Tubuh (IMT) dan Volume Ekspirasi Paksa Detik 1 (VEP1) / Kapasitas Vital Paksa (KVP) di

Hasil penelitian menyatakan terdapat perbedaan nilai rerata Kapasitas Vital Paksa (KVP) yang bermakna (p < 0,05) antara mahasiswa yang berolahraga teratur dan

Hubungan antara Indeks Masa Tubuh (IMT) dengan Kapasitas Vital Paksa (KVP) pada Pasien PPOK Stabil Derajat III di Balai Besar Kesehatan Paru Masyarakat.. (BBKPM)

Pada pengukuran nilai Kapasitas Vital Paksa (KVP) Paru pada kelompok bukan perokok dengan menggunakan alat spirometri didapatkan responden yang mempunyai KVP terendah berjumlah

Berdasarkan hasil penelitian dan pembahasan, makadapat disimpulkan sebagai berikut: Nilai kapasitas vital paru dan kapasitas vital paksa pada qori’ lebih tinggi daripada

Diajukan Kepada Program Vokasi Universitas Muhammadiyah Yogyakarta untuk.. Memenuhi Sebagian Persyaratan Guna Memperoleh Gelar Ahli Madya (A.Md.) Program Studi D3