Secondary Electrons

X- Ray Intensity Emitted from Thick, Solid Specimens

4.5 X-Ray Fluorescence

59

4

s d= cscψ (4.15)

Normalizing by the intensity generated in each layer, the ϕ(ρz) histogram gives the probability, with a value between 0 and 1, that a photon generated in that layer and emitted into the solid angle of the EDS detector will escape and reach the detector, as shown in . Fig. 4.16b for each histogram bin of the silicon ϕ(ρz) distribution. The escape probability of X-rays integrated over the complete ϕ(ρz) histogram gives the parameter designated “f(χ),” which is the overall escape probability, between 0 and 1, for an X-ray generated any- where in the ϕ(ρz) distribution.

. Figure 4.21b shows a sequence of calculations of the C K ϕ(ρz) distribution and subsequent absorption as a func- tion of incident beam energy. As the incident beam energy increases, the depth of electron penetration increases so that carbon characteristic X-rays are produced deeper in the tar- get. For pure carbon with E0 = 5 keV, the cumulative value of f(χ) = 0.867; that is, 86.7 % of all carbon X-rays that are gener- ated escape, while 13.3 % are absorbed. As the C X-rays are produced deeper with increasing beam energy, the total X-ray absorption increases so that the value of f(χ) for C K decreases sharply with increasing beam energy, as shown in . Fig. 4.21b and in . Table 4.3.

Thus, with E0 = 2 keV, 97.4 % of the carbon X-rays escape the specimen, while at E0 = 30 keV, nearly 90 % of the carbon X-rays generated in pure carbon are absorbed before they can exit the specimen.

When the parameter f(χ) is plotted at every photon energy from the threshold of 100 eV up to the Duane–Hunt limit of the incident beam energy E0, X-ray absorption is seen to sharply modify the X-ray spectrum that is emitted from the target, as illustrated for carbon (. Fig.  4.22), copper (. Fig. 4.23), and gold (. Fig. 4.24). The high relative intensity of the X-ray continuum at low photon energies compared to higher photon energies in the generated spectrum is greatly diminished in the emitted spectrum because of the higher absorption suffered by low energy photons. Discontinuities

in f(χ) are seen at the critical ionization energy of the K-shell in carbon, the K- and L-shells in copper, and the M- and L-shells in gold, corresponding to the sharp increase in μ/ρ just above the critical ionization energy. Because the X-ray continuum is generated at all photon energies, the continuum is affected by every ionization edge represented by the atomic species present, resulting in abrupt steps in the background.

An abrupt decrease in X-ray continuum intensity is observed just above the absorption edge energy due to the increase in the mass absorption coefficient. The characteristic peaks in these spectra are also diminished by absorption, but because a characteristic X-ray is always lower in energy than the ion- ization edge energy from which it originated, the mass absorption coefficient for characteristic X-rays is lower than that for photons with energies just above the shell ionization energies. Thus an element is relatively transparent to its own characteristic X-rays because of the decrease in X-ray absorp- tion below the ionization edge energy.

60

4

.Fig. 4.22 a Absorption parameter f(χ) as a function of photon energy for carbon and an incident beam energy of E0 = 20 keV. Note the abrupt decrease for photons just above the ionization energy of carbon at 0.284 keV.

b Expansion of the region from 0 to 5 keV. Note the abrupt decrease for photons just above the ionization energy of carbon at 0.284 keV. c Comparison of the generated (black) and emitted (red) X-ray spectra for carbon with an incident beam energy of E0 = 20 keV

1.0 a

0.8

0.6

0.4

0.2

0.0 5 10

Photon energy (keV) 15 20

Absorption parameter f(χ) for C, E0 = 20 keV

f(χ)

Absorption parameter f(χ) for C, E0 = 20 keV

1.0

0.8

0.6

0.4

0.2

0.0

f(χ)

1 0

0

2 3

Photon energy (keV) 4 5

b

c1e+8 Generated and emitted X-ray spectra for carbon E0 = 20 keV

1e+7 1e+6 1e+5 1e+4 1e+3

Intensity

1e+2 1e+1

1e+00 5 10

X-ray photon energy (keV)15 20 Generated

Emitted Chapter 4 · X-Rays

61

4

.Fig. 4.23 a Absorption parameter f(χ) as a function of photon energy for copper and an incident beam energy of E0 = 20 keV. Note the abrupt decrease just above the ionization energies of the three L-shells near 0.930 keV and the K-shell ionization energy at 8.98 keV.

b Comparison of the generated (red) and emitted (blue) X-ray spectra for copper with an incident beam energy of E0 = 20 keV. c Comparison of the generated and emitted X-ray spectra for copper with an incident beam energy of E0 = 20 keV; expanded to show the region of the Cu L-shell and Cu K-shell ionization energies

Absorption parameter f(χ) for copper, E0 = 20 keV

Photon energy (keV)

f(χ)

1.0 a

0.8

0.6

0.4

0.2

0.00 5 10 15 20

c 1e+7

1e+6

1e+5

Emitted Copper, E0 = 20 keV

Generated

1e+4

1e+3

1e+20 2

Photon energy (keV)

X-ray intensity

4 6 8 10

1e+8 1e+7 1e+6

1e+5 Emitted

Generated Copper, E0 = 20 keV

1e+4 X-ray intensity 1e+3

1e+2 1e+1

1e+00 5 10

Photon energy (keV) 15 20

b 4.5 · X-Ray Fluorescence

62

4

1.0

0.8

0.6

0.4

0.2

0.00 5 10

Photon energy (keV)

Absorption parameter f(c) for gold, E0 = 20 keV a

15 20

f(c)

0 5 10

Photon energy (keV) Gold, E0 = 20 keV b 1e+8

1e+7 1e+6 1e+5 1e+4 1e+3 1e+2

1e+0 1e+1

X-ray intensity

15 20

Emitted Generated

0 1 2

Photon energy (keV) Gold, E0 = 20 keV c 1e+8

1e+7

1e+6

1e+5

1e+4

1e+3

1e+2

X-ray intensity

3 4 5

Emitted Generated

Emitted Generated

.Fig. 4.24 a Absorption parameter f(χ) as a function of photon energy for gold and an incident beam energy of E0 = 20 keV. Note the abrupt decrease in f(χ) for photons just above the ionization energies of the gold M-shell and gold L-shell. b Comparison of the generated (red) and emitted (blue) X-ray spectra for gold with an incident beam energy of E0 = 20 keV. c Comparison of the generated (red) and emitted (blue) X-ray spectra for gold with an incident beam energy of E0 = 20 keV; expansion of the region around the gold M-shell ionization edges

Chapter 4 · X-Rays

63

4

References

Crawford J, Cohen D, Doherty G, Atanacio A (2011) Calculated K, L and M-shell X-ray line intensities for light ion impact on selected targets from Z = 6 to 100. ANSTO/E-774, Australian Nuclear Science and Technology Organisation (September, 2011)

Drouin D, Couture A, Joly D, Tastet X, Aimez V, Gauvin R (2007) CASINO V2.42 – a fast and easy-to-use modeling tool for scanning electron microscopy and microanalysis users. Scanning 29:92–101

Fiori CE, Swyt CR, Myklebust RL (1992) Desktop Spectrum Analyzer (DTSA), a comprehensive software engine for electron-excited X-ray spectrome- try. National Institute of Standards and Technology (NIST)  – National Institutes of Health (NIH). United States Patent 5,299,138 (March 29, 1994) (Standard Reference Data Program, NIST, Gaithersburg, MD 20899)

Heinrich KFJ, Fiori CE, Myklebust RL (1979) Relative transition probabili- ties for the X-ray lines from the K level. J Appl Phys 50:5589 Joy D (2006) Monte Carlo simulation. Available at 7 http://www.lehigh.

edu/~maw3/link/mssoft/mcsim.html

Kramers H (1923) On the theory of X-ray absorption and the continuous X-ray spectrum. Phil Mag 46:836

Krause M, Oliver J (1979) Natural widths of atomic K and L levels, Kα X-ray lines and several KLL Auger lines. J Phys Chem Ref Data Monogr 8:329

Lifshin E, Ciccarelli MF, Bolon RB (1980) In: Beaman DR, Ogilvie RE, Wittry DB (eds) Proceedings 8th international conference on X-ray optics and microanalysis, vol 141. Pendell, Midland

Pouchou JL, Pichoir F (1991) Quantitative analysis of homogeneous or stratified microvolumes applying the model ‘PAP’. In: Heinrich KFJ, Newbury DE (eds) Electron probe quantitation, vol 31. Plenum Press, New York

Powell C (1976) Use of Monte Carlo Calculations. In: Heinrich KFJ, Newbury DE, Yakowitz H (eds) Electron probe microanalysis and scanning elec- tron microscopy. National Bureau of standard Special Publication 460. U.S. Government Printing Office, Washington, DC, 97

Ritchie NWM (2015) NIST DTSA-II software, including tutorials, is available free at: 7 www.cstl.nist.gov/div837/837.02/epq/dtsa2/index.html 50%

90%

99%

10 µm Ni-10% Fe E0 = 20 keV

.Fig. 4.25 Range of secondary fluorescence of Fe K-L3 (EK = 7.07 keV) by Ni K-L3 X-rays (7.47 keV). Red arc = extent of direct electron excita- tion of Ni K-L3 and Fe K-L3. Blue arc = range for 50 % of total secondary fluorescence of Fe K-L3 by Ni K-L3; green arc = 90 %; magenta arc = 99 % References

© Springer Science+Business Media LLC 2018

J. Goldstein et al., Scanning Electron Microscopy and X-Ray Microanalysis, https://doi.org/10.1007/978-1-4939-6676-9_5

65

Scanning Electron Microscope