ATOM INDONESIA
Author's Responses to Revisions Required
Article : 1112
Article Title :
Comparisons of water-equivalent diameter measured on images of abdominal routine computed tomography with and without contrast agent
Line # Referee’s Comments Author's Responses
Line 6
This paper is too short, poorly discussed with results and discussion section only around 200 words. The authors should add up more data and discussion
Thank you for your constructive com- ments. We have given the abdominal CT images with and without contrast agent and the difference of the HU between these two images. However, the increase HU is only in some areas depending on the slice location. We added: ”It shows that the axial CT images of the abdomen with the use of a contrast agent and a portal venous phase scan provides better visualization of the por- tal vein and solid organs, i.e. the liver, spleen, kidney, and pancreas than the axial CT images of the abdomen without the use of a contrast agent due to the HU values in- crease in these areas [27,28]. The mean HU value in some areas of the axial CT images of the abdomen was 22.6% greater than without the use of contrast agents.” Please see the revised manuscript, page 5, lines 174-185.
We have added explanation the meaning of R2 value. “The correlation is strong (R2 = 0.974) which indicates that two Dws have strong correlations. Based on this finding, the DW,contrast can be accurately calculated from DW,non-contrast.” Please see the revised manuscript, page 4, lines 191-195.
Line
30
Delete article "a"
Done. Please see the revised manuscript line 31.Line
38
full stop here
We have made changes to the paragraphplease see the revised manuscript, lines 30- 50.
Line 102
Please specifically mention the contrast agent used in the abdominal CT evaluation
The contrast agent was Iopamidol 370 We have mention the contrast agent, please see the revised manuscript, lines 87-88
Line 107
The texts in Fig. 2 are too small, please use bigger font. Also the figure quality is too low
We have enlarged the image. Please see the revised manuscript, page 3, line 150.
Line
191
Please explain why the use of contrast agent
We have added the information in the revised manuscript page 4, lines 251-254.does not significantly affect the Dw, CTDI and SSDE.
Lines 202-
203
Please add up contributorship statement as indicated in the latest version of Atom Indonesia template
Done. We have added the contributorship statement.
Line 204
Please read carefully the author guidelines, particularly how to write references accepted by AI
Done. We have written the references accepted by AI
Please return to Atom Indonesia Editorial Office via
supplementary file in OJS aplication.March 1st, 2021