This is an official electronic document and secured by BMD Laboratory ©BMD Laboratory 2015
BMD Laboratory
–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
LAPORAN
PROGRAM UJI BANDING (PROFICIENCY TEST) ANTAR LABORATORIUM
SKEMA KHUSUS SEMESTER II-2015
Bahan Uji:
AIR LIMBAH I BATCH II (ALDS I-2)
Ref. No. 006/UR/BMD/2015
PARAMETER UJI:
Besi, Timbal, Tembaga, dan Daya Hantar Listrik
FR.06 Laporan Program Uji Banding (Final Report)
Page i of iii
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com Ref. No. 006/UR/BMD/2015
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
KATA PENGANTAR
Puji syukur kiranya kita panjatkan kepada Alloh SWT yang berkenan melimpahkan karunia-Nya sehingga ‘Program Uji Banding Antar Laboratorium (Proficiency Test) Skema
Khusus Semester II-2015’ untuk bahan uji Air Limbah I Batch II (ALDS I-2) dengan parameter uji
besi (Fe), timbal (Pb), tembaga (Cu), dan daya hantar listrik (DHL, 25oC) telah berhasil dilaksanakan dan diselesaikan sampai dengan diterbitkannya laporan ini.
Program uji banding antar laboratorium merupakan salah satu instrumen yang sangat penting sebagai sarana jaminan mutu eksternal (external quality control) bagi laboratorium pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ISO/IEC 17025:2008. Dengan mengikuti program uji banding, laboratorium dapat melakukan evaluasi kinerja pengujian yang dilakukan dalam rangka melakukan peningkatan secara berkesinambungan (continuous improvement) untuk memberikan layanan terbaik kepada pelanggan.
Program skema khusus kali ini diikuti oleh 8 (delapan) laboratorium dimana 5 (lima) laboratorium diantaranya terakreditasi SNI ISO/IEC 17025:2008 oleh Komite Akreditas Nasional (KAN). Setiap laboratorium peserta diberikan 2 (dua) buah bahan uji yang keduanya harus diukur dan dilaporkan masing-masing secara independen. Hal ini dimaksudkan sebagai salah satu studi homogenitas bahan uji yang dibagikan kepada peserta.
Dalam program uji banding ini masih digunakan nilai konsensus (consensus value) dari laboratorium peserta dalam melakukan evaluasi, sehingga tidak memuaskan dalam ketertelusuran (traceability) secara metrologi. Akan tetapi BMD Laboratory meyakini bahwa hal tersebut tidak mengurangi manfaat dari pelaksanaan program uji banding ini. Untuk itu BMD Laboratory akan terus mengupayakan perbaikan demi perbaikan guna menghadirkan program uji banding yang semakin bermanfaat bagi laboratorium-laboratorium di Indonesia.
BMD Laboratory sebagai provider uji banding antar laboratorium menyampaikan apresiasi setinggi-tingginya kepada seluruh pihak yang telah membantu dan menyukseskan kegiatan ini. Khususnya kepada seluruh laboratorium peserta atas partisipasi aktifnya mengikuti program uji
Page ii of iii
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com Ref. No. 006/UR/BMD/2015
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
banding ALDS I-2. Ucapan terimakasih juga disampaikan kepada tim ahli yang telah bekerja keras mengolah data dan menyusun laporan ini. Akhirnya, BMD Laboratory menyampaikan permohonan maaf sekiranya terdapat kekurangan dalam penyelenggaraan program uji banding antar laboratorium. Kritik, saran, dan masukan yang membangun sangat kami harapkan untuk menghadirkan program uji banding yang semakin baik ke depannya. Semoga laporan ini dapat dimanfaatkan oleh laboratorium peserta dan pihak-pihak terkait.
Tangerang Selatan, 31 Juli 2015
Ir. Renaldo Moontri CEO BMD Group (Selaku Manajer Puncak BMD Laboratory)
Page iii of iii
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com Ref. No. 006/UR/BMD/2015
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
DAFTAR ISI
KATA PENGANTAR ... i
DAFTAR ISI ... iii
I. PENDAHULUAN ... 1
II. PELAKSANAAN ... 1
III. STATISTIKA PENGOLAHAN DATA ... 2
III.1 Penilaian Laboratorium ... 2
III.2 Penetapan Nilai Evaluasi Uji Banding ... 2
III.3 Penetapan SDPA ... 2
IV. HASIL UJI BANDING ANTAR LABORATORIUM ... 2
IV.1 Parameter Besi (Fe) ... 5
IV.2 Parameter Timbal (Pb) ... 7
IV.3 Parameter Tembaga (Cu) ... 9
IV.4 Parameter Daya Hantar Listrik (DHL, 25oC) ... 11
V. PENUTUP ... 12
VI. REFERENSI ... 13
Annex A Rekapitulasi Hasil z-score Program Uji Banding ... 14
Annex B Rekapitulasi Bias Nilai Peserta Terhadap Nilai Konsensus ... 15
Page 1 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
I. PENDAHULUAN
Air limbah industri diatur persyaratan baku mutunya sesuai Keputusan Menteri Negara Lingkungan Hidup KEP-51/MENLH/10/1995. Persyaratan baku mutu untuk parameter besi (Fe), timbal (Pb), dan tembaga (Cu) dapat dilihat seperti pada Tabel 1.
Tabel 1. Persyaratan baku mutu air limbah industri[1]
No. Parameter Satuan Persyaratan
1. Besi (Fe) mg/L Maks. 10
2. Timbal (Pb) mg/L Maks. 1
3. Tembaga (Cu) mg/L Maks. 3
Salah satu jaminan mutu bagi laboratorium pengujian dan kalibrasi sesuai persyaratan dalam SNI ISO/IEC 17025:2008 adalah partisipasi dalam uji banding antar laboratorium atau uji profisiensi[2] sebagai jaminan mutu eksternal (external quality control). Untuk itu BMD Laboratory menyelenggarakan program uji banding antar laboratorium (proficiency test) untuk bahan uji Air Limbah I Batch II (ALDS I-2) dengan tujuan sebagai berikut:
1. Sebagai sarana jaminan mutu eksternal (external quality control) bagi laboratorium peserta khususnya laboratorium yang menerapkan standard SNI ISO/IEC 17025:2008. 2. Membantu laboratorium dalam mengevaluasi unjuk kerja pengujian kadar besi (Fe),
timbal (Pb), tembaga (Cu), dan daya hantar listrik (DHL) untuk bahan uji air limbah.
II. PELAKSANAAN
Berikut adalah jadual pelaksanaan program uji banding antar laboratorium (proficiency test) yang diselenggarakan oleh BMD Laboratory untuk bahan uji Air Limbah I Batch II (ALDS I-2): a. Penutupan pendaftaran: 29 Mei 2015.
b. Pendistribusian bahan uji: 15 s.d. 26 Juni 2015.
Page 2 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
d. Pendistribusian laporan program uji banding oleh BMD Laboratory: Maks. 03 Agustus 2015.
III. STATISTIKA PENGOLAHAN DATA III.1 Penilaian Laboratorium
Penilaian terhadap unjuk kerja laboratorium menggunakan z-score dengan tiga kriteria penilaian sebagai berikut:
Untuk |z-score|≤2,0 dikategorikan memuaskan diberi lambang ‘OK’ Untuk 2,0<|z-score|<3,0 dikategorikan meragukan diberi lambang ‘$’ Untuk |z-score|≥3,0 dikategorikan kurang memuaskan diberi lambang ‘$$’
III.2 Penetapan Nilai Evaluasi Uji Banding
Penetapan nilai evaluasi uji banding menggunakan dua pendekatan statistika yaitu: Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E)[3]. Nilai tersebut
merupakan nilai konsensus dari peserta (consensus value)
Seleksi data menggunakan perangkat lunak Distribution AnalyzerTM
, Taylor Enterprises, Inc.[4]
III.3 Penetapan SDPA
SDPA (simpangan baku yang digunakan dalam evaluasi) ditetapkan menggunakan tiga pendekatan statistika yaitu:
Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E)
Simpangan Baku Horwitz[5]
SDPA yang ditetapkan (Assigned SDPA)
IV. HASIL UJI BANDING ANTAR LABORATORIUM
Sebanyak 8 (delapan) laboratorium telah berpartisipasi dalam program uji banding antar laboratorium (proficiency test) yang diselenggarakan oleh BMD Laboratory untuk bahan uji
Page 3 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
Air limbah I Batch II (ALDS I-2) dengan parameter uji besi (Fe), timbal (Pb), tembaga (Cu), dan daya hantar listrik (DHL, 25oC). Adapun 5 (lima) laboratorium diantaranya merupakan laboratorium terakreditasi SNI ISO/IEC 17025:2008 oleh Komite Akreditasi Nasional (KAN). Masing-masing laboratorium diberikan 2 (dua) buah bahan uji dengan kombinasi kode seperti pada Tabel 2. Laboratorium dengan kode BMD-0004-15B2 dan BMD-0011-15B2 mengukur bahan uji untuk parameter Fe, Pb, dan Cu menggunakan AAS I dan AAS II sehingga melaporkan masing-masing 2 (dua) laporan untuk setiap bahan uji. Oleh karena itu diperoleh sebanyak 18 (delapan belas) laporan hasil pengujian untuk parameter Fe, Pb, dan Cu dan 14 (empat belas) laporan hasil pengujian untuk parameter DHL (25oC).
Tabel 2. Kombinasi Kode Bahan Uji
Rekapitulasi hasil pengujian oleh laboratorium peserta disajikan seperti pada Tabel 3. Seluruh data hasil uji banding dapat diolah dengan baik menggunakan Robust Statistic
Algoritma A Mean sesuai ISO 13528: 2005(E). Selain itu dilakukan juga evaluasi data
menggunakan perangkat lunak Distribution AnalyzerTM guna memastikan bahwa data yang
diolah mengikuti pola distribusi normal pada tingkat kepercayaan 95%. Nilai evaluasi uji
No. Kode Bahan Uji
1. BMD-0001-15B2 BMD-0014-15B2 2. BMD-0002-15B2 BMD-0013-15B2 3. BMD-0003-15B2 BMD-0012-15B2 4. BMD-0004(I)-15B2 BMD-0004(II)-15B2 BMD-0011(I)-15B2 BMD-0011(II)-15B2 5. BMD-0005-15B2 BMD-0010-15B2 6. BMD-0006-15B2 BMD-0009-15B2 7. BMD-0007-15B2 BMD-0008-15B2 8. BMD-0015-15B2 BMD-0020-15B2
Page 4 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
banding (assigned value) merupakan nilai konsensus (consensus value) dari laboratorium peserta sehingga nilai tersebut bukan merupakan nilai acuan yang tertelusur secara metrologi ke sistem satuan internasional (SI).
Tabel 3. Rekapitulasi Hasil Pengujian Laboratorium Peserta
No. Kode Lab. Besi (Fe) Timbal (Pb) Tembaga (Cu)
Daya Hantar Listrik (DHL 25oC) mg.L-1 mg.L-1 mg.L-1 mS.cm-1 1. BMD-0001-15B2 1.55±0.0107 0.51±0.0094 1.22±0.0091 4.93 Metoda SNI 6989.4:2009 SNI 6989.8:2009 SNI 6989.6:2009 SNI 06-6989.1:2004 2. BMD-0002-15B2 1.64±0.12 0.65±0.10 1.04±0.04 4.90 Metoda SNI 6989.4-2009 SNI 6989.8-2009 SNI 6989.6-2009 SNI 06-6989.21-2004 3. BMD-0003-15B2 2.28±0.03 0.35±0.04 1.42±0.09 4.74 Metoda SNI.6989.4. 2009 SNI 6989.8.2009 SNI 6989.6.2009 SNI.6989.1.2004 4. BMD-0004(I)-15B2 1.58 0.43 1.09 2.45 Metoda Spektrometri (AAS I) Spektrometri (AAS I) Spektrometri (AAS I) 5. BMD-0004(II)-15B2 1.51 0.43 1.14 Metoda Spektrometri (AAS II) Spektrometri (AAS II) Spektrometri
(AAS II) Konduktometri 6. BMD-0005-15B2 0.70 0.34 0.58 5.01 Metoda SNI 6989.4-2009 SNI 6989.8-2009 SNI 6989.6-2009 SNI 06-6989.1-2004 7. BMD-0006-15B2 1.51 0.50 1.07 3.91 Metoda SNI 06-6989.4-2009 SNI 06-6989.8-2004 SNI 06-6989.6-2009 SNI 06-6989.1-2004 8. BMD-0007-15B2 1.53 0.48 1.10 4.95 Metoda SNI 06-6989.4-2009 SNI 06-6989.8-2009 SNI 06-6989.6-2009 SNI 06-6989.1-2004 9. BMD-0008-15B2 1.54 0.48 1.08 4.94 Metoda SNI 06-6989.4-2009 SNI 06-6989.8-2009 SNI 06-6989.6-2009 SNI 06-6989.1-2004
Page 5 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
IV.1 Parameter Besi (Fe)
Sebanyak 18 (delapan belas) laporan hasil pengujian untuk parameter besi (Fe) diterima oleh BMD Laboratory dan 8 (delapan) diantaranya disertai dengan laporan nilai estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan
No. Kode Lab.
Besi (Fe) Timbal (Pb) Tembaga (Cu) Daya Hantar Listrik (DHL 25oC) mg.L-1 mg.L-1 mg.L-1 mS.cm-1 10. BMD-0009-15B2 1.52 0.51 1.07 3.94 Metoda SNI.06-6989.4-2009 SNI.06-6989.8-2004 SNI.06-6989.6-2009 SNI.06-6989.1-2004 11. BMD-0010-15B2 0.85 0.35 0.70 4.75 Metoda SNI 6989.4:2009 SNI 6989.8-2009 SNI 6989.6:2009 SNI 06-6989.1-2004 12. BMD-0011(I)-15B2 1.44 0.38 1.07 2.85 Metoda Spektrometri (AAS I) Spektrometri (AAS I) Spektrometri (AAS I) 13. BMD-0011(II)-15B2 1.55 0.38 1.08 Metoda Spektrometri (AAS II) Spektrometri (AAS II) Spektrometri
(AAS II) Konduktometri 14. BMD-0012-15B2 2.26±0.03 0.28±0.01 1.50±0.08 4.60 Metoda SNI.6989.4. 2009 SNI 6989.8.2009 SNI 6989.6.2009 SNI.6989.1.2004 15. BMD-0013-15B2 1.66±0.12 0.65±0.10 1.04±0.04 4.90 Metoda SNI 6989.4-2009 SNI 6989.8-2009 SNI 6989.6-2009 SNI 06-6989.21-2004 16. BMD-0014-15B2 1.56±0.0107 0.51±0.0094 1.22±0.0091 4.93 Metoda SNI 6989.4:2009 SNI 6989.8:2009 SNI 6989.6:2009 SNI 06-6989.1:2004 17. BMD-0015-15B2 1.28±0.27 0.38±0.14 0.97±0.58 - Metoda SNI 06-6989.4-2004 SNI 06-6989.8-2004 SNI 06-6989.6-2004 - 18. BMD-0020-15B2 1.27±0.27 0.38±0.14 0.97±0.58 - Metoda SNI 06-6989.4-2004 SNI 06-6989.8-2004 SNI 06-6989.6-2004 -
Page 6 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
data secara statistika diperoleh sebanyak 12 (dua belas) data yang dilaporkan laboratorium masuk dalam kategori ‘memuaskan’ (|z-score|≤2,0), 2 (dua) data yang dilaporkan laboratorium masuk dalam kategori ‘meragukan’ (2,0<|z-score|<3,0), dan 4 (empat) data yang dilaporkan laboratorium masuk dalam kategori ‘kurang memuaskan’ (|z-score|≥3,0). Hasil pengolahan data disajikan pada Tabel 4 dan Gambar 1.
Tabel 4. Hasil Pengolahan Data Parameter Besi (Fe)
No. Kode Peserta Nilai Laporan Peserta (mg/L)
U, k=2
(mg/L)
Nilai Evaluasi Uji Banding (mg/L) SDPA (mg/L) z-score Kategori 1. BMD-0001-15B2 1.55 0.0107 1.53 0.12 0.19 OK 2. BMD-0002-15B2 1.64 0.12 1.53 0.12 0.95 OK 3. BMD-0003-15B2 2.28 0.03 1.53 0.12 6.37 $$ 4. BMD-0004(I)-15B2 1.58 - 1.53 0.12 0.45 OK 5. BMD-0004(II)-15B2 1.51 - 1.53 0.12 -0.15 OK 6. BMD-0005-15B2 0.70 - 1.53 0.12 -7.00 $$ 7. BMD-0006-15B2 1.51 - 1.53 0.12 -0.15 OK 8. BMD-0007-15B2 1.53 - 1.53 0.12 0.02 OK 9. BMD-0008-15B2 1.54 - 1.53 0.12 0.11 OK 10. BMD-0009-15B2 1.52 - 1.53 0.12 -0.06 OK 11. BMD-0010-15B2 0.85 - 1.53 0.12 -5.73 $$ 12. BMD-0011(I)-15B2 1.44 - 1.53 0.12 -0.74 OK 13. BMD-0011(II)-15B2 1.55 - 1.53 0.12 0.19 OK 14. BMD-0012-15B2 2.26 0.030 1.53 0.12 6.20 $$ 15. BMD-0013-15B2 1.66 0.12 1.53 0.12 1.12 OK 16. BMD-0014-15B2 1.56 0.0107 1.53 0.12 0.28 OK 17. BMD-0015-15B2 1.28 0.27 1.53 0.12 -2.09 $ 18. BMD-0020-15B2 1.27 0.27 1.53 0.12 -2.18 $ Keterangan:
Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust
Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis
Page 7 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com 0.60 0.75 0.90 1.05 1.20 1.35 1.50 1.65 1.80 1.95 2.10 2.25 2.40 B M D -0 0 0 5 -1 5 B 2 B M D -0 0 1 0 -1 5 B 2 B M D -0 0 2 0 -1 5 B 2 B M D -0 0 1 5 -1 5 B 2 B M D -0 0 1 1 (I )-1 5 B 2 B M D -0 0 0 6 -1 5 B 2 B M D -0 0 0 4 (I I) -1 5 B 2 B M D -0 0 0 9 -1 5 B 2 B M D -0 0 0 7 -1 5 B 2 B M D -0 0 0 8 -1 5 B 2 B M D -0 0 0 1 -1 5 B 2 B M D -0 0 1 1 (I I) -1 5 B 2 B M D -0 0 1 4 -1 5 B 2 B M D -0 0 0 4 (I )-1 5 B 2 B M D -0 0 0 2 -1 5 B 2 B M D -0 0 1 3 -1 5 B 2 B M D -0 0 1 2 -1 5 B 2 B M D -0 0 0 3 -1 5 B 2 F e (m g /L ) Lab. Code Cons. value +2SDPA +3SDPA -2SDPA -3SDPA Cons. value = 1.53 mg/L
Algoritma A Mean ISO 13528:2005(E) SDPA = 0.12 mg/L (SD Robust) SD Horwitz = 0.23 mg/L
©BMD Laboratory
simpangan baku yang diperoleh dari pengolahan data secara statistika menggunakan
Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E). U yaitu ketidakpastian
diperluas dengan faktor cakupan (k)=2 pada tingkat kepercayaan 95%.
Gambar 1. Grafik Hasil Uji Banding Parameter Besi (Fe)
IV.2 Parameter Timbal (Pb)
Sebanyak 18 (delapan belas) laporan hasil pengujian untuk parameter timbal (Pb) diterima oleh BMD Laboratory dan 8 (delapan) diantaranya disertai dengan laporan nilai estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan data secara statistika diperoleh sebanyak 16 (enam belas) data yang dilaporkan laboratorium masuk dalam kategori ‘memuaskan’ (|z-score|≤2,0) dan 2 (dua) data yang
Page 8 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
dilaporkan laboratorium masuk dalam kategori ‘meragukan’. Hasil pengolahan data disajikan pada Tabel 5 dan Gambar 2.
Tabel 5. Hasil Pengolahan Data Parameter Timbal (Pb)
No. Kode Peserta Nilai Laporan Peserta (mg/L)
U, k=2
(mg/L)
Nilai Evaluasi Uji Banding (mg/L) SDPA (mg/L) z-score Kategori 1. BMD-0001-15B2 0.51 0.0094 0.44 0.08 0.91 OK 2. BMD-0002-15B2 0.65 0.10 0.44 0.08 2.67 $ 3. BMD-0003-15B2 0.35 0.04 0.44 0.08 -1.11 OK 4. BMD-0004(I)-15B2 0.43 - 0.44 0.08 -0.10 OK 5. BMD-0004(II)-15B2 0.43 - 0.44 0.08 -0.10 OK 6. BMD-0005-15B2 0.34 - 0.44 0.08 -1.23 OK 7. BMD-0006-15B2 0.50 - 0.44 0.08 0.78 OK 8. BMD-0007-15B2 0.48 - 0.44 0.08 0.53 OK 9. BMD-0008-15B2 0.48 - 0.44 0.08 0.53 OK 10. BMD-0009-15B2 0.51 - 0.44 0.08 0.91 OK 11. BMD-0010-15B2 0.35 - 0.44 0.08 -1.11 OK 12. BMD-0011(I)-15B2 0.38 - 0.44 0.08 -0.73 OK 13. BMD-0011(II)-15B2 0.38 - 0.44 0.08 -0.73 OK 14. BMD-0012-15B2 0.28 0.01 0.44 0.08 -1.99 OK 15. BMD-0013-15B2 0.65 0.10 0.44 0.08 2.67 $ 16. BMD-0014-15B2 0.51 0.0094 0.44 0.08 0.91 OK 17. BMD-0015-15B2 0.38 0.14 0.44 0.08 -0.73 OK 18. BMD-0020-15B2 0.38 0.14 0.44 0.08 -0.73 OK Keterangan:
Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust
Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis
menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C), sedangkan SDPA adalah simpangan baku Horwitz (SDHorwitz). U yaitu ketidakpastian diperluas dengan faktor cakupan (k)=2 pada tingkat kepercayaan 95%.
Page 9 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com 0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 B M D -0 0 1 2 -1 5 B 2 B M D -0 0 0 5 -1 5 B 2 B M D -0 0 0 3 -1 5 B 2 B M D -0 0 1 0 -1 5 B 2 B M D -0 0 1 1 (I )-1 5 B 2 B M D -0 0 1 1 (I I) -1 5 B 2 B M D -0 0 1 5 -1 5 B 2 B M D -0 0 2 0 -1 5 B 2 B M D -0 0 0 4 (I )-1 5 B 2 B M D -0 0 0 4 (I I) -1 5 B 2 B M D -0 0 0 7 -1 5 B 2 B M D -0 0 0 8 -1 5 B 2 B M D -0 0 0 6 -1 5 B 2 B M D -0 0 0 1 -1 5 B 2 B M D -0 0 0 9 -1 5 B 2 B M D -0 0 1 4 -1 5 B 2 B M D -0 0 0 2 -1 5 B 2 B M D -0 0 1 3 -1 5 B 2 P b (m g /L ) Lab. Code Cons. value +2SDPA +3SDPA -2SDPA -3SDPA Cons. value = 0.44 mg/L
Algoritma A Mean ISO 13528:2005(E) SDPA = 0.08 mg/L (Horwitz Function) SD Robust = 0.10 mg/L
©BMD Laboratory
Gambar 2. Grafik Hasil Uji Banding Parameter Timbal (Pb)
IV.3 Parameter Tembaga (Cu)
Sebanyak 18 (delapan belas) laporan hasil pengujian untuk parameter tembaga (Cu) diterima oleh BMD Laboratory dan 8 (delapan) diantaranya disertai dengan laporan nilai estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan data secara statistika diperoleh sebanyak 14 (empat belas) data yang dilaporkan laboratorium masuk dalam kategori ‘memuaskan’ (|z-score|≤2,0) dan 4 (empat) data yang dilaporkan laboratorium masuk dalam kategori ‘kurang memuaskan’ (|z-score|≥3,0). Hasil pengolahan data disajikan pada Tabel 6 dan Gambar 3.
Page 10 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
Tabel 6. Hasil Pengolahan Data Parameter Tembaga (Cu)
No. Kode Peserta Nilai Laporan Peserta (mg/L)
U, k=2
(mg/L)
Nilai Evaluasi Uji Banding (mg/L) SDPA (mg/L) z-score Kategori 1. BMD-0001-15B2 1.22 0.0091 1.08 0.09 1.61 OK 2. BMD-0002-15B2 1.04 0.04 1.08 0.09 -0.42 OK 3. BMD-0003-15B2 1.42 0.09 1.08 0.09 3.87 $$ 4. BMD-0004(I)-15B2 1.09 - 1.08 0.09 0.14 OK 5. BMD-0004(II)-15B2 1.14 - 1.08 0.09 0.71 OK 6. BMD-0005-15B2 0.58 - 1.08 0.09 -5.61 $$ 7. BMD-0006-15B2 1.07 - 1.08 0.09 -0.08 OK 8. BMD-0007-15B2 1.10 - 1.08 0.09 0.26 OK 9. BMD-0008-15B2 1.08 - 1.08 0.09 0.03 OK 10. BMD-0009-15B2 1.07 - 1.08 0.09 -0.08 OK 11. BMD-0010-15B2 0.70 - 1.08 0.09 -4.26 $$ 12. BMD-0011(I)-15B2 1.07 - 1.08 0.09 -0.08 OK 13. BMD-0011(II)-15B2 1.08 - 1.08 0.09 0.03 OK 14. BMD-0012-15B2 1.50 0.08 1.08 0.09 4.77 $$ 15. BMD-0013-15B2 1.04 0.04 1.08 0.09 -0.42 OK 16. BMD-0014-15B2 1.22 0.0091 1.08 0.09 1.61 OK 17. BMD-0015-15B2 0.97 0.58 1.08 0.09 -1.21 OK 18. BMD-0020-15B2 0.97 0.58 1.08 0.09 -1.21 OK Keterangan:
Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust
Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis
menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C). SDPA adalah simpangan baku yang diperoleh dari pengolahan data secara statistika menggunakan
Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E). U yaitu ketidakpastian
Page 11 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 B M D -0 0 0 5 -1 5 B 2 B M D -0 0 1 0 -1 5 B 2 B M D -0 0 1 5 -1 5 B 2 B M D -0 0 2 0 -1 5 B 2 B M D -0 0 0 2 -1 5 B 2 B M D -0 0 1 3 -1 5 B 2 B M D -0 0 0 6 -1 5 B 2 B M D -0 0 0 9 -1 5 B 2 B M D -0 0 1 1 (I )-1 5 B 2 B M D -0 0 0 8 -1 5 B 2 B M D -0 0 1 1 (I I) -1 5 B 2 B M D -0 0 0 4 (I )-1 5 B 2 B M D -0 0 0 7 -1 5 B 2 B M D -0 0 0 4 (I I) -1 5 B 2 B M D -0 0 0 1 -1 5 B 2 B M D -0 0 1 4 -1 5 B 2 B M D -0 0 0 3 -1 5 B 2 B M D -0 0 1 2 -1 5 B 2 C u (m g /L ) Lab. Code Cons. value +2SDPA +3SDPA -2SDPA -3SDPA Cons. value = 1.08 mg/L
Algoritma A Mean ISO 13528:2005(E) SDPA = 0.09 mg/L (SD Robust) SD Horwitz = 0.17 mg/L
©BMD Laboratory
Gambar 3. Grafik Hasil Uji Banding Parameter Tembaga (Cu)
IV.3 Parameter Daya Hantar Listrik (DHL, 25oC)
Sebanyak 14 (empat belas) laporan hasil pengujian untuk parameter daya hantar listril (DHL, 25oC) diterima oleh BMD Laboratory dan 2 (dua) diantaranya disertai dengan laporan nilai estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan data secara statistika diperoleh sebanyak 10 (sepuluh) data yang dilaporkan laboratorium masuk dalam kategori ‘memuaskan’ (|z-score|≤2,0) dan 4 (empat) data yang dilaporkan laboratorium masuk dalam kategori ‘kurang memuaskan’ (|z-score|≥3,0). Hasil pengolahan data disajikan pada Tabel 7 dan Gambar 4.
Page 12 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
Tabel 7. Hasil Pengolahan Data Parameter Daya Hantar Listrik (DHL, 25oC)
No. Kode Peserta Nilai Laporan Peserta (mS/cm)
U, k=2
(mS/cm)
Nilai Evaluasi Uji Banding (mS/cm) SDPA (mS/cm) z-score Kategori 1. BMD-0001-15B2 4.93 2.72 4.91 0.25 0.09 OK 2. BMD-0002-15B2 4.90 - 4.91 0.25 -0.03 OK 3. BMD-0003-15B2 4.74 - 4.91 0.25 -0.68 OK 4. BMD-0004-15B2 2.45 - 4.91 0.25 -10.02 $$ 5. BMD-0005-15B2 5.01 - 4.91 0.25 0.42 OK 6. BMD-0006-15B2 3.91 - 4.91 0.25 -4.07 $$ 7. BMD-0007-15B2 4.95 - 4.91 0.25 0.17 OK 8. BMD-0008-15B2 4.94 - 4.91 0.25 0.13 OK 9. BMD-0009-15B2 3.94 - 4.91 0.25 -3.94 $$ 10. BMD-0010-15B2 4.75 - 4.91 0.25 -0.64 OK 11. BMD-0011-15B2 2.85 - 4.91 0.25 -8.39 $$ 12. BMD-0012-15B2 4.60 - 4.91 0.25 -1.25 OK 13. BMD-0013-15B2 4.90 - 4.91 0.25 -0.03 OK 14. BMD-0014-15B3 4.93 2.72 4.91 0.25 0.09 OK Keterangan:
Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust
Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis
menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C), sedangkan SDPA adalah simpangan baku yang ditetapkan (assigned SDPA) sebesar 5% dari nilai konsensus (consensus value). U yaitu ketidakpastian diperluas dengan faktor cakupan (k)=2 pada tingkat kepercayaan 95%.
V. PENUTUP
Program uji banding antar laboratorium merupakan salah satu sarana jaminan mutu eksternal bagi laboratorium pengujian dan kalibrasi. Hasil dari program uji banding ini
Page 13 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 B M D -0 0 0 4 -1 5 B 2 B M D -0 0 1 1 -1 5 B 2 B M D -0 0 0 6 -1 5 B 2 B M D -0 0 0 9 -1 5 B 2 B M D -0 0 1 2 -1 5 B 2 B M D -0 0 0 3 -1 5 B 2 B M D -0 0 1 0 -1 5 B 2 B M D -0 0 0 2 -1 5 B 2 B M D -0 0 1 3 -1 5 B 2 B M D -0 0 0 1 -1 5 B 2 B M D -0 0 1 4 -1 5 B 3 B M D -0 0 0 8 -1 5 B 2 B M D -0 0 0 7 -1 5 B 2 B M D -0 0 0 5 -1 5 B 2 D H L 2 5 oC (m S /c m ) Lab. Code Cons. value +2SDPA +3SDPA -2SDPA -3SDPA Cons. value = 4.91 mS/cm
Algoritma A Mean ISO 13528:2005(E) SDPA = 0.25 mS/cm (5% assigned value) SD Robust = 0.05 mS/cm
©BMD Laboratory
diharapkan dapat digunakan sebagai sarana evaluasi bagi laboratorium untuk meningkatkan kinerja dalam kegiatan pengujian yang dilakukan.
Gambar 4. Grafik Hasil Uji Banding Parameter Daya Hantar Listrik (DHL, 25oC)
VI. REFERENSI
[1] Keputusan Menteri Negara Lingkungan Hidup Nomor: KEP-51/MENLH/10/1995 [2] SNI ISO/IEC 17025:2008
[3] ISO 13528:2005(E)
[4] Distribution AnalyzerTM, Copyright © by Taylor Enterprises, Inc. USA
[5] M. Thompson (2004) The Amazing Horwitz Function, AMC Technical Brief No.17, Royal Society of Chemistry
Page 14 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
Annex A
Rekapitulasi Hasil z-score Program Uji Banding
No. Kode Peserta
z-score Fe Pb Cu DHL 25oC 1. BMD-0001-15B2 0.19 OK 0.91 OK 1.61 OK 0.09 OK 2. BMD-0002-15B2 0.95 OK 2.67 $ -0.42 OK -0.03 OK 3. BMD-0003-15B2 6.37 $$ -1.11 OK 3.87 $$ -0.68 OK 4. BMD-0004(I)-15B2 0.45 OK -0.10 OK 0.14 OK -10.02 $$ 5. BMD-0004(II)-15B2 -0.15 OK -0.10 OK 0.71 OK 6. BMD-0005-15B2 -7.00 $$ -1.23 OK -5.61 $$ 0.42 OK 7. BMD-0006-15B2 -0.15 OK 0.78 OK -0.08 OK -4.07 $$ 8. BMD-0007-15B2 0.02 OK 0.53 OK 0.26 OK 0.17 OK 9. BMD-0008-15B2 0.11 OK 0.53 OK 0.03 OK 0.13 OK 10. BMD-0009-15B2 -0.06 OK 0.91 OK -0.08 OK -3.94 $$ 11. BMD-0010-15B2 -5.73 $$ -1.11 OK -4.26 $$ -0.64 OK 12. BMD-0011(I)-15B2 -0.74 OK -0.73 OK -0.08 OK -8.39 $$ 13. BMD-0011(II)-15B2 0.19 OK -0.73 OK 0.03 OK 14. BMD-0012-15B2 6.20 $$ -1.99 OK 4.77 $$ -1.25 OK 15. BMD-0013-15B2 1.12 OK 2.67 $ -0.42 OK -0.03 OK 16. BMD-0014-15B2 0.28 OK 0.91 OK 1.61 OK 0.09 OK 17. BMD-0015-15B2 -2.09 $ -0.73 OK -1.21 OK - - 18. BMD-0020-15B2 -2.18 $ -0.73 OK -1.21 OK - - Keterangan: OK : Kategori ‘memuaskan’ $ : Kategori ‘meragukan’
Page 15 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com -1.00 -0.80 -0.60 -0.40 -0.20 0.00 0.20 0.40 0.60 0.80 1.00 b ( m g /L ) Lab. Code Parameter Besi (Fe)
-0.20 -0.15 -0.10 -0.05 0.00 0.05 0.10 0.15 0.20 0.25 b ( m g /L ) Lab. Code Parameter Timbal (Pb)
Annex B
Page 16 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com -3.00 -2.50 -2.00 -1.50 -1.00 -0.50 0.00 0.50 b ( m S /c m ) Lab. Code
Parameter DHL 25
oC
-0.60 -0.40 -0.20 0.00 0.20 0.40 0.60 b ( m g /L ) Lab. CodePage 17 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
No Kode Peserta Nilai Laporan
Peserta (mg/L) 1 BMD-0001-15B2 0.51 2 BMD-0002-15B2 0.65 3 BMD-0003-15B2 0.35 4 BMD-0004(I)-15B2 0.43 5 BMD-0005-15B2 0.34 6 BMD-0006-15B2 0.50 7 BMD-0007-15B2 0.48 8 BMD-0008-15B2 0.48 9 BMD-0009-15B2 0.51 10 BMD-0010-15B2 0.35 11 BMD-0011(I)-15B2 0.38 12 BMD-0012-15B2 0.28 13 BMD-0013-15B2 0.65 14 BMD-0014-15B2 0.51 15 BMD-0004(II)-15B2 0.43 16 BMD-0011(II)-15B2 0.38 17 BMD-0015-15B2 0.38 18 BMD-0020-15B2 0.38 Sample Size = 18 Average = 0.444 Standard Deviation = 0.102 Skewness = 0.66 Excess Kurtosis = 0.10 Test of Fit: p-value = 0.3865 (SK All) Decision = Pass (SK Spec) Decision = No specs
With 95% confidence more than 99% of the values are between 0.0674 and 0.8204
0.2800 0.4650 0.6500
Annex C
Evaluasi Distribusi Normal Data Program Uji Banding
Normal Distribution Analysis (Parameter Fe)
Powered by:
Distribution AnalyzerTM (licensed software) ©Taylor Enterprises, Inc.
No Kode Peserta Nilai Laporan Peserta (mg/L) 1 BMD-0001-15B2 1.55 2 BMD-0002-15B2 1.64 3 BMD-0003-15B2 2.28 4 BMD-0004(I)-15B2 1.58 5 BMD-0005-15B2 0.70 6 BMD-0006-15B2 1.51 7 BMD-0007-15B2 1.53 8 BMD-0008-15B2 1.54 9 BMD-0009-15B2 1.52 10 BMD-0010-15B2 0.85 11 BMD-0011(I)-15B2 1.44 12 BMD-0012-15B2 2.26 13 BMD-0013-15B2 1.66 14 BMD-0014-15B2 1.56 15 BMD-0004(II)-15B2 1.51 16 BMD-0011(II)-15B2 1.55 17 BMD-0015-15B2 1.28 18 BMD-0020-15B2 1.27 Sample Size = 18 Average = 1.076 Standard Deviation = 0.211 Skewness = -0.38 Excess Kurtosis = 1.79 Test of Fit: p-value = 0.1972 (SK All) Decision = Pass (SK Spec) Decision = No specs
With 95% confidence more than 99% of the values are between 0.2943 and 1.8568
0.5800 1.0400 1.5000
Normal Distribution Analysis (Parameter Pb)
Powered by:
Distribution AnalyzerTM (licensed software) ©Taylor Enterprises, Inc.
Page 18 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
Normal Distribution Analysis (Parameter Cu)
Powered by:
Distribution AnalyzerTM
(licensed software) ©Taylor Enterprises, Inc.
No Kode Peserta Nilai Laporan
Peserta (mg/L) 1 BMD-0001-15B2 1.22 2 BMD-0002-15B2 1.04 3 BMD-0003-15B2 1.42 4 BMD-0004(I)-15B2 1.09 5 BMD-0005-15B2 0.58 6 BMD-0006-15B2 1.07 7 BMD-0007-15B2 1.10 8 BMD-0008-15B2 1.08 9 BMD-0009-15B2 1.07 10 BMD-0010-15B2 0.70 11 BMD-0011(I)-15B2 1.07 12 BMD-0012-15B2 1.50 13 BMD-0013-15B2 1.04 14 BMD-0014-15B2 1.22 15 BMD-0004(II)-15B2 1.14 16 BMD-0011(II)-15B2 1.08 17 BMD-0015-15B2 0.97 18 BMD-0020-15B2 0.97 Sample Size = 18 Average = 1.076 Standard Deviation = 0.211 Skewness = -0.38 Excess Kurtosis = 1.79 Test of Fit: p-value = 0.1972 (SK All) Decision = Pass (SK Spec) Decision = No specs
With 95% confidence more than 99% of the values are between 0.2943 and 1.8568
0.5800 1.0400 1.5000
Normal Distribution Analysis (Parameter DHL, 25oC)
Powered by:
Distribution AnalyzerTM (licensed software) ©Taylor Enterprises, Inc.
No Kode Peserta Nilai Laporan
Peserta (mS/cm) 1 BMD-0001-15B2 4.93 2 BMD-0002-15B2 4.90 3 BMD-0003-15B2 4.74 4 BMD-0004-15B2 2.45 5 BMD-0005-15B2 5.01 6 BMD-0006-15B2 3.91 7 BMD-0007-15B2 4.95 8 BMD-0008-15B2 4.94 9 BMD-0009-15B2 3.94 10 BMD-0010-15B2 4.75 11 BMD-0011-15B2 2.85 12 BMD-0012-15B2 4.60 13 BMD-0013-15B2 4.90 14 BMD-0014-15B3 4.93 Sample Size = 14 Average = 4.414 Standard Deviation = 0.830 Skewness = -1.65 Excess Kurtosis = 1.71 Test of Fit: p-value = 0.0244 (SK All) Decision = Fail (SK Spec) Decision = No specs
Page 19 of 19
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 006/UR/BMD/2015
BMD Laboratory
–Provider of Proficiency Testing ProgramHead Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: info@bmdstreet.com
www.ujiprofisiensi.com
No Kode Peserta Nilai Laporan
Peserta (mS/cm) 1 BMD-0001-15B2 4.93 2 BMD-0002-15B2 4.90 3 BMD-0003-15B2 4.74 4 BMD-0005-15B2 5.01 5 BMD-0007-15B2 4.95 6 BMD-0008-15B2 4.94 7 BMD-0010-15B2 4.75 8 BMD-0012-15B2 4.60 9 BMD-0013-15B2 4.90 10 BMD-0014-15B3 4.93 Sample Size = 10 Average = 4.865 Standard Deviation = 0.126 Skewness = -1.19 Excess Kurtosis = 0.71 Test of Fit: p-value = 0.1531 (SK All) Decision = Pass (SK Spec) Decision = No specs
With 95% confidence more than 99% of the values are between 4.3046 and 5.4254
4.6000 4.8050 5.0100
Normal Distribution Analysis (Parameter DHL, 25oC)
Powered by:
Distribution AnalyzerTM
(licensed software) ©Taylor Enterprises, Inc.
BMD-0004-15B2, BMD-0006-15B2, BMD-0009-15B2,