• Tidak ada hasil yang ditemukan

Chapter 1: Origin and prediction of free-solution interaction studies performed label-free

N/A
N/A
Protected

Academic year: 2023

Membagikan "Chapter 1: Origin and prediction of free-solution interaction studies performed label-free "

Copied!
244
0
0

Teks penuh

We then postulate a new theory: changes in molecular conformation and hydration are the source of the free solution refractive index signals. The chapter concludes with a fairly simple method that accurately predicts both the magnitude and direction of free solution RI signals. In addition, this work provides detailed instructions on how to perform free solution assays (FSA) using a backscattering interferometer (BSI).

Theory

Origin and prediction of free-solution interaction studies performed label-free

While we, and others, have postulated the origin of the free-solution signal, no explicit explanation for the physical phenomenon has emerged. We describe here how the relative mass of the binding partners plays a minor role in determining the Free-Solution Response Function (FreeSRF). This simple equation demonstrates that the edge shift (in radians) quantified by an interferometer when a folding, binding or hybridization event is measured in free solution (no labels) is directly proportional to: a) the magnitude of structural change (mainly conformation and hydration) of the sample; b) the dθ/dη sensitivity of the interferometer (which includes the optical path length); and c) the concentration of the analyte.

Instrumentation & Engineering

A Highly Compensated Interferometer for Biochemical Analysis

The result was a beam that was approximately 12 mm in the long axis of the channel and. As shown, 1 µL is pipetted into the chip inlet on either side of the channel (Fig. 2.2C). Then the temperature of the entire chip was increased (by the Peltier) in increments of 0.5°C and the phase shifts were recorded for one minute.

Longitudinal Pixel Averaging for Improved Compensation in Backscattering Interferometry Compensation in Backscattering Interferometry

Illumination of the microfluidic channel with a laser extending along the axis of the channel produces a series of elongated interference fringes (Fig. 3.1). By placing a hole in the center of the microfluidic channel, the two solutions to be compared can be separated. In each case, two regions of the same "object" (microfluidic channel) are simultaneously interrogated from a common source.

Upon further investigation, we realized that small imperfections in the fabrication process of the microfluidic channel serving as an interferometer can result in variations of up to 5% in depth/diameter. This variation can affect the distribution of the beams or laser energy that forms the interference pattern. The result is some non-uniformity in the RI response function between individual regions of the channel.

We report here that interrogating longer edge pattern regions averages out these imperfections, resulting in interference patterns that have increased “similarity” between multiple regions. So the noise from a 2°C change in room when using a 2.75mm window is only 2% of the link signal. The standard deviation of the baseline noise over a 3-second period, in the absence of electronic filtering, is 0.68 milliradians, providing a 3-sigma.

In summary, this work shows that by measuring two larger regions of diffuse and elongated fringe patterns produced in the CBSI optical setup, a significant compensation of environmental disturbances in interferometric measurements can be obtained.

Compensated Interferometric Reader Signal Extraction and Analysis

Then, after a transition period, test sample 2 fills both detection regions, yielding a phase shift of 0, shown as time E in Fig. The average phase change over the duration of B is called the "test/reference phase shift (+p )", while the average phase change over the duration of D is the "reference/test phase shift (-p)." 4.2, there are several methods. which can be used to obtain a value for the “signal.” The first method, as demonstrated in Fig.

Performing the phase shift – baseline calculation results in an n=3, with an average phase shift of +p (indicated by the three vertical arrows in Fig. 4.2A). A similar phase shift – baseline calculation can also be performed by subtracting the negative reference/test swing (eg Fig. 4.1 region D, -p). Because the same test and reference sample are used to produce the alternating droplets, the phase shift from Fig.

The data in Fig 4.3 can be tabulated in the +p method, the -p method and the 2×p method, and these results are shown in Fig. It should be noted here that as demonstrated by the data in Figure 3, a 5 test/reference solution pair results in 19 regions, 10 of which will be "baseline". The first "baseline" is typically longer than the others, due to the first droplet in the standard droplet train configuration having a larger volume than the other drops (described in more detail below). An expanded view of a set of test/reference solution pairs is displayed in Fig 4.5, expressed by showing the time each droplet takes to traverse the detection region.

The lengths of all droplets used in the standard array of droplets are presented in Fig.

Biomedical Applications

Rapid assay development and quantification of two chemical nerve agents in serum

In the modern era, some of the most widely used chemical warfare weapons belong to the family of organophosphorus nerve agents (OPNAs). The response from the reported off-target screens, or interference, signals well below the limit of quantification (LOQ) of the assay for binding to the best-in-class aptamer probe. The screen was performed by measuring the FSA signal for a high concentration of OPNA acid (500 nM) in the absence and presence of aptamer at a concentration of 100 nM.

The reference sample was prepared by adding 50 µl of the 1 µM aptamer solution to 50 µl of a 50% serum/50% PBS solution without OPNA acid target. This resulted in 12 sample-reference pairs with concentrations ranging from 100 to 0 nM of the OPNA acid target, with 500 nM aptamer in the binding sample and no aptamer in the reference. Here, a large concentration of the OPNA acid target (100 nM) was incubated with 500 nM of each aptamer in 25% serum (the binding sample), and 100 nM of the OPNA acid target was incubated with the 25% serum solution (the reference sample). sample).

From this simple approach we were able to identify which of the aptamer probes displayed the greatest signal for each OPNA acid target. These measurements also provide insight into the potential value of the assays in the context of the intended quantification. Saturation isotherms for each of the two best aptamer candidates for each of the OPNA acid targets enable quantification of the binding affinity, KD, to the ligand (OPNA acid).

Even the 2nd best performing aptamers produced a signal near the LOQ for the respective off-target quantification assay.

Quantitation of Opioids and the Prospect of Improved Diagnosis of Neonatal Abstinence Syndrome Improved Diagnosis of Neonatal Abstinence Syndrome

Before analyzing the next Dropix sample well, the capillary tube is thoroughly rinsed with PBS. The reference solution consisted of the same concentration of opioid nM incubated with a 50% urine/50% PBS solution without aptamer. As detailed in the SI, calibration curves for the opioid tests were obtained by creating a dilution series of the target at 0–100 nM in 50% urine/49.5% PBS/0.5% methanol with 1 nM aptamer, an RI matched reference and performing FSA-CIR analysis.

The slope in the linear range was used to calculate LOD (3×σ instrument baseline noise/slope) and LOQ (3×σ (of replicate determinations)/slope). Urinary oxycodone concentration: As illustrated in Figure B.7, the urinary oxycodone concentration is found by estimating the fraction (𝑓 ) of oxycodone that is metabolized to noroxycodone and assuming that the remaining oxycodone (𝑓) is present (excreted) in the urine. The CIR, described in detail in the SI, is a relatively simple device consisting of a sample introduction method (a commercial droplet generator (Mitos Dropix, Dolomite Microfluidics, UK), a diode laser (Lasermate, USA), a capillary flow cell (Polymicro , Molex, USA), a camera (Basler avA2300 2D CCD array, USA) and a syringe pump (Figures 5.1 and B.5).

Here, the large concentration of the aptamer measurably contributes to the signal (mass RI of the solution), therefore we reset the CIR by subtracting this aptamer background signal. With "cross-reactivity" signal magnitudes close to or below the LOQ of the assay (Figure B.3), the concentration of off-target species must be several times higher than the target concentration to introduce quantifiable imprecision into the LOQ. S.; Zhou, C.; Luo, F.; Xu, L., The Economic Burden of Prescription Opioid Overdose, Abuse, and Addiction in the United States, 2013.

T.; Kalso, E., Pharmacokinetics and metabolism of oxycodone after intramuscular and oral administration in healthy subjects.

Preclinical Evaluation of a Free

The free solution assay (FSA) presented here appears to represent a viable alternative [91] to many existing and emerging assays. After illumination of the capillary, a series of high-contrast elongated interference fringes is generated from the laser-capillary interaction. Block diagram of the CIR showing the droplet generator (left), the laser and edge detector (middle), and the syringe pump (right).

Spiked "unknowns" were prepared by adding a known amount of the CYFRA 21-1 biomarker to the pooled serum and then processing the samples in the same manner as patient samples. Briefly, comparison of sample-reference pairs, separated by a drop of oil, enables quantification of the target biomarker. Comparison of positional displacements in the adjacent regions of this reflected beam pattern (edges) as a train of sample and reference solution droplets traverses the beam provides a measure of the antibody-biomarker-target complex.

To quantify the target, an excess of antibody probe is added to one of the aliquots, giving Overall, within the linear range of the calibration curve (0 to 4 ng/mL), the FSA-CIR provided an average percent difference between actual and detected concentration of 10.9%. A Clinical Laboratory Improvement Amendments (CLIA)-accredited laboratory at the University of Maryland then independently analyzed aliquots of the same serum samples using the gold standard Roche Cobas 601 module, electrochemiluminescence (ECL) technology.

Here we tested a hypothesis that it was possible to redefine the concentration of the candidate biomarker CYFRA 21-1 in the control population of patients diagnosed with IPNs due to lower LOQs obtained by FSA-CIR measurements be provided.

CIR Analysis Tools

To calculate the ΔRIU of the test samples, the concentration of the product must first be calculated using the mass balance equation. This result is primarily due to the large differences between the masses of the species. A fixed concentration solution of the receptor is prepared in the matrix and then combined with the ligand dilution series, which is typically prepared in buffer.

Reference solutions are matrix solutions without receptor/target that have been combined with a dilution series. This process helps to ensure that the binding signal arises from a change in solution composition and not from uptake on the walls. Repetition and multiple iterations of the analysis tool to mitigate the contribution of injection to analysis error.

The success of the free solution test with our interferometer depends on a general understanding of the optical train reporting the signal. To illustrate the complexity of BSI, Figure A.7 presents optical beam traces of the optical train using. Methanol by adding 100 µl of the 200 nM target (prepared above) to 100 µl of pooled human urine (Valley Biomedical).

To produce the reference solution, 10 µL of each target dilution was combined with 10 µL of the RI-matched control solution. Specificity studies were performed for each of the aptamers to the target metabolite and cortisol. The reference sample was created by combining 10 µL of the unknown sample with 10 µL of PBS without aptamer.

Referensi

Dokumen terkait

Question: 3 Senator Kristina Keneally: In addition to the Assistance and Access Act, what additional, specific FTE requirements does the Ombudsman have in relation to the oversight