SPESIFIKASI TEKNIS
NO. NAMA ALAT DAN SPESIFIKASI ALAT VOLUME
PERALATAN ICU
1 VENTILATOR ADULT PEDIATRIC AND NEONATUS 2 Unit
Specifications :
VENTILATION MODES : ADULT AND PEDIATRIC
VOLUME (VCV) - Controlled
- Assisted/Controlled
- SIMV with support pressure with Backup - SIMV with support pressure without Backup
PRESSURE (PCV) - Controlled
- Assisted/Controlled
- SIMV with support pressure with Backup - SIMV with support pressure without Backup
SUPPORT PRESSURE (PSV) - PSV + CPAP
- Mandatory Minute Ventilation (MMV) + PSV - PSV + Assured Tidal Volume
- Biphasic Pressure (APRV / BIPAP)
NIV (NON INVASIVE VENTILATION) - with leakage compensation
BACKUP VENTILATION (SUPPORT) - Assisted/Controlled Volume
- Assisted/Controlled pressure
VENTILATION MODES : NEONATOLOGY - Assisted/Controlled pressure (PCV) - support pressure (PSV) / CPAP - NASAL CPAP
- SIMV (PCV) + PSV
- continuous flow, Time cycled with adjustable pressure
- Backup Ventilation (support) By Assisted and Controlled pressure
CONTROLS
- FiO2 21 to 100% (electronically adjusted on panel with graphic display visualization) - Inspiratory Time : 0.1 to 10 s.
- Ventilator frequenc: 1 to 150 rpm - Tidal volume : 10 to 2500 ml
- Apnea (with adjustable time) from 5 to 60 s.
- Sensibility:
By f: 0.5 to 10 l/min
By pr: -0.5 to -15.0 cmH2O (compensated PEEP)
- Controlled pressure (2 to 70 cmH2O over PEEP (with adjustable rise time) - Support pressure (0 to 70 cmH2O over PEEP (with adjustable rise time) - Inspiratory pressure: 2 to 120 cmH2O
- Expiratory sensibi: Adjustable from 5 to 60% of initial flow - PEEP/CPAP : 0 to 50 cmH2O
- Nebulization : Synchronized with inspiration, from 1 to 20 min - TGI : Synchronized with expiration
- Inspiratory flow:
In VCV : Automatic adjustment In PCV and PSV : up to 180 l/min Continuous flow in ne: 2 to 15 l/min Inspiratory flow in ne: 2 to 30 l/min : 0 to 120 l/min
- Expiratory flow
- Sigh (VCV mode) :- Cycles per hour - Quantity
- Maximum Tidal Volume - Manual activation - Inspiratory pause (VCV: 0 to 2.0 seconds - Manual activation
- O2 100% (Oxygenation for aspiration with synchronized system by 5, 10 or 15min, returning after that the initial FiO2)
- Flow waveform :
- In VCV : - Rectangular - Descendent - Sinusoidal - Ascendant - In PCV and PSV :- Descendent
- Gas net automatic By-pass (O2-AIR) in case of failure of one of them, the equipment continues operating normally.
- Inpiratory pressure inner safety valve adjusted in 120 H2Ocm - Stand By : to maintain the configuration without cycling
- Scales : for vertical and horizontal analysis with Automatic adjustment - Freeze : to analyze the graphs
OUTPUT PARAMETERS
- Airway pressure: peak, plateau, mean, base (PEEP) - Inspiratory Time
- E:I ratio / spontaneous breathings - Expired Tidal Volume
- Inspired flow peak - Dynamic compliance - Total frequency
- FiO2 concentration
RESPIRATORY MECHANICS - Auto PEEP
- Dynamic compliance - Static compliance
- Static Inspiratory resistance - Slow vital capacity
- PO. 1
- Tobin index (RSBI)
- P-V curve of insufflation In Low flow to determine inflection points
ALARMS
- High Inspiratory pressure - Low Inspiratory pressure - Apnea
- Maximum respiratory frequency - Maximum and minimum Tidal Volume - Maximum and minimum Minute Volume
- Maximum and minimum PEEP and continuous pressure - Gas supply source (air-O2)
- Power outage - Low battery charge
- Microprocessor (Technical failure) - Interrupted cycle In pressure modes - Inverted E:I ratio
- Alarms are triggered according to priority order with light, sound and screen message.
GRAPHICS
- pressure + Volume + flow / Time - pressure + flow / Time
- pressure / Volume Loop - flow / Volume Loop
TENDENCY CURVES (LAST 6, 12 OR 18 HS) - pressure
- flow
- Tidal Volume - Minute Volume - Dynamic compliance - frequency
OTHER FUNCTIONS
- Alarms log with last 24hs or last 1000 alarms with date, hour and cause of alarm - log of consulted values In respiratory mechanics
- circuit test without turning off the equipment
- Menu of indication of use hours and services provided - Altitude adjustments for Volume compensation - Language and Time/date selection
SELF TESTS
- Patient circuit leakage - Expiratory flow
POWER SUPPLY
- Universal input : 90 to 264 Vac – 47 to 63Hz
- Rechargeable inner battery with at minimum 120 min autonomy Termasuk Humidifier dan Air Compressor
2 TOUCH SCREEN PATIENT MONITOR 2 Unit
Specifications :
* Application
- Neonatal, pediatric and adult patient
* Operation Environtment - Power
+ Sourch : External AC power or internal battery + AC Power : 100 -240 VAC, 50/60 Hz, < 150 VA Rechargeable Lad-Acid
Operating time under the normal condition : 1 hour
Operatng time after first alarm of low battery : 10 mines - Temperature
+ Working : 5 -40° C + Storage : -20 - 65° C - Relative
+ Working : 30 - 75 % + Storage : < 80 % - Altitude
+ Operating Altitude : Up to 5,000 Meters + Hyperbaric Pressure : Up to 405.3 kPa
* Performance Spesifications
- Display : 12.1 " color TFT, Touch Screen - Resolution : 800 x 600 Pixels
- Trace : 4, 6, 8 or 9 waveforms - Indicator : Alarm indicator
Power indicator
QRS beep and alarm sound - Trend Time : 72 Hours
- Recorder : Built-in, thermal array, 2 channels - OxyCRG : Combinies HR, SPO2 and Respiration trends into a single graph
- ECG
+ Input : 5 - Lead ECG and standard AMMI line for connector
+ ECG waveforms : 7 channels
+ Penetration Voltage : 4000 VAC 50/60 Hz - ECG (cont.)
+ Sweep Speed : 12.5, 25 and 50 mm/s + HR Display Range : 15 - 300 bpm
+ Accuracy : ± 1 % 0r ± 1 bpm, whichever is greater + S-7 Segment Detection : 7 Channels
+ Arrythmia Analysis : 13 types - RESP
+ Sweep Speed : 12.5 and 25 mm/s - NIBP
+ Measuring Technologi : Automatic oscillsting measurement + Cuff Inflating : < 30 s (0 - 300 mmHg, standard adult cuff) + Measuring Period : AVE < 40 s
+ Mode : Manual, Auto, STAT + Measuring Interval in AUTO mode : 2 min - 4 hrs + Resolution : 1 mmHg
+ Accuracy : Pressure
Maximum Mean Error : Maximum Standard deviation : 8mmH + Overpressure Protection
Adult Mode : 280 (mmHg) Neonatal Mode : 150 (mmHg) + Alarm Limit Setting
SYS : 50 - 240 DIA : 15 - 180 - TEMP
+ Range : 0 - 50 ° C (32 - -122 ° F) + Accuracy : Without Sensor
+ Display Resolution : 0.1 ° C + Alarm Upper - Lower Limit Setting
Upper Limit : 0 - 50 ° C Lower Limit : 0 - 50 ° C + Channel : 2 Channels, Provide T1, T2, T - SPO2
+ ASpO2 : Anti-motion SpO2 + SpO2% Range : 0 - 100 %
+ SpO2 Accuracy : ± 2 % 70 - 100 %, non motion ± 3 % 70 - 100 %, motion + Pulse Rate Range : 30 - 250 bpm
+ Pulse Rate Accuracy : ± 2 bpm (non motion) ± 3 bpm (motion) + Alarm Upper - Lower Limit Setting + SpO2 Probe
Red light LED wavelenght Infrared light LED wavelenght
3 ECG 3 CHANNEL 1 Unit
Specifications :
Power Supply : AC Main class II and rechargeable NiMH batteris, 8 X 1,2 Volt 1500 mAh
Battery Power Cappacity : 3 hours
Writing System : Thermal printer, 8 dot/mm, useable print height 50 mm Thermosensitive Paper
Rolls : height 60 mm, leght 15 mm, gridded Z - Fold : Leght 20 m, page 70 x 60 mm, gridded
Monitor : Backlit graphic display 32 x 120 pixels (60 x 16mm)
Signal memory : 10 seconds each lead in automatic mode Print Channels : 3
Print format : 1, 2, 2 + R and 3 Operating Modes
Manual : acquisition and printing in real time Automatic : Simultaneous acquisition
Pre- programed : Simultaneous acquisitions at programmable intervals
Recording Sensitivity
Manual : 5 - 10 - 20 mm/mV + 5 %
Automatic : according to number of channels printed Paper Transport speed : 25 - 50 mm/s + 5 % Defibrillator Protection : Internal
Input Dinamics : '+ 300 mV @ 0 Hz + 5 mV in pass band Input Impedance : > 100 M ohm o eavh electrode Common Mode Rejection : > 90 dB
Frequency Response : 0,05 + 150 Hz (-3dB) without filters time constant : 3,3 S
Acquisition : 11 bit
1000 samples/s/channels printings and filetrs 500 samples/s/channels in calculations and filters resulotions 5uV/bit
Pace maker recognition : Recognizes pulse in accordance with current IEC standart Filters
Mains and muscle interference : modified digital notch 50 - 60 HZ, Switch on / switch off filters
Anti - drift : Digital high pass 0,5 HZ, linier phase, always enabls and without morpological distortion
Serials Interface : Infrared
Diagnostics Programs : HES electrocardiogram interpreter : develop at at medizinische
Dimension : 250 x 60 x 185 mm (length x height x depth) Weight : 1000 grams with batteries, without paper
Safety and conformity : Class II, type CF, Ref : EN 60601 - 1 to standard EN 60601 - 2 - 25, IEC 60601-2-51 CE0470.
4 TRANSPORTASI INCUBATOR , Servo Control 1 Unit
Specifications :
Power source : AC: 220 V/ 50 Hz DC: 6 x 12V/7.2 Ah
Auxiliary Battery (Car plug) DC 12V Power consumption : 250 W
Power supply controller : Inverter 220 VAC/500 W
Battery : 6 x 12V/7.2 Ah Sealed Lead Acid Rechargeable batteries Battery life time : 3 hours
Physical Specifications
Incubator dimension : App.75 L x 54 W x 67 H cm
Trolley dimension : App. 102 L x 57 W x (31 - 60) H cm (folded trolley for ambulance fixation) Weight : Incubator : App. 80 kg
IV Plole : Adjustable with fixation clamp Tray : For instrument with max. load 2.5 kg
Hood Specifications
Mattress size : App. 60 cm x 35 cm Mattress to hood : App. 22 cm
Access type : 3 port doors with cuff, 1 front access panel, 1 side access panel, 3 tubing grommets Characteristics
Control panel type : Hidden keys type, auto self-test system Control mode : Servo (Skin mode) and manual (Air mode) Controller type : Microcontroller
Control panel lock : Yes, automatic key lock after 10 seconds unchanged Air model control range : 25.0 °C - 37.0 °C, Override 39.0 °C, Resolution 0.1 °C Skin mode control range : 34.0 °C - 37.0 °C, Override 38.0 °C, Resolution 0.1 °C Warm-up time : < 30 minutes
Humidity range : 15 - 90 % RH(A) displayed by LCD hygrometer Noise level in hood : < 50 dB(A)
Air filter : Micro filter
Alarm indicator : Visible (16 x 2 dot-matrix LCD with blacklight) Audible (Buzzer, tri-mode sound)
Alarm silencer : 2 minutes for low backup battery, 10 minutes for others Information on LCD : Heater level, alarms, serial number, backup battery level and User ID (by request)
Alarms
• System faliure • High temperature (38.0°C or 40.0°C) • Low backup baterry • Skin over temperature (37.5°C or 38.5°C) • Sensor failure • Air over temperatur (37.5°C or 39.5°C) • Blower failure • Warmn-up failure (> 30 minutes)
• Airflow failure • Skin deviation ± 1°C and air deviation ± 3°C • Probe failure (probe disconnected)
5 INFUSION PUMP 2 Unit
Specifications :
Peristaltic finger - Delivery rate :
* Infusion sets : 15, 19 or 20 drops/mL * Infusion sets : 60 drops/mL
- Accuracy :
- "mL/h" control mode : ± 10% (using water of sodium chloride Solution and the recommended infusion set)
- Drops/min. control mode : ± 2% (10 minutes after the start of infusion, and after solution drips more than 500 times)
- Alarms :
AIR/OCCLUSION/EMPTY/FLOW ERR./DOOR/LOW BATT/FREE FLOW/BATT ERR./DRIP SENSOR ERR - Infusion complete function :
When the total volume delivered reaches the preset delivered limit, light and buzzer sound notifies the operator
Outputs an alarm relay contact (can be connected to nurse-call) - Special functions :
* Volume memory function * Repeat alarm function * Start reminder alarm function - Power supply :
* AC 100 to 240 V, 50/60 Hz
* Internal Nickel-Cadmium batt. can run the pump continuously for about 2 hours. (If the battery is new & the pump operates at the delivery rate of 25 mL/h after 12 hour's changing with the power off)
- Power consumption : Less than 16 VA
6 SYRINGE PUMP 2 Unit
Specifications :
Compatible Syringe 10, 20, 30, 50 mL Flow Rate Display Change :
0.1mL/h steps (for settings from 0.1 to 99.9 mL/h) 1mL/h steps (for settings from 100 to 999 mL/h) Flow Rate Accuracy :
Mechanical accuracy : ± 1%
Accuracy including syringe : within ± 3% Priming (purge rate) :
300mL/h - 1200 mL/h (depending on syringe size) Alarms :
Occlusion, Nearly empty, Low Battery, Plunger/Clutch disenganged, Syringe disenganged, AC/DC Cable disconnected.
Special functions :
Functions are selectable via internal switches : - Delivery Limit setting
- Nearly Empty alarm setting
The time until the syringe is empty can be programmed (from 3 to 60min) Power Supply :
100 to 240VAC, 50/60Hz 12 to 15VDC
Operable duration using internal battery NiMH:
Approx.5 hours (using a new battery at flow rate of 5mL/h and ambient temperature of 25ºC, after charging at least
15 hours with the power off)
7 ICU-ICCU BED ELECTRIC TYPE PLASTIC PANEL + MATTRESS 3 Unit
Specifications :
CONSTRUCTION - MAIN FRAME
Made of square steel pipes - PANELS
Head and foot panel made of plastic (can be removed in 3 second) - MATTESS DECK
Made of ABS (easy to be cleaned) ADJUSTMENT RANGE
Back rest adjustable by actuator up to 80° Knee rest adjustable by actuator up to 45° Hi - Low adjustable by actuator
Trendelenburg -15° / 15° ACCESSORIES
- Bumper Ø 4" - Patient board - Infusion stand
- Fold down side guard made of plastic FINISHING
Powder coating CASTOR
Ø 5", central lock castor DIMENSION
2215 (L) X 900 (W) X 1037 - (1290) (H) mm Include : Mattress
PERALATAN NEONATOLOGI
1 VENA SECTION INSTRUMENT SET 2 Set
Consisting of :
SCALPEL HANDLE NO 3 1
FEILCHENFELDT SPL FCPS 11.5 2
TISSUE FCPS 10,5CM 1X2T 1
BABY-METZENBAUM SCISSORS CVD 14,5CM 1
SURG SCS STR SH/SH 16,5 CM 1
KOENIG LIGATURE GUIDE 19,5CM 1
DESCHAMPS NEEDLE BL R 20,5 CM 1
DESCHAMPS NEEDLE BL L 20,5CM 1
MAYO HEGAR NH 16 CM 1
CONVERSE HOOK SM 18,5 CM 2
CONVERSE HOOK LG 18,5 CM 2
HALSTEAD MOSQ FCPS STR 12 1
HALSTEAD MOSQ FCPS CVD 12 1
GROSS DRESS FCPS STR 20 CM 1
BACKHAUS TOWEL FCPS 9 CM 2
FINSEN RETRACTOR BLUNT 5 CM 1
DESMARRES LID RETR 13X18 MM 1
APPLYING/REMOVING FCPS 23 CM 1
TITAN BULLDOG CL STR 3 CM V 1
TITAN BULLDOG CL STR 3 CM A 1
INSTRUMENT BOX 205X105X35MM 1
2 MINOR SURGERY SET 1 Set
CLINIC SCALPEL HANDLE NO 3 1
GILLIES TISSUE FCPS 1X2 15CM 1
MCINDOE DRESS FCPS 15 CM 1
MAYO HEGAR NH 16 CM 1
CLINIC CRILE WOOD NH 15 CM 1
CLINIC HALSTED-MOSQ FCPS CVD 12CM 4
CLINIC MICRO-HALSTED FCPS CVD 12CM 4
CLINIC MAYO DISS SCS CVD 14,5 CM 1
CLINIC MAYO DISS SCS STR 14,5 CM 1
CLINIC BABY METZENBAUM SCS CVD 14,5 1
CLINIC FRAZIER SUCTION TUBE 4.0 MM 1
BALFOUR MOR. RETR W RATCH 215 1
BABY BALFOUR RETRACTOR 1
BABCOCK INT TISS FCPS 17CM 1
RICHARDSON RETR 24CM 43X39MM 1
CLINIC BACKHAUS TOWEL FCPS 13,5 CM 4
BABY KOCHER FCPS STR 14 CM 4
KOCHER ART FCPS 1X2 STR 13CM 4
CLINIC FOERSTER FCPS STR 24,5 SERRATED 1
07ROUND BOWL 80X40MM 0,14L 1
3 MANUAL RESUSCITATION SET NEONATAL 1 Set
Consist of :
Resuscitation Valve Unit, Child/Neonatal combined 1
Silicone Resuscitation Bag, Neonatal 1
Inhalation Valve Unit 1
Oxygen Reservoir Valve Unit 1
Oxygen Reservoir Bag 600 ml 1
Silicone Mask, Neonatal, Large 1
Silicone Mask, Neonatal, Small 1
Airway, Neonatal, Large 1
Airway, Neonatal, Small 1
E agi ® Case 1
4 INFANT WARMER 1 set
Specifications :
OPERATIONS
Skin temperatur controller heater - Heating power : 500 W Quartz
- Skin temperature setting range : 34°C - 37.9°C - Setting heating power manuel : 0% - 100% - Display range LCDisplay : 0.0°C - 99.0°C - Warm-up time accordin to IEC : 45 minutes - Bed tilting : ± 10°
- Horizontal swiveling : ± 60°
- Observation lamp : 12V / 50W / 40° ALARMS
Supply voltage control Lost of mains power Temperature control
Completed with :
- Skin Temperature Probe 1
- Stand 1
- X-Ray Tray 1
- Compartement with 2 drawer 1
- Mattress 1
5 LARYNGOSCOPE SET 1 Set
Consists of :
Laryngoscope Blade w/lamp, Adult 1
Laryngoscope Blade w/lamp, Child 1
Laryngoscope Blade w/lamp, Neonatal 1
Laryngoscope Handle w/batteries 1