BALAI BESAR PENGAWAS OBAT DAN MAKANAN DI DENPASAR
Jl. Tjut Nya Dien No.5 Denpasar Bali, Telp. (0361) 223763, 234597 Fax. (0361) 225395
A D D E N D U M D O K U M E N P E N G A D A A N
Nomor: PR.01.05.107.03.12.0775
Tanggal: 2 Maret 2012
untuk
PENGADAAN ALAT LABORATORIUM
PANITIA:
BALAI BESAR PENGAWAS OBAT DAN MAKANAN (BBPOM)
DI DENPASAR
Tahun Anggaran:
BALAI BESAR PENGAWAS OBAT DAN MAKANAN DI DENPASAR
Jl. Tjut Nya Dien No.5 Denpasar Bali, Telp. (0361) 223763, 234597 Fax. (0361) 225395
BAB I. UMUM
TETAP / TIDAK ADA PERUBAHAN
BAB II. PENGUMUMAN PELELANGAN UMUM DENGAN PASCAKUALIFIKASI
4. Jadwal Pelaksanaan Pengadaan
berubah menjadi
Jadwal dapat dilihat pada website LPSE.
No
Kegiatan
Hari/Tanggal
Waktu (wib)
a.
Pendaftaran dan Down Load
Dokumen Pengadaan
Senin, 27 Pebruari s/d Sabtu, 10 Maret
2012
b.
Pemberian Penjelasan
Jumat, 2 Maret 2012
08.00 – 11.00
c.
Upload Dokumen Penawaran
dan Dokumen Kualifikasi
Sabtu, 3 Maret s/d Minggu, 12 Maret
2012
d.
Pembukaan Dokumen
Penawaran
Senin, 13 Maret 2012
08.00 - 15.00
e.
Evaluasi Penawaran
Selasa, 13 Maret s/d Minggu 25 maret
2012
f.
Pengumuman Pemenang
Senin, 26 Maret 2012
g.
Masa Sanggah
Selasa, 27 Maret s/d Senin, 2 April 2012
h.
Penerbitan SPPBJ
Selasa, 3 April 2012
i.
Kontrak
Selasa,17 April – Selasa, 14 Agustus 2012
BAB III. INSTRUKSI KEPADA PESERTA
B.
DOKUMEN PENAWARAN
8. Isi Dokumen Pengadaan
8.2 k (10) Surat Pernyataan memberikan Jaminan Purna
Jual dari Distributor (pemberian garansi alat
selama 1 tahun dan suku cadang minimal 10
tahun)
untuk alat No. 1 (AAS), 4 (Automatic
TLC Sample Application Instrument), 5
(Automatic Titration Karl Fischer and
Potensiometer), 7 (Automatic Elisa Reader),
14
(Dissolution
Tester
System),
15
(Dissintegration Tester Digital), 32 (GCMS), 36
& 37 (HPLC), 43 (LCMS-MS), 46 (Micro
Balance), 73 (UV-Vis Spectrofotometer).
8.2 k (16) Daftar Populasi
untuk alat No. 1 (AAS), 4
(Automatic
TLC
Sample
Application
Instrument), 5 (Automatic Titration Karl
Fischer and Potensiometer), 7 (Automatic
Elisa Reader), 14 (Dissolution Tester System),
15 (Dissintegration Tester Digital), 32 (GCMS),
36 & 37 (HPLC), 43 (LCMS-MS), 46 (Micro
Balance), 73 (UV-Vis Spectrofotometer).
BALAI BESAR PENGAWAS OBAT DAN MAKANAN DI DENPASAR
Jl. Tjut Nya Dien No.5 Denpasar Bali, Telp. (0361) 223763, 234597 Fax. (0361) 225395
15. Dokumen Penawaran
E.
PEMBUKAAN DAN EVALUASI
PENAWARAN
b.
Hasil pemindaian (scan) Jaminan Penawaran. Dalam
hal pencairan jaminan memerlukan dokumen asli
jaminan maka jaminan penawaran asli dikirim melalui
pos/jasa pengiriman ke alamat
PANITIA PENGADAAN
ALAT LABORATORIUM BBPOM DI DENPASAR, d/a Jl.
Tjut Nya Dien No.5 Denpasar 80235, paling lambat
tanggal 17 Maret 2012 diterima ditempat tujuan
akhir.
e. 12) Surat Pernyataan kesanggupan memberikan
training terhadap 2 (dua) orang personel
laboratorium tentang perawatan alat ditempat
penyedia, biaya sepenuhnya ditanggung oleh
penyedia sesuai SBM (Standar Biaya Masukan)
Menteri Keuangan RI untuk alat No.
1 (AAS), 4
(Automatic
TLC
Sample
Application
Instrument), 32 (GCMS), 43 (LCMS-MS).
13) Surat Pernyataan kesanggupan melakukan
kunjungan service minimal 2 kali dalam masa
garansi untuk alat No.
1 (AAS), 4 (Automatic
TLC Sample Application Instrument), 5
(Automatic
Titration
Karl
Fischer
and
Potensiometer), 7 (Automatic Elisa Reader), 14
(Dissolution Tester System), 15 (Dissintegration
Tester Digital), 32 (GCMS), 36 & 37 (HPLC), 43
(LCMS-MS), 46 (Micro Balance), 73 (UV-Vis
Spectrofotometer).
16) Daftar Populasi
untuk alat No. 1 (AAS), 4
(Automatic
TLC
Sample
Application
Instrument), 5 (Automatic Titration Karl Fischer
and Potensiometer), 7 (Automatic Elisa
Reader), 14 (Dissolution Tester System), 15
(Dissintegration Tester Digital), 32 (GCMS), 36
& 37 (HPLC), 43 (LCMS-MS), 46 (Micro Balance),
73 (UV-Vis Spectrofotometer).
28.11 h. Surat Pernyataan memberikan Jaminan Purna
Jual dari Distributor (pemberian garansi alat
selama 1 tahun dan suku cadang minimal 10
tahun)
untuk alat No. 1 (AAS), 4 (Automatic TLC
Sample Application Instrument), 5 (Automatic
Titration Karl Fischer and Potensiometer), 7
(Automatic Elisa Reader), 14 (Dissolution Tester
System), 15 (Dissintegration Tester Digital), 32
(GCMS), 36 & 37 (HPLC), 43 (LCMS-MS), 46
(Micro Balance), 73 (UV-Vis Spectrofotometer).
BALAI BESAR PENGAWAS OBAT DAN MAKANAN DI DENPASAR
Jl. Tjut Nya Dien No.5 Denpasar Bali, Telp. (0361) 223763, 234597 Fax. (0361) 225395
l. Surat Pernyataan kesanggupan memberikan
training terhadap 2 (dua) orang personel
laboratorium tentang perawatan alat ditempat
penyedia, biaya sepenuhnya ditanggung oleh
penyedia sesuai SBM (Standar Biaya Masukan)
Menteri Keuangan RI untuk alat No.
1 (AAS), 4
(Automatic
TLC
Sample
Application
Instrument), 32 (GCMS), 43 (LCMS-MS).
m. Surat Pernyataan kesanggupan melakukan
kunjungan service minimal 2 kali dalam masa
garansi untuk alat No.
1 (AAS), 4 (Automatic
TLC Sample Application Instrument), 5
(Automatic
Titration
Karl
Fischer
and
Potensiometer), 7 (Automatic Elisa Reader), 14
(Dissolution Tester System), 15 (Dissintegration
Tester Digital), 32 (GCMS), 36 & 37 (HPLC), 43
(LCMS-MS), 46 (Micro Balance), 73 (UV-Vis
Spectrofotometer).
p.
Daftar Populasi
untuk alat No. 1 (AAS), 4
(Automatic
TLC
Sample
Application
Instrument), 5 (Automatic Titration Karl Fischer
and Potensiometer), 7 (Automatic Elisa
Reader), 14 (Dissolution Tester System), 15
(Dissintegration Tester Digital), 32 (GCMS), 36
& 37 (HPLC), 43 (LCMS-MS), 46 (Micro Balance),
73 (UV-Vis Spectrofotometer).
BAB IV. LEMBAR DATA PEMILIHAN
L. BATAS AKHIR WAKTU
UPLOAD DOKUMEN
PENAWARAN
M. PEMBUKAAN PENAWARAN
Hari : Senin
Tanggal : 12 Maret 2012
Jam: 23.59
Hari : Selasa
Tanggal : 13 Maret 2012
Jam: 08.00 – 15.00
BALAI BESAR PENGAWAS OBAT DAN MAKANAN DI DENPASAR
Jl. Tjut Nya Dien No.5 Denpasar Bali, Telp. (0361) 223763, 234597 Fax. (0361) 225395
BAB V. LEMBAR DATA KUALIFIKASI
TETAP / TIDAK ADA PERUBAHAN
BAB VI. BENTUK DOKUMEN PENAWARAN
A. BENTUK SURAT PENAWARAN
Sesuai dengan persyaratan dokumen pengadaan, bersama surat penawaran ini kami
lampirkan:
8) Surat Pernyataan memberikan Jaminan Purna Jual dari Distributor (pemberian
garansi alat selama 1 tahun dan suku cadang minimal 10 tahun)
untuk alat No. 1
(AAS), 4 (Automatic TLC Sample Application Instrument), 5 (Automatic Titration Karl
Fischer and Potensiometer), 7 (Automatic Elisa Reader), 14 (Dissolution Tester
System), 15 (Dissintegration Tester Digital), 32 (GCMS), 36 & 37 (HPLC), 43
(LCMS-MS), 46 (Micro Balance), 73 (UV-Vis Spectrofotometer).
12)
Surat Pernyataan kesanggupan memberikan training terhadap 2 (dua) orang
personel laboratorium tentang perawatan alat ditempat penyedia, biaya
sepenuhnya ditanggung oleh penyedia sesuai SBM (Standar Biaya Masukan)
Menteri Keuangan RI untuk alat No.
1 (AAS), 4 (Automatic TLC Sample Application
Instrument), 32 (GCMS), 43 (LCMS-MS).
13)
Surat Pernyataan kesanggupan melakukan kunjungan service minimal 2 kali dalam
masa garansi untuk alat No.
1 (AAS), 4 (Automatic TLC Sample Application
Instrument), 5 (Automatic Titration Karl Fischer and Potensiometer), 7 (Automatic
Elisa Reader), 14 (Dissolution Tester System), 15 (Dissintegration Tester Digital), 32
(GCMS), 36 & 37 (HPLC), 43 (LCMS-MS), 46 (Micro Balance), 73 (UV-Vis
Spectrofotometer).
16)
Daftar Populasi
untuk alat No. 1 (AAS), 4 (Automatic TLC Sample Application
Instrument), 5 (Automatic Titration Karl Fischer and Potensiometer), 7 (Automatic
Elisa Reader), 14 (Dissolution Tester System), 15 (Dissintegration Tester Digital), 32
(GCMS), 36 & 37 (HPLC), 43 (LCMS-MS), 46 (Micro Balance), 73 (UV-Vis
Spectrofotometer).
B. BENTUK DOKUMEN PENAWARAN TEKNIS
h. Surat Pernyataan memberikan Jaminan Purna Jual dari Distributor (pemberian garansi
alat selama 1 tahun dan suku cadang minimal 10 tahun)
untuk alat No. 1 (AAS), 4
(Automatic TLC Sample Application Instrument), 5 (Automatic Titration Karl Fischer
and Potensiometer), 7 (Automatic Elisa Reader), 14 (Dissolution Tester System), 15
(Dissintegration Tester Digital), 32 (GCMS), 36 & 37 (HPLC), 43 (LCMS-MS), 46 (Micro
Balance), 73 (UV-Vis Spectrofotometer).
l.
Surat Pernyataan kesanggupan memberikan training terhadap 2 (dua) orang personel
laboratorium tentang perawatan alat ditempat penyedia, biaya sepenuhnya
ditanggung oleh penyedia sesuai SBM (Standar Biaya Masukan) Menteri Keuangan RI
untuk alat No.
1 (AAS), 4 (Automatic TLC Sample Application Instrument), 32 (GCMS),
43 (LCMS-MS).
BALAI BESAR PENGAWAS OBAT DAN MAKANAN DI DENPASAR
Jl. Tjut Nya Dien No.5 Denpasar Bali, Telp. (0361) 223763, 234597 Fax. (0361) 225395
m.
Surat Pernyataan kesanggupan melakukan kunjungan service minimal 2 kali dalam
masa garansi untuk alat No.
1 (AAS), 4 (Automatic TLC Sample Application
Instrument), 5 (Automatic Titration Karl Fischer and Potensiometer), 7 (Automatic
Elisa Reader), 14 (Dissolution Tester System), 15 (Dissintegration Tester Digital), 32
(GCMS), 36 & 37 (HPLC), 43 (LCMS-MS), 46 (Micro Balance), 73 (UV-Vis
Spectrofotometer).
p.
Daftar Populasi
untuk alat No. 1 (AAS), 4 (Automatic TLC Sample Application
Instrument), 5 (Automatic Titration Karl Fischer and Potensiometer), 7 (Automatic
Elisa Reader), 14 (Dissolution Tester System), 15 (Dissintegration Tester Digital), 32
(GCMS), 36 & 37 (HPLC), 43 (LCMS-MS), 46 (Micro Balance), 73 (UV-Vis
Spectrofotometer).
BAB VII. SYARAT-SYARAT UMUM KONTRAK (SSUK)
TETAP / TIDAK ADA PERUBAHAN
BAB VIII. SYARAT-SYARAT KHUSUS KONTRAK (SSKK)
S. LAYANAN TAMBAHAN
1. Surat Pernyataan kesanggupan memberikan training
terhadap 2 (dua) orang personel laboratorium
tentang perawatan alat ditempat penyedia, biaya
sepenuhnya ditanggung oleh penyedia sesuai SBM
(Standar Biaya Masukan) Menteri Keuangan RI untuk
alat No.
1 (AAS), 4 (Automatic TLC Sample Application
Instrument), 32 (GCMS), 43 (LCMS-MS).
2.
Surat Pernyataan kesanggupan melakukan kunjungan
service minimal 2 kali dalam masa garansi untuk alat
No.
1 (AAS), 4 (Automatic TLC Sample Application
Instrument), 5 (Automatic Titration Karl Fischer and
Potensiometer), 7 (Automatic Elisa Reader), 14
(Dissolution Tester System), 15 (Dissintegration
Tester Digital), 32 (GCMS), 36 & 37 (HPLC), 43
(LCMS-MS),
46
(Micro
Balance),
73
(UV-Vis
Spectrofotometer).
LAMPIRAN
No. SATUAN JUMLAH
1 Unit 1
Specifications :
a. Analysis Methods : Flame & Graphite Furnace
b. Optic system : double beam method
c. Measured wavelength : 190 - 900 nm or wide
d. Bandwith : minimum 4 - step
e. Detector : Photomultiplier or equivalent
f. Background correction : Deuterium and BGC-SR or Zeeman method
g. Number of lamps : minimum 8-lamps
h. Gas Controllers : Automatic
Flame Operation
a. Burner Head : Titanium
b. Spray Chamber : Engineering plastic or other corrosion resistant material
c. Positioning Control : Automatic
d. Absorption at 5 ppm Cu : > 0.7 AU
e. RSD at 5 ppm Cu : < 0.5 %
Graphite Furnace Operation + Camera
a. Temperature Range : 50° to 2600° C or higher
The Equipment provided with :
a. Autosampler 2 unit
b. Mercury & Hydride Analysis Unit 1 unit
c. Dilution & Online Calibrator Unit 1 unit
d. Water cooling system 1 unit
e. Workstation 1 set
Hardware
Software
- Parameter setting Wizard method or compatible Include of:
Installation Training
Original Operasional manual Guarantee min. a year Factory Certificate Dust cover
Tool kit 1 set
Air/Acetylene burner head 1 set
HC Lamp, Pb, Hg, Cd, Zn, As, Mg, Ca, Na, K, Cu, Fe, Cr, Se, Sn, Ag (each 1 lamp)
Requirement reagent (Mg Perchlorat, KMnO4, NaBH4) 1 set
Requirement reagent for function test 1 set
Logam Cu references solution, aquademineralisata. 1 set
(with water cooling system for GFA)
- PC ( Processor min. Core i5, RAM min. 4 Gb DDR3, HD min. 500 Gb, DVD R+W multy, motherboard, LAN ) + LCD Monitor min 19 " + mouse + Keyboard + Printer inkjet ( color min. 5800 x 1400 dpi ) + OS compatible
NAMA BARANG DAN SPESIFIKASI TEKNIS ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*) 2 set
Compressor 1 unit
Pyro-coated graphite tube 1 set
High density graphite tube 1 set
Ducting system for exhaust gas 1 set
Stabilizer 10 KVA 1 unit
UPS 10 KVA 1 unit
2 UNIT 2 Specifications : a. Capasity : maksimal 220 g b. Readability : 0.0001 g c. Repeatability : ≤ 0.1 mg d. Linearity : ≤ 0.2 mg e. Display : Custom LCD
f. Connecitivity : interface connection for PC or printer
Included of
a. Installation b. Garansi min 3 tahun
c. Original operation manual book d. Dust Cover
e. Weight Standard Class E2 Capasity 100 & 200 g
3 ANAEROBIC JAR UNIT 3
Spesifications:
a. Capacity : min. 3,5 liters
b. Accessories
- Petridish rack : 1 unit
(plate basket 12 petridish)
Include of : Original Operational Manual Book Garansi minimal 1 tahun
4 AUTOMATIC TLC SAMPLE APPLICATION INSTRUMENT *) Set 1
consist of
1 Automatic TLC sampler Automatic TLC sampler : 1 unit
a. object support up to 20 x 20 cm
b. Fully controllable through Operational software c. Suitable for standart sample vial and 96 well plates d.
e. Gas supply 4-6 bar (60-90 Psi) preferabli nitrogent Gas supply 4-6 bar (60-90 Psi) preferabli nitrogent
2 Automatic Developing Chamber : 1 unit
a. Fully Automatic development
b. Two mode of operation: stand alone or PC Control
c. Recording of detected developing distance and time, temperature and humidity d. Including humidity control
3 TLC Scanner 1 unit
Gases (Acetylene, Nitrous Oxide , Asam nitrat , and Argon) complete with tank and regulator
ANALYTICAL BALANCE
Accurate and reproducible, fast and reliable, easy handling, and IQ/OQ qualification for use in GMP/GLP enviroment
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)a. Light source & line spectrum : Deuterium lamp/ 190-400 nm or better
Halogen tungsten lamp/ 350-900 nm or better High pressure mercury lamp/ 254-578 nm or better b. Automatic ignition lamp
c. Optical system : Apochromatic suprasil-fluorite lens system, transmision
range 190-900 nm
d. Monochromator : Concave hologhapic grating, 1200 lines/mm or better
e. Bandwith selectable : 5/20 nm or better
Wavelength range : 190-900 nm
Reproducibility of WL setting : 0,2 nm or better
Accuracy : 1 nm or better
Max speed of spectra recording : 100 nm/s at oversampling rate 40/nm or greater
f. Scanning slit : revolving disk with 20 fixed apertures
g. Length of slit images : 0,5 - 12 nm or better, selectable
Width of slit images : 0,025 - 1,2 nm or better
h. Detector :
Reflectance mode : two matced broadband photomultiplier, multi alkali type
Spectral sensitivity : 185 - 850 nm or better
Linear range : 0 - 3,0 OD
4 TLC Visualizer 1 unit
a. Enhanced illumination unit : visible/ white light, UV 254, UV 366
b. Powerful high resolution CCD digital camera
c. Linearity : 12 bit = 4095 linear intensity levels
d. Fast images transfer e. Automatic image optimization f. Automatic background correction
5 TLC Sprayer : charger unit 1 unit
Pump unit with bottle clamps and connecting tube one each
6SPEKTRO DENSITOMETER TLC-MS INTERFACE 1 unit
a. Automation : semi automatic instrument involving automatic piston
movement, automatic cleaning of piston, manual positioning and switching
b. TLC plates can be positioned accurately and analyzed zone by zone
Include of : Installation
Training
Original operasional manual book Guarantee min. 1 year
Gas N2 + Cylinder + Regulator 2 set
PC ( Processor min. Core i5, RAM min. 4 Gb DDR3, HD min. 500 Gb, DVD R+W multy, motherboard, LAN ) + LCD Monitor min 19 " + mouse + Keyboard + Printer inkjet ( color min. 5800 x 1400 dpi )
2 unit
Software
Sample vial rack for 66 vials 1 unit
Vials 100 pcs
Screw cap bottle 250 mL 2 pcs
Syringe rinsing tube TLC foil down holder
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)Dosing syringe 10 µL for ATS 4 w/ O needle
Compensatin ledge
Humidity Control for automatic chamber Spectrum Library
Track Optimization Multi wavelength scan
IQ/OQ TLC Scanner (inc. Test plate)
Pack of spray heads type A & B 5 pcs
Reagent Bottle with cap (100 & 50 mL) 2 pcs
5 AUTOMATIC TITRATION KARL FISCHER AND POTENSIOMETER *) Unit 1
Spesifications
a. Automatic burette : yes
b. short cut : maksimal 12 per user
c. Display : LCD
d. Number of burette driver : 1 internal + 3 external (optional, fuul titration control)
f. Flexible method creation : extended flexibility
g. KF Titration : yes
h. Sensor board : maks. 2
i. Task list (number of task) : max. 10, max. 7 paralel
j. Hot plug and play : yes
k. Plug & Play sensor : yes
l. End point titration : very suitable
m. GLP conform printout : yes
n. Password protection : User manager/user rights
o. EQP Titration : very suitable
p. pH-stat : very suitable
q. User guidance : Learn titration, online help
r. Redox, volta, complexom, non aqueous very suitable
s. Connection to printer : USB printer (PCL)
t. Connection to barcode reader : yes (USB)
Include of :
- Titration stand , solven manager and accessories 1 Unit
- Set for Karl Fischer with general titration stand 1 Unit
- Inter changeable burrete 5 mL 1 Unit
- Solvent manager kit 1 Unit
- Manual Titration kit with titration stand , compact stirrer and accessories 1 Unit
- USB Printer 1 Unit
Interface : USB
Printing speed : min. 2.0 lines per second
paper type : standar paper 57,5 mm (W), max. Ø 60 mm
Number of character per line : 5 x 7 , 24-character line length
Power line voltage, frequency : Internal power supply, 100-240VAC, 50/60 Hz
- pH glass electrode with moveable sleeve diaphragm-non-aqueous titration 1 Unit
- pH glass electrode 1 Unit
-Combined platinum ring electrode for redox titration (3M KCL,saturated with AgCl) 1 Unit
- Combined silver ring electrode for argentometric titration (1M KNO3) 1 Unit
- 20 mL burette 1 Unit
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)- Glass beaker, 80 mL, set of 20 pcs 1 Unit
-Installation -Training
-Original operasional manual book -Guarantee min. 1 year
6 UNIT 2
Specifications:
a. System : Automatic
b. Range of humidity control : up to 25 %
c. Power source : AC 100 - 220 V ; 50/60 Hz
d. Electric, door built-in hygrometer and temperature
e. Design : Vertical orientation
f. Dimensi : min. 280 x 270 x 480 mm
g. include of : Step down/up
Instalation
Guarantee min a year Original operation manual book 3 shelves
plate for desicant
7 AUTOMATIC ELISA READER *) UNIT 1
Spesifications
a. Type : Process the ELISA/EIA from sample identification (option) to
interpretation and print out of the result. Perform single phases, such as sample dilution , washing, readind etc.
b. Main Performances : Open systemfully programamble, walk away, random access,
Multi-batch execution, Continuous loading of samples and test kit
c. Maximum no. Of assay per run : maks. 12
d. Result calculation : Programamble Qualitative formula, standard curve
calculation in quantitative tests
e. Standard Curves : Point to point, linear and square regression, 4 parameters,
ratio. The curves can be stored and retrieved later, furthermore the curve can be adjusted using only 1 or 2 standards
f. Sampler
Micro-plate working capacity : minimal 3
No. of syring : 2
Syringes capacity : min. 800 µL
No. of Needles : 2
Tips : plastik mono-use tip, non conductive and low cost
Liquid detect ion : Pneumatic type
Clot detection : Pneumatic type
Aspired volume detection : Pneumatic type
g. Pipetting precision and accuracy : Less than 1%@10% of the syringe volume, based on gravimetric evaluation using H2O
h. Samples minimum detectable volume : min. 200 µL
i. Carry-Over : zeroed because of the disposable tips usage
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)j. Sample capacity : up to 96 primary tubes, 88with bar-code reader option
k. Micro-plates Incubators
No of incubators : 3
Type : Closed and darkened
Temperature range : each incubator can be programmed from + 2 ⁰C RT to 50 ⁰C,
in 0,1 ⁰C steps
Temperature precision : ± 0,2 ⁰C
Incubation time : programmable from 1 minute to 24 hours
micro-plates shaking : programmable
l. Micro-plates washer
Manifold needles : 8-way distribution channels and 8-way aspiration channels
No of washing buffers : 3, in 2,5 liter bottles
Feature : Programmable steps, delivery volume, soak time
m. Reader
Reading type : Single/dual wavelengthmeasurement
Wavelength range : from 400 to 750 nm ( 320 to 800 nmon request )
Include of
Computer : Standard / industrial PC integrated in the unit with hard Disk,
keyboard, mouse, Printer, and LCD display
Interfaces : USB, Ethernet, Parallel, Rs232, VGA,etc
Host connection : Using files and serial port
Original operation manual book Guarantee min. a year Instalation
Training
8 AUTOCLAVE DIGITAL UNIT 1
Spesifications :
a. Effective capacity : min. 110 L
b. Chamber size : min. Ø 420 mm x D 790 mm
c. Outer dimension : min. 660 x 1180 x 650 mm
d. max. Sterile Temperature : 135 ⁰C
e. Process mode : min. 4 mode
f. Net Weight : max. 85 kg
g. Electromechanical lock system : Dual sensing mechanism; Double sensing system for air evacuation; Programmable autostart; Process status display; Process mode for agar
h. Accessories : Forced cooling unit, Digital printer, automatic water supply
Included of : - min. 2 pcs SS wire basket - Instalation
- Guarantee min a year - Original operation manual book
9 AYAKAN SET 1
Each set consist of:
1 Ayakan Mesh 7
2 Ayakan Mesh 20
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)4 Ayakan Mesh 48
5 Ayakan Mesh 80 6 Ayakan Mesh 100 7 Ayakan Mesh 200 Include of a. Penutup + wadah 10 UNIT 3 Specifications : a. Speed (rpm) : min. 6000 b. Capacity : 15 mL, 50 mL, 100 mL c. Power requirements : 220-240; 50-60 Hz
d. Accessories : Swing-out Rotor, capacity 100 mL
Fixed-angle rotor, capacity 50 mL
Tube 100 mL 4 tubes
Tube 50 mL 6 tubes
Tube 15 mL 12 tube
Included of
a. Original operation manual book b. Garansi min 1 tahun
11 UNIT 1
Specifications :
a. speed : min. 4000 rpm
b. RCF : min. 2819 x g
c. capacity : 15, 100 mL
d. time : 1 - 99 menit atau lebih baik
e. display : Min. Digital atau lebih baik (speed, time, temp, RCF)
f. temperature : -9 °C to + 40 °C atau range lebih lebih lebar
g. stainless steel chamber h. Motor over heating protection i. replaceable rotor
j. automatic opening of lid
k. Electric supply : 230 V, 50 Hz
Included of
- 1 x angle rotor 16 x 15 mL - 1 x swing out rotor 4 x 100 mL - centrifuge tube 15 mL (16 pcs) - centrifuge tube 100 mL (16 pcs)
12 CLASS II MICROBIOLOGI BIOSAFETY CABINET *) UNIT 2
Spesifications: a. Dimensi external : min .1700 x 800 x 1530 mm internal : min.1570 x 610 x 670 mm b. Inflow average : 0.48 m/s (95 fpm) c. Down flow : 0.35 m/s (70 fpm )
d. Sound emision : EN 12469 < 56 dBA
e. ULPA filter downflow : > 99,999%
CENTRIFUGE
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)f. ULPA filter Exhaust : > 99,999%
g. Working Opening : 210 mm
h. Lamp intensity : 1200 lux (111 foot-candles)
i. Main body : 1.2 mm (0.05") 18 gauge electrogalvanized steel with white
oven
baked epoxy , isoside antimicrobial powder coated finish
j. Work surface : 1.5 mm ( 0.06") 16 gauge stainlees steel, type 304 with 4B
finish with arm rest
k. Side wall : 0.9 mm ( 0.035") 20 gauge stainlees steel, type 304 with 4B
finish
l. Accessories : UV lamp , support Stand, Electric outlet, motorized window
m.
Included of : - Instalation
- Guarantee min a year - Original operation manual book - Training
- UV Lamp
13 CLAMP FOR TWO BURETTE PCS 3
Spesification
a. Material : Light Alloy
b. panjang (mm) : 110
c. For rod Ø 12 mm
14 DISSOLUTION TESTER SYSTEM *) Unit 1
Spesifications
1. Dissolution Tester 1 unit
a. Flask temperatur Range : 30 to 40° C or better Accuracy : 0.5 ° C or better b. Spindle speed Range : 25 to 200 rpm or better Accuracy : ± 1 rpm or better
c. Shaft centering to vessel : < 1 mm
d. Vibration : maksimum 0,1 mils
e. Shaft woble : maksimum 1 mm TIR
g. Voltage : 220 V-240 V, 50/60 Hz
h. Supporting system including : 8 pcs USP/NF glass dissolution flask
8 pcs USP paddles
8 basket assembly 40 mesh basket 8 pcs sampling probe
2. Autosampler 1 unit
a. Include precision syringe pump sampling Sample Volume Accuracy
0.5 to 5.0 mL ≤ 0.06 mL
5.1 to 10.0 mL ≤ 0.12 mL
10.1 to 15.0 mL ≤ 0.18 mL
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)Change and save minimum up to 20, protocols, easy-to-use routines, password
protect, Built in Validation, calibration and diagnostics, plus unique, advanced features, Teflon valves, media replace and system rinse & wash.
Stainless probes for media replace and waster; 8 sample system 24 hours sampling time (sustained release). Media replace and system rinse
3. Sampler Collector 1 unit
a. Removable collection racks allow for sample handling. Auto-Archive and further assay
b. Equiped 1 collection rack for 16 x 100 mm test tubes
c. Interface bundle and 1 teflon tubing harness (output to waster or return)
4. Media Mate 1 unit
a. Pre heat tank capacity : min. 10 L
b. Pre heat temperature range : min 30 °C ± 0.5 °C
c. Delivery volume : min. 250 mL
d. Accuracy : ± 1 %
e. Power consumption : min 220 Volt ± 10 %
f. dimension (W x H x D) cm : min 50 x 100 x 80
g. weight (kg) : min 100 kg
h. include of :
calibration gauge 26 glade set
5. Accessories : Validation tool kit 1 set
UV-vessel 8 pcs
Replacement filters 100 pcs
Base-kit, including device box
Glass tube and caps 250
pcs/pkg
Sinkers (ukuran kapsul OO) 16 pcs
Sample and return probe kit 8 pcs
Padle 8 pcs
Basket 8 pcs
E-prob actuator kit
Vessel cover 8 pcs
Include of : Installation
Training
Original operasional manual book Guarantee min. A year
Dust cover
Prednison tablet for calibration Alignment kit
15 DISINTEGRATION TESTER DIGITAL *) Unit 1
Spesifications
a. Number of basket : 2
b. Basket material : Stainless steel
c. Volume of beaker : min. 1000 ml
d. Stroke : min. 55 mm
e. Speed Stroke : 30 ± 1
f. Thermostat, Power Requirement : 230 V/115 V, 50/60 Hz
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)
Include of : Installation
Training
Original operasional manual book Guarantee min. a year
16 SET 1
Specifications :
1. Scrubber unit 1 unit
a. Centrifugal scrubber condenses and neutralizes
b. Works in two step, separating and washing out acid fumes c. Can be connect to the digestion unit
2. Digestion unit 1 unit
a. Kjeldahl Nitrogen and Protein Digestion unit b. Digital Display
c. Temperatur : maksimum 430°C
d. Tube volume : min 250 ml
e. Rapid and reliable process f. Acid resistant material
g. Heating place : 20 tubes
h. Voltage : 220 V - 240 V, 50/60 Hz
i. Controller : Electronic, with holder
j. Cooling water valve
3. Cooling Unit 2 unit
Include of : Installation
Training
Original operation manual book Dust Cover
Tool kit
Guarantee min. a year
17 DIGITAL MICROPIPETTE VOL. 1 - 10 µL UNIT 1
Specifications :
a. Volume range : 1 - 10 µL
b. Akurasi : < 1 %
Include of
a. Tips autoclavable : 1 box
b. Rak Tips Autoclavable : 1 pcs
18 DIGITAL MICROPIPETTE VOL. 5 - 50 µL UNIT 1
Specifications :
a. Volume range : 5 - 50 µL
b. Akurasi : < 1 %
Include of
a. Tips autoclavable : 1 box
b. Rak Tips Autoclavable : 1 pcs
19 UNIT 3
Specifications :
a. Volume range : 100 - 1000 µL
DIGESTION + SCRUBBER UNIT FOR PROTEIN/NITROGEN ANALYZER *)
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)b. Akurasi : < 1 %
Include of
a. Tips autoclavable : 1 box
b. Rak Tips Autoclavable : 1 pcs
20 DIGITAL DISPENSER PIPPETE 1 - 5 mL UNIT 1
Specifications : a. Volume : 1- 5 mL (adjustable) b. Imprecision (CV %) : < 0,1% c. Chemical Resistance d. Autoclavable Include of
a. Original operation manual book b. Garansi min 1 tahun
21 DIGITAL DISPENSER PIPPETE 2 - 20 mL UNIT 6
Specifications : a. Volume : 2- 10 mL (adjustable) b. Imprecision (CV %) : < 0,1% c. Chemical Resistance d. Autoclavable Include of
a. Original operation manual book b. Garansi min 1 tahun
22 DIGITAL DISPENSER PIPPETE 10 - 50 mL UNIT 1
Specifications : a. Volume : 10 - 50 mL (adjustable) b. Imprecision (CV %) : < 0,1% c. Chemical Resistance d. Autoclavable Include of
a. Original operation manual book b. Garansi min 1 tahun
23 DIGITAL DISPENSER PIPPETE 10 - 100 mL UNIT 1
Specifications : a. Volume : 10- 100 mL (adjustable) b. Imprecision (CV %) : < 0,1% c. Chemical Resistance d. Autoclavable Include of
a. Original operation manual book b. Garansi min 1 tahun
24 UNIT 9
Specifications :
a. Air flow fast : min. 270 sq.m/h
b. Air flow slow : min. 150 sq.m/h
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)c. Temperature operating range : 0 to 27⁰C
d. Power supply : 220 - 240 V; 50/60Hz
Include of
a. Original operation manual book b. Garansi min 1 tahun
25 DIGITAL VISCOMETER UNIT 1
Spesification: a. Display : LCD b. Accuracy : ± 1.0 % c. Repeatability : ± 0.2 % d. Viscosity range : 100 to 40.000.000 cPs e. Speed : minimal 0.01 rpm
Included of : 6 spindle for medium viscosity software
Spindle Guard Leg Lab Stand Carrying case
26 ELECTRONIC BURNER UNIT 6
Spesifications:
a. Technology : Foot pedal standard (Flame during pressed foot pedal)
Start stop with timer
b. Program Safety : safety control system with gas safety cut off
c. Gas supply and consumption : Gas connection : ¼” left with gas filter
d. Electrical : Power consumption : 2 VA
e. Power supply : 100-240 V, 50/60 Hz, 0.3 A
f. Acessories : - Gas LPG Cylinder 12 Kgs content
- Regulator with scale and pressure gauge - Rubber slank
Included of : - Instalation
- Training
- Guarantee min. a year - Original operation manual book
27 SET 1
Completed with Waterbath Each set consist of :
1 Extraction Soxhlet System : 1 unit
2 Hydrolysis system : 1 unit
3 Chiller/ Cooling Water : 1 unit
Specifications :
1. Extraction Soxhlet System
Consists of Extraction Unit and Control Unit
a. Extraction system : Fully automated
b. Solvent volume : minimum 100 ml
c. Capacity per-batch : 6 samples
d. Power supply of control unit : 230V/50/60 Hz
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)e. Accessories : Extraction beaker glass macro
Holder for extraction Thimbles Extraction thimbles 33 x 80 mm Pair of Tongs for extraction beaker Box of boiling stones 250 g Insert rack for 6 extraction beaker Tubing set complete
Line filter 16 A (just for 230V) Plug for safety temperature 135 °C Plug for safety temperature 200 °C Plug for safety Temperature 300 °C
Level control for solvent tank Acoustic and visual error messages
Compressor 2. Hydrolysis system
a. Max. Temperature : 425°c
b. Capacity perbatch : 6 samples simultaneusly
c. Power : 230 V/ 50/60 Hz
d. Accessories : Support & Forked Rod
3 Chiller
a. Cooling fixed temperature : max 7°C
Included of : Installation
Operation Training
Original operasional manual book Guarantee min. a year
28 FUME HOOD PROVIDE WITH MICROPROSSESORCONTROL *) UNIT 1
Spesifications:
a. Exterior dimension ( W in mm) : min. 1800
b. Interior work zone dimension (W in mm) : min. 1600 c. Exhaust air volume for minimum face
velocity : 500 cmh
d. Inner exhaust outlet diameter : min. 12.0"
e. Outer exhaust outlet diameter : min. 12.0"
f. Construction
- Main body : Coated steel
- Coating : Epoxy powder coating
g. Controller : Air flow alarm kit
Include of : - Instalation
- Guarantee min a year - Original operation manual book - Training
29 FUME HOOD PROVIDE WITH MICROPROSSESORCONTROL *) UNIT 1
Spesifications:
a. Exterior dimension ( W in mm) : min. 1500
b. Interior work zone dimension (W in mm) : min. 1300 c. Exhaust air volume for minimum face
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)d. Inner exhaust outlet diameter : min. 12.0"
e. Outer exhaust outlet diameter : min. 12.0"
f. Construction
- Main body : Coated steel
- Coating : Epoxy powder coating
g. Controller : Air flow alarm kit
Include of : - Instalation
- Guarantee min a year - Original operation manual book - Training
30 FIBER ANALYZER UNIT 1
Spesifications:
a. Configuration : Automatic acid addition
Automatic caustic addition Automatic rinse circle Automatic suction Cooling water control Complete function control
b. Cooling water consumption : min. 5 L/min
c. Compressed air pressure : min. 4 bar
d. Capacity Boiling vessel : min. 1,8L
e. Nominal wattage : max. 2000 W
f. Nominal voltage : 220 V ± 10%
g. Weight : max. 50 kg
h. Dimensions (WxDxH) : max. 400 x 700 x 900 mm
i. Interface : min. 2
j. Program : min. 9
k. Water connection : min. 2
l. Compressor : min. 4 bar
Included of : Fibrebags, small mesh size for determination of crude fibre 100 pcs Fibrebags, small mesh size for determination of ADF/NDF 100 pcs SPF Glassspacer
Set of storage tanks, 3 x 20 L with automatic level sensor Cooling water
Guarantee min. 1 year
Original operation/instruction manual Instalation
Training
31 FOGGING SYSTEM UNIT 2
Spesifications:
a. size : min. 420 x 200 x 290 mm
b. Power suppy : 800 w ( 220 V, 50 HZ)
c. Chemical tank : min. 6 Liter
d. Fogger speed : min. 0 - 400 mL/min
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)f. Net weight : min. 3,5 kg
g. Mist size : Less than 20 micron
Included of : - Surface Disinfectans
- Original operation manual book - Guarantee min. a year - Instalation
- Training
32 GAS CHROMATOGRAPY - MASS SPECTROMETER (GC-MS) *) Unit 1
Spesifications
a. Mode : EI, PCl dan NCl 1 set
b. Energi Ionisasi : 2 - 300 eV
c. Ion source Temperature : 100 - 400° C
d. Filaments : Dual
e. Dynamic range : 1-10 x 10^6
f. Scan rate : min 10.000 u/s
g. Mass Range : min 1000 m/z
h. Pumping System : Turbomolekular Pump
i. Turbomolekular Pump : min 150 L/s
j. Automatic Injector : 1 unit
k. Automatic Sampler : 1 unit
l. Multo Mode Inlet Multi Mode Inlet : 1 unit
Consumable and Accessories complete with :
a. FID Detector for GC 1 unit
b. Compressor Air 1 unit
c. Gas Supply (Helium, Hydrogen, Air)+ Cylinder, Regulator, Moisture and Trap 3 set
d. Tubing for gas line 1/8" 3 set
e. Column for GC-MS 3 unit
f. GC-MS Software (min Scan & SIM type) 1 set
g. 1 set
h. Tool Kit and Spare part 1 set
i. Syringe 10 uL, 6/pk 1 set
j. Vial screw 100/pk 2 set
k. Cap + septa 100/pk 2 set
l. Ferulle 0,5 (for ID 0,32 mm), 10 pcs/pkt 2 set
m. Ferulle 0,5 (for ID 0,25 mm), 10 pcs/pkt 2 set
n. Library : Willey 9 edition, NIST 2008 1 set
o. UPS 10 KVA 1 unit
Include of : Guarantee min 1 year
Installation
Original operating manual book Factory Certificate
After Sales Support On site Operator Training
33 SET 2
PC ( Processor min. Core i5, RAM min. 4 Gb DDR3, HD min. 500 Gb, DVD R+W multy, motherboard, LAN ) + LCD Monitor min 19 " + mouse + Keyboard + Printer inkjet ( color min. 5800 x 1400 dpi ) + OS compatible
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)
Each set consist of
1 Vaccum Pump
2 Glass Filter + Holder Specifications :
1. Vaccum Pump 1 unit
a. Power suply : 220 - 240 Volts, 50/60 Hz
b. Delivery : min. 6 L/min
c. Absolute Vaccum : 100 mbar
d. Noise level : max. 55 dBA
e. Dimensi (LxHxW) : min. 160 x 140 x 90 mm
2. Glass Filter Holder 1 set
a. Diameter : 47 mm
b. Joint connection : sesuai
c. Material : Glass (funnel and flask): Aluminium (clamp)
d. Funnel capasity : min. 300 mL
e. Flask capasity : min. 1000 mL
Included of
a. Garansi minimal 1 tahun b. Original operation manual book c. Sillicon tubing for vaccum
d. Membrane filter nylon 0.45µm diameter 47 mm 1 pack @ 100 pcs e. Membrane filter cellulose 0.45µm diameter 47 mm 3 pack @ 100 pcs f. Membrane filter PTFE 0.45µm diameter 47 mm 1 pack @ 100 pcs
34 PCS 6
Spesifications
a. Bahan : Stainless Steel
b. Ukuran (panjang, cm) : min. 14 cm
35 HEATING MANTLE Unit 2
Spesifications
a. Maximum temperature : 400°C atau lebih
b. Flask size : 500 mL
Included of : Guarantee min 1 tahun
Original operation manual book
36 UNIT 1
Specifications :
1. Pompa 2 unit
a. Tipe : Quartenary
b. Rentang Laju Alir : 0.1-5.0 mL/menit atau lebih besar
c. Akurasi Laju Alir : ± 1 %
d. Presisi Laju Alir : ≤ 0.1 % RSD
e. Gradient accuracy : ± 1 % atau lebih kecil
f. Jumlah wadah fase gerak : minimal 3
2. Detektor 1 unit
a. Lampu : D2 lamp
b. Rentang Panjang Gelombang : 200 - 600 nm atau lebih besar
c. Presisi Panjang Gelombang : ± 0.1 nm
GUNTING STAINLESS STEEL
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)3. Oven Kolom 1 unit
a. Rentang Suhu : 20 °C - 65 °C atau lebih besar
b. Temperature control precision : ≤ 0.1 °C
4. Autosampler 1 unit
5. Degasser 1 unit
6. Data Processor 1 unit
LAN ) + LCD Monitor min 19 " + mouse + Keyboard + Printer inkjet ( color min. 5800 x 1400 dpi ) 7. Software
Parameter setting. Wizard method or compatible
Included of :
a. Intalasi b. Pelatihan c. Dust Cover
d. Kolom C18, panjang 25cm, ukuran partikel 5µm, masing-masing 1 pcs e. Kolom C18, panjang 15cm, ukuran partikel 5µm, masing-masing 1 pcs f. Kolom C8, panjang 25cm, ukuran partikel 5µm, masing-masing 1 pcs g. Kolom C8, panjang 25cm, ukuran partikel 5µm, masing-masing 1 pcs h. Garansi minimal 1 tahun
i. Tool kits
j. Syringe 100 µL, 2 pcs k. Buku petunjuk pemakaian asli l. Factory certificate
m. Reagen untuk uji fungsi (metanol dan aquabidest) n. Manual Injektor o. Vial, 100 pcs 37 UNIT 1 Specifications : 1. Pompa 2 unit a. Tipe : Quartenary
b. Rentang Laju Alir : 0.1-5.0 mL/menit atau lebih besar
c. Akurasi Laju Alir : ± 1 %
d. Presisi Laju Alir : ≤ 0.1 % RSD
e. Gradient accuracy : ± 1 % atau lebih kecil
f. Jumlah wadah fase gerak : minimal 3
2. Detektor 2 unit
Detektor UV
a. Lampu : D2 lamp
b. Rentang Panjang Gelombang : 200 - 600 nm atau lebih besar
c. Presisi Panjang Gelombang : ± 0.1 nm
Detektor Fluoresence
a. Lampu : Xenon lamp
b. Rentang Panjang Gelombang : 200 - 650 nm atau lebih besar
3. Oven Kolom 1 unit
a. Rentang Suhu : 20 °C - 65 °C atau lebih besar
b. Temperature control precision : ≤ 0.1 °C
4. Autosampler 1 unit
5. Degasser 1 unit
HPLC UV + FLUORESENCE DETECTOR WITH AUTOSAMPLER *)
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)6. Data Processor
LAN ) + LCD Monitor min 19 " + mouse + Keyboard + Printer inkjet (color min. 5800 x 1400 dpi) 7. Software
Parameter setting. Wizard method or compatible
Include of a. Intalasi b. Pelatihan c. Dust Cover
d. Kolom C18, panjang 25cm, ukuran partikel 5µm, masing-masing 1 pcs e. Kolom C18, panjang 15cm, ukuran partikel 5µm, masing-masing 1 pcs f. Kolom C18, panjang 25cm, ukuran partikel 5µm, masing-masing 1 pcs g. Kolom C18, panjang 25cm, ukuran partikel 5µm, masing-masing 1 pcs h. Garansi minimal 1 tahun
i. Tool kits
j. Syringe 100 µL, 2 pcs k. Buku petunjuk pemakaian asli l. Factory certificate
m. Reagen untuk uji fungsi (metanol dan aquabidest) n. Manual Injektor
o. Vial, 100 pcs
38 UNIT 6
Specification:
a. Dimensions (cm) : Min. 300 (W) x 250 (D) x 100 (H)
b. Heating plate material : Ceramik
c. Stirring capacity with H₂O : Min. 15 L
d. Speed range : Min. 100 - 1500 rpm
e. Temperature range : Min. 50 - 500 °C
Include of
a. Original operation manual book b. Garansi min 1 tahun
c. Magnetic bar : 4 pcs/ unit
39 INOCULATING LOOP WITH HOLDER UNIT 24
Spesifications :
a. Material loop : Stainless Steel
- size: Ø loop (mm) ,length (mm) : 1,5mm, 50mm
b. Material handle : Nickel Plated Brass
- length (mm) : 230 mm
40 INOCULATING NEEDLE WITH HOLDER UNIT 24
Spesifications :
a. Material Needle : Stainless Steel
- size: length (mm) : 50mm
b. Material handle : Nickel Plated Brass
- length (mm) : 230 mm
min. Core i5, RAM min. 4 Gb DDR3, HD min. 500 Gb, DVD R+W multy, motherboard,
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)
41 INKUBATOR UNIT 1 Spesifications : a. Interior dimensions - Capacity (L) : min. 400 L - w x h x d (mm) : min. 1000 x 800 x 500 mm b. Exterior w x h x d (mm) : min. 1100 x 1000 x 650 mm c. Power (W) : min. 1800 d. Electric supply : 230V 1 Ph e. Weight (kg) : max. 125 kg
f. Temperature control Information
- No. Of temperature sensor PT 100 : 2
- Display Resolution (⁰C) : max. 0.1
- Temperature Accuracy at 37 ⁰C(<=±⁰C) : max. 0.05 - temperature Uniformity at 37 ⁰C(<=±⁰C) : max. 0.7
g. Feature for Incubator :
- Manually adjustable fresh air supply via pre-heating chamber - Fan speed control in 10% steps (UF)
- Fuzzy - supported PID microprocessor controller - 2 PT 100 sensors Class A
- Temperature range from 30 ⁰C up to 70 ⁰C - 7-day-programme timer with real time clock
- long-term documentation of all relevant data via ring store 1024 kB - Works calibration certificate at 37 ⁰C
- Adaption of effective heating power depending on setpoint - IQ/OQ check list for validation by customer (Optional)
Include of : - Conection cable for computer interface - Instalation
- Guarantee min a year - Original operation manual book - Shelves stainless steel 2 pcs - Stainless steel Support stand
42 KOMPOR LISTRIK UNIT 1
Specifications :
a. double over heating with thermostat and thermofuse b. non stick cast iron heating system
c. indicator light
d. power suply : 220 ± 10 % / 600 W
43 LIQUID CHROMATOGRAPHY MASS SPECTROMETER (LCMS-MS) *) Unit 1
Specifications : 1. Mass Spectrometer
a. Triple Quadrupole
b. Ionization Method : min. ESI & APCI (without source exchange)
c. MS Sensitivity (signal to noise ratio) : (ESI +) max. 200 m/z (APCI+) max. 150 m/z
- Integral programme timer for one profile with 4 segments, each 1 min. Up to 999 hours
- Triple overtemperature protection - extra relay following the setpoint , electronic overtemperature controller TWW protection class 3. 1 and mechanical temperature limiter TB class 1
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)d. Mass Resolution : Max. 1 Da FWHM
e. Mass Range : min. 10 - 2.000 m/z or better
f. Mass Stability : < 0,1 Da/ 24 h or better
g. Scan rate : min. 5.200 Da/s
2. Pump
a. Flow rate setting range : min. 0,001 to 2.000 mL/min or better
b. Flow rate precision : max. 0,075% RSD
c. Flow rate accuracy : ± 1,0 %
3. Column
a. Operating temperature range : min. 4°C - 35°C atau dengan rentang yang lebih lebar
b. Temperature stability/precision : ± 0,05 °C or better
c. Temperature acuracy : ± 0,5 °C with calibration
4. Detector
a. Number of Diode elements : min. 512
b. Wavelength range : 190 - 640 nm atau dengan rentang lebih lebar
5. Auto Sampler/ Auto Injektor
a. Injection-volume setting range : 0,1 - 10 µl atau dengan rentang yang lebih lebar
b. Cross Contamination/carry over : < 0,004 % or better
6. Accessories : UPS 10 KW 1 unit
Nitrogen Generator 1 unit
Gas Argon UHP + Cylinder 2 set
Pressure Regulator Argon 2 unit
Sample vial 500 pcs
PC ( Processor min. Core i5, RAM min. 4 Gb DDR3, HD min. 500 Gb, DVD R+W multy, motherboard, LAN ) + LCD Monitor min 19 " + mouse + Keyboard + Printer inkjet ( color min. 5800 x 1400 dpi ) + OS compatible
1 set
HUB w/ cable 1 set
Software
Column C18 6 pcs
Column C8 3 pcs
Solvent Filtration 1 unit
Reservoir Tray 1 unit
Botol Eluen 5 btl
Ultrapure Water Purification Apparatus 1 set
Include of : Guarantee min 1 year
Tool kits Installation
Original operating manual book Factory Certificate
After Sales Support On site Operator Training
44 MUFFLE FURNACE UNIT 1
Spesifications:
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)b. Time to Temp max : max. 65 min
c. Volume : min. 10 L
d. Power Electrical : 220 V ± 10%; max. 60 Hz
e. Opening door : vertical
Include of : ceramic shelf
Guarantee min. 1 year Original operation manual book
45 MICROWAVE DIGESTION *) Unit 2
Spesifications
a. Magnetron supply : Dual magnetrons, min. 1800 watt
b. Controller and display : LED, acid resistant touch screen
c. Data/method export/import : USB Port and Ethernet port
d. Loading capacity : min. 24 vessels per loading
e. Vessels size : min. 100 mL
f. Accesoris : Vessels 24 set, 100 mL
Turntable with 24 composite sleeves
Include of : Installation
Operation Training
Original operasional manual book Guarantee min. a year
46 UNIT 1
Specifications :
a. Balance provides with : Level sensor
Automatic self Calibrating Printer port
Touch screen
b. Weighing capacity : min. 5 gram
c. Readability : 0.001 mg
d. Repeatability : maksimum 1 µg
e. Linearity : maksimum 4 µg
Include of
a. Garansi minimal 1 tahun b. Original operation manual book c. Installation d. Training e. Dust Cover f. Sertifikat Kalibrasi 47 SET 1 Consist of:
a. Afla Test-P HPLC Kit-100 Column & 1 set standards 1 set
b. Ocra Test HPLC Kit-100 Column & 1 set standards 1 set
c. Afla Test M1 Kit-100 Column & 1 set standards 1 set
d. 10x Concentrate of PBS 7 btl
e. Microfibre Filter 1.5 micron 11 cm (100 pcs/box) 3 box
f. Fluted Filter Paper 24 cm (100 pcs/box) 3 box
g. HPLC Post Column Derivatisation (w/1 setspare part) 1 unit
MYCOTOXIN EQUIPMENT MICRO BALANCES *)
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)h. HPLC C-18 Column (4.6 x 250 mm, 5 µm, Ultra High Silica) 4 unit
i. 6-Position Pump Stand (w/ WB Column Coupling) 1 unit
j. 4-Position Pump Stand (w/ WB Column Coupling) 1 unit
k. Glass Syringe 10 ml 12 pcs
l. Disposible Cuvettes 250 pcs
m. Stainless Steel Blender w/ Stainless Steel Jar (1 L) 1 set
Include of : Installation
Training
Original operation manual book Tool kit
Guarantee min. a year
48 MICROBIAL AIR SAMPLER *) UNIT 1
Spesifications:
a. Height : min. 25 cm
b. Diameter : min. 11 cm
c. Weight : min. 2.38 kg
d. Material : Anodized aluminium
e. Diameter of sampling head : 10 cm
f. Nominal air flow : 100 L/min ± 2.5 % validated to ISO 14698
g. Standar sampling volume : 50, 100, 250, 500, 1000 lters
h. Freely definable sampling volume : 1 - 2,000 liters
i. Battery Pack : Li-ion, Rechargable battery, 7.4 V/ 6.9 Ah
j. Motor : 6 V
k. Display characters : Alphanumeric liquid cryst displ, 32 characters
l. Lifetime RTC battery : RTC (Real Time Clock) battery; good for about 10 years
m. Driving motor : PWM frecuency for driving motor
n. Airflow regulation : Hot- wire anemometer , numeric contr.
o. Power Unit/battery Charger : 110-240 volt, 50-60 KHz
Include of : - Instalation
- Guarantee min a year - Original operation manual book - Training 49 PCS 2 Spesifications a. Volume range : 50 - 300 µL b. Channel : 8 channel 50 OVEN UNIT 1 Spesifications: a. Dimensions - Interior (W x H x D) : min. 1000 x 800x 500 mm - Capacity (L) : min. 416 L
- 2 stainlees steel grids
Exterior stainless steel casing :
- fully insulated stainless steel doors with double locking and 4-point adjustment
MULTI CHANNEL PIPPETOR
- easy - to- clean interior, made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four side
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)- rear zinc-plated steel
b. temperature range : from + 30 ⁰C (however, at least 5 ⁰C above ambient) up to +
250 ⁰C
c. Voltage/Power rating :
- 400 V (+/10%), 3 phases N, 50/60 Hz - ca. 4000 W (during heating)
- adaption of effective heating power depending on the set temperature d. Ventilation and control
- natural convection
- continous adjustment of pre-heated fresh air admixture - vent connection with restrictor flap
- solid state switching unit
- "Loop" fuction to repeat profile 1 - 99 times or endless
f. Accessories : - Stainless steel shelves 2 pcs
Included of : - incl. Works calibration certificate for + 160 ⁰C - Instalation
- Guarantee min. a year - Original operation manual book - Support stand stainless steel
51 UNIT 1
Specifications :
a. Motion : orbital motion
b. Speed : 20 to 400 rpm atau rentang speed yang lebih luas
c. Load Capacity : minimal 10 kg
d. Power : 220 ± 10 % Volt, 50/60 Hz
e. Display : digital
f. Accessories : - frame with tension roller 1 set
- spare tension roller 6 pcs
Included of
a. Installation b. Garansi min 1 tahun
c. Original operation manual book
52 UNIT 4
Specifications :
a. pH Range : -2 - 19,999 atau range yang lebih lebar
- Integrated digital timer for temperating profile with max.4 segments, each segment adjustable hour from 1 min to 999
- digital display (LED) of all set parameters, such as temperature , week days, time, programme status and set-up value
- resolution of display for set value 0,5 ⁰C ; for actual value 0,1 ⁰C below 99,9 ⁰C, 0,5 ⁰C above 100 ⁰C - Long-term logging (ring store) of all relevant data, GLP-conforming as data logger 1024 kB
ORBITAL SHAKER
pH METER
- fuzzy-supported PID microprocessor controller with integrated autodiagnostic system with fault indicator
- 2 Pt 100 sensors Class A 4-wire-circuit, mutually monitoring and taking over the perfomance at the temperature value
- digital 7-day-programme-timer with real time clock , precise minute setting,for one set value or ramp operation in combination with
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)b. mV Range : - 1999 to 19999 atau range yang lebih lebar
c. pH Relative Accuracy : ± 0.005 atau lebih baik
d. pH Resolution : min. 0.001
e. mV Resolution : min. 0.1
f. mV Relative Accuracy : min. 0.1
g. Temperature Range : -5.0 to 105 °C atau range yang lebih lebar
h. Temperature Resolution : min 0.1
i. Temperature Accuracy : ± 0.2 °C atau yanglebih baik
j. Display : Custom LCD
k. Interface : Connectin to printer or PC
Included of
a. Installation b. Garansi min 3 tahun
c. Original operation manual book d. Buffer pH 4, pH 7, pH 9 e. Storage solution f. Electrolyte solution
53 PLATINA INOCULATING LOOP WITH HOLDER UNIT 2
Spesifications :
a. Material : Platinum
b. Diameter : 5 mm, 60 mm long, 0,5 mm dia wire
c. Holder : metal-Alumunium, long 250 mm
54 PCS 6
Spesifications
a. Bahan : Stainless Steel
b. Ukuran (panjang, cm) : min. 18 cm
55 UNIT 1
Consist of Specification :
1. Evaporator 1 unit
a. Hand/motor lift : motor lift
b. height adjustment speed : 30 mm/s
c. height adjustment : 155
d. rotary speed : 20 - 280
e. rotary speed setting : 4.3" LCD color graphic display
f. drive : brushless DC motor with electronic speed control
g. heating capacity : 1300
h. temperature range heating bath : 20 - 210 °C
i. temperature accuracy bath : ± 1
j. overheat protection bath : cutt-off at 5 °C over set temperature via separate PT 1000
k. bath temperature setting : 4.3" LCD color graphic display
l. heating bath temperature control : electronc / digital
m. secondary overtemperature cutt-off : 250
n. material heating bath : stainless steel AISI 316 L
o. diameter heating bath : 255
PINSET
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)p. volume heating bath : 4,5
q. power supply : 1400
r. size : minimum 395 x 490 x 430
s. protection class : IP 20
t. operation conditions : 0 - 40 at 80 % rel. humidity
u. Supply voltage : 230 V
v. Include of :
Auto Acurate sensor : 1 unit
2. Vaccum Pump 1 unit
a. suction capacity : min. 1.7 m3 / hour
b. power input : min. 160 W
c. weight : min. 5,4 kg
d. archive an ultimate vacuum : minimal 2 mbar
e. accessories : condensor
3. Chillers 1 unit
a. temperature range : -10 °C to + 40 °C
b. precise temperature control : ± 0.5 °C
c. footprint of chiller : min. 470 x 580 x 440 mm
4. Accessories
a. condensor for vacuum pump 1 unit
b. evaporating flask 250 mL 1 unit
c. receiving flask 500 mL (standard) 1 unit
d. spider with 5 flask for 100 mL 1 unit
Include of
a. Original operation manual book b. Garansi min 1 tahun
c. Training d. Instalasi
56 PCS 4
Spesifications
a. Bahan : Polypropylene dan PVC
b. Capacity : max. 94 pipet
c. Dimensi
Height Stand : min. 48 cm
Rotating Disk : min. 18 cm
Base : min. 23 cm
57 ROD FOR STAND + STAND PCS 2
a. Material : Rod : Aluminium; Stand : Stainless steel
b. panjang (mm) : 800 mm
c. Diameter : 12 mm
d. Rectangular base stand
58 UNIT 2
Specifications :
a. capasity : 100 - 120 g
b. Readability : 0.01 mg/ 0,1 mg
c. Repeatability : maksimum 0.015 mg
ROTARY PIPPETE STAND
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)d. Linearity : ± 0.1 mg
f. Display : LCD
g. Minimum weight (according to USP) : 45 mg
h. Minimum weight (@ U=1%, 2 SD) : 3 mg
i. Sensitivity offset : 0,3 mg (100 g)
Included of
a. Installation b. Garansi min 1 tahun
c. Original operation manual book d. Dust Cover
e. Weight Standard Class E2 Capasity 100 & 200 g
59 UNIT 1
Spesifications:
a. Useful capacity : min 400 mL
b. Power supply : 220 V. 50/60 Hz
c. Programmable padle speed and time
d. Speed : 4 - 10 strokes/s
e. Blending time : 1 sec - 1 hour
f. Security drip tray
Included of : - Guarantee min a year
- Original operation manual book
- Storage Rack for 10 bags 400 mL, stainlees steel 3 unit
60 SEROLOGICAL ELECTRICAL PIPPETTE UNIT 22
Spesifications:
a. Battery : Lithium battery
b. Runtime between charges : min. 12 hours of continous use or better
c. Recharge time : Approxinately 1 hour
d. Volume Range : 0,1-100 ml (available glass pipette)
Include of : - Original operation manual - Charger
- Wall mounted holder and table mounted stand
61 STERILIZER, DRY BEAD UNIT 4
Spesifications :
a. Physicochemical properties
- Dimensi (H x W x D ) : min. 20,5 x 12,5 x 13,5 cm
- Diameter chamber : min. 14 cm
b. General description :
- plunge metal instruments into hot glass beads for instant sterilization
- The glass beads can be cleaned by any cleaning liquid (or soapy water)
- Chamber deep for sterilizing small and medium sized instruments - Supplied with a 2-pin EU plug
Include of :
- Just rinse the glass beads thoroughy with hot water, dry completely and refill the sterilizer
STOMACHER
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)- Units are supplied with one package of glass beads
- Original operation manual book - Guarantee min. a year - Instalation
- Training
62 SET 1
Specification:
a. Vacuum Extraction Manifold set complete with min. 20 position 1 pcs
Include of : min. 20 needles tips,
min. 25 plug and ejector tool
b. Rack of minimal 20 position 1 pcs
c. Reservoir Adaptor for suitable cartridge 2 pack
d. Vacuum pump with power suply 220 ± 10 % Volt, 50 Hz 1 unit
e. Valve with needle tips 1 pack
Include of
a. Cartridge SPE Hydrophillic and Lipophilic, min. 20 cc / 1 gram 2 box
b. Cartridge SPE Hydrophillic and Lipophilic, min. 35 cc / 6 gram 2 box
c. Reservoir Cartridge for min. 30 cc 2 pack
d. Reservoir Cartridge for min. 60 cc 2 pack
63 SHOWCASE COOLER 2 PINTU UNIT 1
Spesifications
a. Capacity : minimal 1000 liters
b. Dimensi (P x L x T) mm : minimal 1100 x 700 x 2000 mm
c. Total input power (w) : ± 800 watt
d. Interior temp. : ± 2 - 8 ⁰ C
Include of
- Original operation manual book - Guarantee min. a year - Rak
64 SPATULA WITH SPOON PCS 8
Spesification
a. Material : Stainless steel
b. panjang (mm) : min. 210
c. Width (mm) : 29
65 SPATULA WITH MICRO SPOON PCS 6
Spesification
a. Material : Stainless steel
b. panjang (mm) : min. 150
c. Width (mm) : 5
66 TOTAL ORGANIC CARBON (TOC) METER *) UNIT 1
Spesifications:
a. Measurement method : Combustion - Non Dispersive Infrared Detection (NDIR)
b. PC Controlled
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)c. Measured item : TC, IC, NPOC, POC
d. Measurement Range : TC = min. 30,000 mg/L
IC = min. 35,000 mg/L POC = min. 500 mg/L
e. Detection Limit : ± 4 µg/L
f. Reproducibility : max. 1,5 %
g. Sample injection : automatic, volume 10 - 2000 µL
h. Automatic addition of acid
i. Automatic sample dilution with accuracy max. 2%
Include of : High purity Carrier gas (CO, CO2, HC) + cylinder + gas pressure regulator installed
2 unit
CO2 Absorber 1 unit
Computer + Printer 1 unit
Pipa Udara, stainless steel 1 set
Gas purifier 1 unit
Molecular sieves 13x 500 g
Compressed Air 1 unit
Regulator Udara 1 unit
Stabilizer 1,5 KVA 1 unit
Installation On site Training
Original Operasional manual book Guarantee min. 1 year
Factory Certificate
67 UNIT 3
Spesifications:
a. Speed : 2500 rpm or better ( adjustable )
b. Power : 220 ±10% V. 50/60 Hz
c. Stirring volume : up to 2000 mL
d. Material : heavy cast- aluminium
e. Accessories : stand attachment
round base for flat foam & test tube foam insert flat foam insert attach
test tube foam attach microtiter plate base stirrer base
Included of : Guarantee min a year
Original operational manual book
68 UNIT 4 Specifications : a. Capacity : max. 2200 g b. Readibility : 10 mg c. Reapeatility : maks. 10 mg d. Reproducibility : maks. 0.01 g e. Linearity : maks. 20 mg f. Display : LCD TOUCH MIXER
No. SATUAN JUMLAH ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) + AUTOSAMPLER + GFA*)g. Internal calibration and adjustment fully automatically with ISO Call
h. Respon time : max.1,5 s
Included of : - Dust cover
- Guarantee min 1 a year - Original operational manual book
69 UNIT 12
Spesifications :
a. Display : Dual display
b. Temperature range
- Temperature range : (-) to 50 up to 70⁰C or better
- Accuracy : ± 0.1 ⁰C
c. Relative Humididy
- Humidity range : up to 99 % RH
- Accuracy at 20 ⁰C : ± 1 % RH
: - Guarantee min 6 month - Battery 2 pcs
70 PCS 4
Spesifications
a. Material : stainlees steel
b. Ø /Dime (mm) : 10 mm/ 11,5 mm
c. Quantity pcs : 50 tube
71 PCS 4
Spesifications
a. Material : stainlees steel
b. Ø /Dime (mm) : 15 mm/ 16,5mm
c. Quantity pcs : 50 tube
72 PCS 10
Spesifications
a. Material : stainlees steel
b. Ø /Dime (mm) : 21 mm/ 22,5 mm c. Quantity pcs : 50 tube 73 UNIT 1 Specification: a. wavelenght range : 190 - 1100 nm b. Slit width : 1 nm
c. wavelenght accuracy : ± 0.3 nm for entire range and ± 0.1 nm at D2 peak 656.1 nm
atau 0.5 at 541.94 nm
d. stray light : less than 0.02% at 220 nm (NaI), less than 0.02 % at 340 nm
(NaNO2), less than 1.0 % at 198 nm (KCl) atau < 0.03 % at 340 nm (NaNO2), < 0.05 % at 220 nm (NaI)
e. photometric system : double beam optics
f. photometric range : absorbance : - 4 to 4 Abs, Transmittance 0 % to 400 %
g. photometric accuracy : < ± 0.005 A
h. baseline flatness : < 0.001 A or better
TEST TUBE RACK
TEST TUBE RACK
TEST TUBE RACK
UV - VIS SPECTROFOTOMETER *) THERMOHYGROMETER