• Tidak ada hasil yang ditemukan

PENGADAAN SIMULTANEOUS INDUCTIVELY COUPLED PLASMA OPTICAL EMISSION SPECTROMETER (ICP - OES)

N/A
N/A
Protected

Academic year: 2022

Membagikan "PENGADAAN SIMULTANEOUS INDUCTIVELY COUPLED PLASMA OPTICAL EMISSION SPECTROMETER (ICP - OES)"

Copied!
8
0
0

Teks penuh

(1)

Page 1 of 8

KERANGKA ACUAN KERJA

PENGADAAN SIMULTANEOUS INDUCTIVELY COUPLED PLASMA OPTICAL EMISSION SPECTROMETER (ICP - OES)

PT AntamTbk

Geomin &Technology Development Unit Laboratory Bureau

Jl. Pemuda No. 1 Pulogadung, Jakarta Timur 13250, Indonesia

Jakarta, 24 Februari 2022

DISIAPKAN OLEH : DISETUJUI OLEH : MENGETAHUI :

Rahmat Fauzi Chemical & Instrument

Laboratory Dept Head

Rustini

Laboratory Bureau Head

Abdul Bari General Manager

(2)

Page 2 of 8

KERANGKA ACUAN KERJA (TERM OF REFERENCE)

PENGADAAN SIMULTANEOUS INDUCTIVELY COUPLED PLASMA OPTICAL EMISSION SPECTROMETER (ICP-OES) & ACCESSORIES

I. LATAR BELAKANG

Laboratorium PT ANTAM Tbk Unit Geomin & Technology Development merupakan rangkaian penting dalam proses eksplorasi untuk menghasilkan data hasil analisis yang akurat dan presisi dalam menentukan perhitungan cadangan. Untuk menjamin hasil pengujian yang akurat & presisi sesuai persyaratan SNI ISO/IEC 17025:2017 maka alat/instrument analisis yang digunakan harus memenuhi standar serta sesuai dengan kriteria yang dipersyaratkan dan selalu dimutakhirkan.

Salah satu alat/instrument yang digunakan laboratorium yaitu Simultaneous inductively coupled plasma optical emission spectrometer (ICP-OES) yang berfungsi untuk

mengukur/analisis kandungan mineral dalam sampel dengan kadar sampai level ppm dengan cara sampel yang sudah dilarutkan dengan bahan kimia tertentu dirubah kedalam bentuk aerosol yang kemudian dengan adanya energi akan diatomisasi dengan bantuan sinar dari panjang gelombang tetentu akan diemisikan dan sinar emisi yang dihasilkan akan dterima oleh detector yang akan dikonversi kedalam satun konsentrasi larutan.Konsentrasi larutan akan diperhitungkan dengan volume dan berat sampel yang akan didapat kadar suatu unsur//element

II. TUJUAN PENYUSUNAN TOR

Secara umum tujuan penyusunan TOR pengadaan Simultaneous inductively coupled plasma optical emission spectrometer (ICP-OES) adalah sebagai pedoman bagi mitra

kerja dalam proses pengadaan agar sesuai target yang telah ditetapkan dan disetujui oleh PT Antam Tbk. Unit Geomin & Technology Development.

Secara detil tujuan penyusunan TOR pengadaan Simultaneous inductively coupled plasma optical emission spectrometer (ICP-OES)adalah

(3)

Page 3 of 8

1. Sebagai salah satu pedoman bagi Mitra Kerja dalam membuat penawaran harga.

2. Sebagai acuan bagi mitra kerja dalam memenuhi permintaan PT Antam Tbk. Unit Geomin & Technology Development.

3. Sebagai salah satu dasar pembuatan draft kontrak kerjasama.

4. Sebagai pedoman bagi Tim Pegadaan untuk mengevaluasi penawaran dari para calon mitra kerja atau peserta lelang.

III. RUANG LINGKUP PEKERJAAN

1. Pengadaan barang sesuai spesifikasi hasil pelelangan 2. Pengiriman barang

3. Install dan training pengoperasian alat dan aplikasi dengan CRM

4. Garansi minimal 1 tahun & after sales service support serta respon maksimal 2x24 jam selama masa garansi bila ada kendala alat.

5. BASTP

IV. IV. SPESIFIKASI & JUMLAH ALAT

Jumlah Spesifikasi

SIMULTANEOUS INDUCTIVELY COUPLED PLASMA OPTICAL EMISSION SPECTROMETER (ICP-OES) WITH 5 CHANNEL PERISTALTIC PUMP

1 unit Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) Simultaneous System; CCD Detector; 5-Channel Peristaltic Pump and Mass Flow Gas Control System

Vertical Dual View (VDV), which offers deep insights from the use of smart tools and is also upgradable on-site to the Synchronous Vertical Dual View (SVDV) configuration.

feature the unique 3rd generation detection system, that delivers fast simultaneous measurement over the full 167 to 785 nm wavelength range. Inclulude the vertical torch and a new optical design that includes Freeform optical technology.

Fully PC-controlled Five channels peristaltic pump with variable

(4)

Page 4 of 8

speed from 0–80 rpm for sample, drain, internal standard/ ionization buffer, and MSIS vapor generation solution.

RF generator

27 MHz Solid State, maintenance free, water cooled RF generator.

Power output of 750-1500 W in 10 W increments. Robust free running design rapidly reacts to changes in plasma load, providing stable and consistent power supply into the plasma when switching between samples of high or varying matrix. Coupling efficiency of more than 75%.

Optical system

Vertical dual view pre-optics allow axial and radial plasma viewing from the vertical torch. Unique Dichroic Spectral Combiner (DSC) technology allows you to run axial and radial view analysis at the same time in SVDV mode, for the fastest analysis and lowest gas usage. Cooled Cone Interface (CCI) prevents the cooler plasma tail from being viewed by the optics when viewing axially, reducing interferences and increasing linear dynamic range.

CCD detector

Systems that feature the VistaChip III detector. This is a high speed, continuous wavelength coverage CCD detector with anti blooming protection on every pixel. It is a zero gas consumption design and enables fast warmup, high throughput, high sensitivity and the largest dynamic range.

 Utilizing Image Mapping Technology (I-MAP), the photosensitive pixels are arranged to exactly match the image from the echelle optics. This provides full wavelength coverage from 167–785 nm on a single detector, from a single entrance slit.

 Adaptive Integration Technology (AIT) allows intense and trace signals to be measured simultaneously at the optimum signal to noise ratio.

Software

ICP software has a familiar worksheet interface, easy method development and software applets that include pre-set method templates, saving you time. Completed with PC and Printer (LED monitor 32 inch, mouse & keyboard wireless).

Power requirement

2.9 kVA, single phase mains input voltage of between 200-240 VAC (50-60 Hz), drawing a maximum of 15 amps.

ACCESSORIES FOR OPERATION :

1x ICP Exhaust System Kit (Blower and ducting system) 1x Recirculating refrigerated water cooler system

1x Argon Regulator

(5)

Page 5 of 8

3x ICP-OES Wavelenght Calibration Standard

2x PVC peristaltic pump tube,grey (sample channel)(12/pk) 2x PVC peristaltic pump tube,blue (drain channel) (12/pk) 1x UPS 6 KVA

1x IT Trafo 6 KVA Complete with :

1 Set Inert Sample Introduction Kit for HF digest

 Inert double pass spray chamber with ball joint socket

 OneNeb Series 2 nebulizer (non AVS 6/7 use)

 HF resistant semi-demountable Easy Fit torch with 1.8 mm ID ceramic injector, longer outer tube with slot for radial viewing

 Large torch clamp for inert torch

Solvaflex peristaltic pump tubing for sample (black/black tabs) and waste (grey/grey tabs).

1 Unit SPS-4 AUTOSAMPLER

 Sample capacity : 4 rack support 360 samples

 Flexible rack configuration enables wide range of sample capacities

 Integrated environmental enclosure option protects your samples and your laboratory environment

 Dual-wash reservoir option eliminates potential carryover

 Three-channel peristaltic pump for ultimate flow-through rinse flexibility

 Multiple probe size options for a diverse range of applications Complete with : Integrated Cover Kit & Extraction Duct Housing.

1 unit Multi-mode Sample Introduction System (MSIS) MSIS provides simultaneous measurement of hydride and non-hydride elements including As, Se, and Hg to sub-ppb levels. Simultaneous measurement eliminates changeover and allows routine and hydride elements to be determined simultaneously using the same setup.

V. HAK DAN KEWAJIBAN PT. ANTAM Tbk UNIT GEOMIN & TECHNOLOGY DEVELOPMENT

1. Menetapkan jadwal kedatangan alat dari perusahaan pemenang pengadaan.

(6)

Page 6 of 8

2. Menegur mitra kerja apabila tidak sesuai atau lalai dalam hal melaksanakan pekerjaan yang diberikan.

3. Berhak untuk menghentikan pekerjaan Mitra Kerja sebagai pemenang, apabila tidak sesuai dengan lingkup pekerjaan yang telah disetujui.

4. Wajib melakukan pembayaran kepada Mitra Kerja yang akan diatur lebih lanjut dalam kontrak

5. Berhak mengembalikan barang/alat kepada perusahaan pengadaan atau mitra kerja apabila barang/alat tidak sesuai spec yang diminta.

VI. HAK DAN KEWAJIBAN MITRA KERJA

1. Membuat penawaran harga sesuai dengan permintaan.

2. Membuat time line waktu pelaksanaan pekerjaan yang berisikan:

a. Indent alat b. Pengiriman alat

c. Instalasi alat dan ducting d. Training

e. Trial/validasi dengan CRM f. BASTP

3. Mengirim dan menjamin bahwa mesin/alat yang akan diserahkan ke PT. Antam Tbk Unit Geomin & Technology Development sesuai spec yang diminta.

4. Melaksanakan install mesin/alat dan memberikan training kepada analis di laboratorium PT Antam Tbk. Unit Geomin & Technology Development di Pulogadung.

5. Wajib mematuhi peraturan K3 PT. Antam Tbk Unit Geomin & Technology Development dan biaya yang timbul akibat pelanggaran dari ketentuan ini menjadi beban dan tanggungjawab mitra kerja.

6. Wajib menanggung asuransi selama perjalanan dan menanggung segala resiko dan biaya yang timbul apabila terjadi kecelakaan saat pengadaan mesin/alat.

7. Wajib memberikan garansi minimal 1 tahun & after sales service support serta respon maksimal 2x24 jam selama masa garansi bila ada kendala alat.

8. Mendapatkan informasi dan akses untuk melaksanakan pekerjaan.

(7)

Page 7 of 8

9. Membuat berita acara serah terima barang.

10. Mendapatkan pembayaran atas hasil pekerjaan, sesuai dengan aturan yang ada.

VII. KUALIFIKASI PERUSAHAAN MITRA KERJA

1. Mitra kerja harus memiliki semua ijin yang diperlukan dan dipersyaratkan oleh peraturan perundang-undangan yang berlaku dan sesuai dengan buku rantai pasokan PT. Antam Tbk no 333.K/92/DAT/2017.

2. Mitra kerja merupakan agen/distributor yang dibuktikan dengan surat keagenan/dukungan.

3. Berpengalaman lebih dari 2 tahun dalam menyuplai alat/barang , dibuktikan dengan melampirkan bukti list kontrak .

4. Memiliki referensi perusahaan yang telah menggunakan alat sesuai spesifikasi di dunia industri (Populasi alat) .

VIII. KRITERIA PENILAIAN TEKNIS DAN HARGA

NO URAIAN BOBOT (%)

1 Spesifikasi alat/mesin, harus sesuai dengan yang di TOR

dan hasil andwidjing bila ada tambahan Mandatory

2 Surat keagenan atau dukungan dari agen serta populasi alat Mandatory

3 Waktu penyelesaian pekerjaan dan time line proses

pengadaan Mandatory

4

Masa garansi/ jaminan adalah minimal 1 tahun dengan respon service bila ada kendala maksimal 2x24 jam selama masa garansi

Mandatory

100 Kriteria Penilaian Teknis :

Harga

(8)

Page 8 of 8

IX. CARA PEMBAYARAN

1. Mitra kerja membuat invoice dengan dilampiri dokumen pendukung.

2. Pembayaran dilakukan setelah invioce beserta bukti – bukti pendukung dari perusahaan mitra kerja dinyatakan lengkap dengan mengikuti siklus pembayaran di keuangan PT Antam Tbk. Unit Geomin & Technology Development.

X. JANGKA WAKTU

Jangka waktu pengadaan alat Simultaneous inductively coupled plasma optical emission spectrometer (ICP-OES)paling lama 120 (seratus dua puluh) hari kalender setelah terbit PO.

XI. LOKASI PEKERJAAN

Lokasi penyerahan pekerjaan dan training di laboratorium PT Antam Tbk. Unit Geomin

& Technology Development, Jl. Pemuda no. 1 Pulogadung Jakarta Timur.

XII. LAIN-LAIN

Persyaratan Administrasi yang bersifat umum tertuang di dalam dokumen pelelangan dan pada saat aanwijzing.

Referensi

Dokumen terkait

In the present study, extraction of La, Ce and Nd with DBDTC as chelating extractant using organic solvents toluene, diethyl ether and petroleum ether, and

Penelitian ini meliputi proses dekstruksi kering kemudian analisis kuantitatif kobalt, molibdenum, kalium dan natrium dilakukan dengan menggunakan metode Inductively Coupled Plasma